Summary of Contents for inbody 970S

  • Page 1 User's Manual...
  • Page 2: Intended Use

    • The InBody device is not a diagnostic device. To make an accurate diagnosis, the physician needs to commission thorough examinations and take their results into account in addition to the results of the InBody.
  • Page 3: Intended User Profile

    3. Contraindication Individuals with medical implant devices, such as pacemakers, or essential support devices, such as patient monitoring systems, must not use this device. Safe, micro alternative currents will flow through the body during the test, which may cause malfunctioning of the device or endanger lives.
  • Page 4 Setup Thank you for purchasing the InBody. This user's manual describes all the features of the InBody. Please read before use and keep it in a safe place. By following the instructions in the manual, you will be able to use the InBody more safely and effectively.
  • Page 5: Table Of Contents

    8 Frequently Asked Questions (FAQ) .25 4 Setting ...........12 Regarding the Device ......25 Regarding Serious Incidents ....26 Initial Setup ..........12 Regarding the InBody Test ....26 IT Security Measures ......14 Residual Risks and Undesirable Side 5 Connecting Compatible Device ..15 Effects ...........27 Printer ...........15...
  • Page 6: Safety

    • Individuals with medical implant devices such as pacemakers, • Repairs and inspections can only be performed by InBody's or essential support devices such as patient monitoring technician. For repairs and inspections, contact the customer systems, must not use this device.
  • Page 7: Device Overview

    • Keep children away from the packing box. • Do not put protective packaging vinyl on your face. • Refer to the description below when transporting InBody.  Number Button: Enter numeric data such as age and height.  Setup Button: Enter the 'Setup' in the Administrator Menu 2.2 Device Components...
  • Page 8  Fingerprint Recognition Sensor: The sensor recognizes the or Lookinbody installed on the PC via a wire connection. fingerprint on the left thumb and saves or loads personal * You can connect the InBody to LookinBody installed on information automatically. the PC if at least one of ports is connected.
  • Page 9: Installation

    3 Installation 3.1 Installation Environment Raise the device upper section to the end and then remove the protective packaging vinyl. Remove the protective packaging material placed over the footplate. Check the environment before installing the device. This device is suitable for indoor use. If installing this device outdoors, the following requirements must be fulfilled.
  • Page 10 Installation Adjust the foot supporters to center the bubble in the Connect the power adapter ( ) to the power inlet on level indicator and to level the device for accurate weight the rear of the device ( ). Connect the power cable ( measurement.
  • Page 11 Installation Press the power switch to turn it on. Warning Power Switch • Do not pull the power cord violently. • Do not plug in or pull out the power cord with wet hands. There is a risk of an electric shock. •...
  • Page 12: Setting

    : This option allows you to change the basic settings of the device. Self Mode/Professional Mode • Self Mode: The subject takes the InBody Test by entering only his/her height. Throughout the test, instructions and the InBody Information will be shown on the screen.
  • Page 13 Once the device is connected to the Internet, the testing a specific age group or gender. InBody Test results can be sent to the cloud, or the LookinBody that can be connected remotely. View/Print/Delete Data...
  • Page 14: Security Measures

    4.2 IT Security Measures Refer to additional information on how to use the InBody. InBody would like to clarify that the user access to the Refer to FAQ checklist when there are problems that occur InBody is only granted for the authorized users, who have while using the device.
  • Page 15: Connecting Compatible Device

    Interpretations for Result Sheet' in Setup in the Administrator Menu and then select blood pressure in the sub-options, then you can print the blood pressure value on the InBody Result The Result Sheet can be printed if the printer is connected to Sheet.
  • Page 16: Lookinbody

    USB-AB type cable to the USB SLAVE port on the rear of InBody device instead, and connect the other end of the USB Enter the password to enter the Administrator Menu. cable to the USB port of the PC.
  • Page 17: Connecting Internet

    * You may need to enter the IP address or DNS address * If the Bluetooth ID does not appear or the connection is manually depending on the service area. In this case, not good, please contact InBody Customer Service. press the corresponding button to use.
  • Page 18 Connecting Compatible Device Connecting Wi-Fi You can connect to LookinBody Web or use the Cloud Service through Wi-Fi after completing internet connection. Press the [Administrator Menu] on the test standby screen. Enter the password to enter the Administrator Menu. Go to '16 Internet Options' in Setup.
  • Page 19: Inbody Test

    Wipe your palms and soles of your feet with an InBody Tissue (wet tissue) before measuring. • When stepping onto or off of the device, do not grab or shove the device.
  • Page 20 Connecting InBodyBAND. If your fingerprint is recognized, the personal information will be automatically displayed on the * If InBodyBAND is connected with iOS, turn off the screen, as shown below. Bluetooth setting and use it. * If fingerprint is placed differently from when it was initially * The InBodyBAND firmware version must be "V038N"...
  • Page 21 * For proper test posture, refer to '6.3 Test Posture'. Press [Exit] to end the measurement. * When the hand and foot electrode is recognized, the InBody Test begins automatically. InBody Test begins. * It will take 90 seconds to complete the test.
  • Page 22: Test Posture

    InBody Test 6.3 Test Posture The subject must maintain proper posture to have accurate test results. Your arms must not touch Keep your arms straight. the sides of your body. The thighs should not touch each other. Hold the hand electrode so that the...
  • Page 23: Maintenance And Storage

    • Turn off the device if you are not using it for a day or longer. • Do not allow any liquid substances to contact the device directly. Wrap the InBody with protective plastic vinyl. Cover the Keep food and drinks away from the device. Substances getting...
  • Page 24: Storage Environment

    Protective Protective packaging material packaging material (Paper) (Paper) Caution • When repacking the device, the protective packaging materials provided by InBody must be used. Caution To transport the device safely, two people should keep the device in a horizontal position.
  • Page 25: Frequently Asked Questions (Faq)

    Insert the power cord completely into the adapter. • The problem may occur if you are using the power adapter that is not provided by InBody. Always connect a power adapter (DC 12 V, 5.0 A/3.34 A) provided by InBody.
  • Page 26: Regarding Serious Incidents

    8.2 Regarding Serious Incidents 8.3 Regarding the InBody Test If you are aware of a serious incident involving your device, The questions and answers regarding the InBody Test are as you must report it as quickly as possible to both the follows.
  • Page 27: Residual Risks And Undesirable Side Effects

    InBody. As part the InBody has already obtained the national and of a comprehensive risk management protocol, the stainless European medical certificate.
  • Page 28: Classifications And Specifications

    Type of the Applied Parts BF Type • Body Composition History (Weight, Skeletal Muscle Mass, Percent Body Fat, ECW Ratio) EMC Emission Class B • InBody Score • Whole Body Phase Angle (History) Degree of Protection IPX0 • SMI (History) against Water Infiltration •...
  • Page 29 Left Arm, Trunk, Right Leg, Left Leg) • Body Water Composition History (Weight, Total Body Water, Intracellular Water, Extracellular Water, ECW Ratio) • InBody Score • Whole Body Phase Angle (History) • SMI (History) • Visceral Fat Area (Graph) • Body Type (Graph) •...
  • Page 30 Classifiaaiiossaons Spiifiaaiios • Body Composition Summary (Fat Free Mass, • Bioelectrical Impedance Vector Analysis Outputs Outputs Fat Mass, Intracellular Water, Extracellular (BIVA) (Research (Evaluation Water, Total Body Water, ECW/TBW: Whole • Whole Body Phase Angle_50 kHz (PhA, Ø) (M Result Sheet) Result Sheet) Body, Right Arm, Left Arm, Trunk, Right Leg, ±...
  • Page 31: Functional Specifications

    Recognition measurer and automatically inputs Function personal information to the InBody Backup Data Backup data saved in InBody by using an InBody USB QR code By scanning QR codes, you can send and verify the InBody results. * QR code is registered trademark of DENSO WAVE...
  • Page 32: Symbols Used On The Device

    Classifiaaiiossaons Spiifiaaiios 9.3 Symbols Used on the Device Disposal of old Electrical & Electronic Devices (Application in the European Union and other European countries with separate collection Indicators system.,) This symbol indicates that this device shall not be treated as household waste. Instead, it shall be handed over to the applicable collection 9-pin Serial Port (Female, RS-232C) point for the recycling of electrical and electronic...
  • Page 33: Guidance And Manufacturer's Declaration

    Classifiaaiiossaons Spiifiaaiios 9.4 Guidance and Manufacturer's Declaration The InBody device is intended for use in the electromagnetic environment specified below. The customer or the user of the InBody device should ensure that it is used in such an environment. Electromagnetic Emissions...
  • Page 34 Recommended separation distances between portable and mobile communication devices and InBody device The InBody device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the InBody device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications devices transmitters) and the InBody device as recommended below, according to the maximum output power of the communications devices.
  • Page 35 Classifiaaiiossaons Spiifiaaiios Electromagnetic Immunity Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance Portable and mobile RF communications devices should not be used closer to any part of Conducted RF the Ultrasound System, including cables, than 150 kHz to 80 MHz the recommended separation distance.
  • Page 36 Electromagnetic Emissions The InBody device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. A portable RF communications devices should be used no closer than 30 cm (12 inches) to any part of the InBody device.
  • Page 37: Key Performance Claims

    InBody. The Body Composition Analyzer of InBody provides clinical benefits to support the aforementioned intended use, as the InBody is mainly used for healthy and acutely or chronically ill populations in hospitals, medical practices and inpatient care facilities in accordance with the national regulations.
  • Page 40 © 2024 InBody Co., Ltd., All rights reserved. IM-ENG-U9-A-240314...