Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Seven Zones 1200° C Max. 5" Split Tube Furnace With Temperature Controller & Quick Flange OTF-1200X-VII-5UL Operation Manual 860 South 19th Street, Richmond, CA 94804, USA Tel: 510-525-3070 Fax: 510-525-4705 Email: info@mtixtl.com...
MTI reserves the right to update or upgrade the product without informing customers of the data change(s) in this manual. Please visit www.mtixtl.com...
Vacuum pressures may only be safely used up to 1500° C For the use with flammable, toxic or corrosive gases, you must contact MTI engineering group for the consultation to know all the prerequisites and precautionary measures ...
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Safety Note Description Do not operate the equipment when objects such as a piece of metal, water or other liquid get inside of the equipment. You should shut off the power first. Unplug the power cord. And ask the professional electrician to take the objects out if possible or contact with us for further assistance.
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 1. Introduction OTF-1200X-VII-5 is a CE certified & UL Ready Seven-zone 5“ split tube furnace with separated temperature control unit and max working temperature of 1200 C . This seven zone furnace can create step thermal gradient or one long constant temperature zone up to 35"...
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 3. Specification 1. Double layer steel case with air cooling to ensure furnace surface temperature < 60° C 2. Energy saving high purity fibrous alumina insulation Furnace structure 3. Heating Element : 1300º C grade Fe-Cr-Al Alloy doped by Mo 4.
Page 8
Warning 3.Vacuum pressures may only be safely used up to 1500° C 4.For the use with flammable, toxic or corrosive gases, you must contact MTI engineering group for the consultation to know all the prerequisites and precautionary measures.
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 4. Installation 4.1 Equipment Inspection Once receiving the furnace, please inspect the equipment and all accessories for damage incurred from shipping and check the packing list for missing items. If anything is damaged or missing, please inform us at info@mtixtl.com...
Page 10
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL MTI Corporation | www.mtixtl.com...
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Attention MTI Corporation will not be responsible for any equipment or property damage that results from improper or incorrect power connections. 4.3 Wire Connection Thermocouple connection: Connect the thermocouples in the sequence; MTI engineers have already labeled the thermocouple connection from 1 to 7.
Page 12
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Terminal Block: MTI engineers have already labeled the wires. Connect to the terminals which are the same as the label. The terminal block delivers the power to the heating elements.
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 5. Operation If the furnace is being used for the first time (or use again after a long period of inactivity), please pre- heat the furnace to 300 for at least 2 hours to remove moisture inside the chamber. This helps to prevent the furnace refractory ceramics from cracking.
Page 14
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 3. Install the vacuum sealing flanges on each side of the tube using the picture below as reference. Seal the flanges by tightening the screws to finish the assembly. 4. Limit the pressure regulator attached to the gas tank to below 1.1 atm before filling the tube with gas.
Page 15
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 5. When the installation is finished, please do a test for gas leaks to check if the sealing parts were assembled well. 6. If your system has a big gas leaking rate, you may locate the leaking part(s) by using soap suds and apply vacuum grease on the part(s).
Page 16
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Attention 1. A small to medium heating rate with a small temperature difference between the adjacent segments should be set when the furnace is being heated up from room temperature. Please take the characteristics of the material being sintered into consideration when setting the heating rate.
Page 17
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 4. Check every parameter before run this program. Press the green button.The main contactor closed with a sound “bang”. Press the RUN button for 2 seconds. The Stop Process 1. After the end of program, press STOP button, the regular shown as the picture.
The flanges on MTI furnaces can be readily configured to accept various fittings for vacuum pump connections. Typically, connection to the pump is made by securing stainless steel bellows to either side of the flange by quick locking clamps.
7.1 PC software Control 1.1 Computer communicating with controller on the furnace. MTI engineers have already assigned the craft 1 to zone 1; craft 2 to zone 2; craft 3 to zone 3 etc. For example, user only need to set up the parameters in the craft 7, the temperature of zone 7 would follow the craft 7 program.
Page 20
1.3 In the parameter setting, user only needs to access “program setting” and “monitoring” in general. MTI engineers have already set up the parameters for each heating zone. 1.4 In the program setting, seven zones are corresponded to seven crafts in sequence and each craft or program has max.
Page 21
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 1.5 In the monitoring, the heating process in each zone can be monitored. MTI Corporation | www.mtixtl.com...
Page 22
Step Present step in the recipe Notice: Since MTI engineer team has already preset the all of the parameters for the PC software and temperature controller, there is a one-to-one correspondence among the heating program, craft and zone. For instance: Program 1 (PC software) – Craft 1 (Controller) – Zone 1 (Furnace).
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 7.2 Quick Temperature Control Setting 2.1 Start up When start the controller, press “ ” to enter MENU display and then press CraftCon, choosing Craft1 or (2.3.4.5.6.7.8.9.10) to enter temperature control setting.
Page 24
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL MTI’s Engineer has set up a quick heating program for testing the furnace of the curve below. Please follow these steps in order to run the test program: Prompt Input Data...
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 7.3 Interface and operation Current System Locked.(red:locked. white:unlocked) screen information Time. Modified in“system setting” Current channel Unit. Modified in“channel ” Measurement value(PV) Bit number. Modified in“channel” click the area can quickly enter the settings Input specifications.
Page 26
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL General curve information shown in fig7.2. Each screen display real-time curve and output in bar-graph of four channels, comparing the measured values and adjusting the output trends. If channels are more than 4, press the "PAGE"...
Page 27
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL General information of four channels shown in fig7.6. Each screen display 4 channels, press“CHNU”to page. The six circles at the bottom indicate six alarm status, from left to right are: HA-upper limit alarm,LA-lower limit alarm,DH-positive bias alarm, DL-negative...
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Process curve shown in fig7.10. Each screen can display 5 or 30 segment process curves and running track, parameter setting in“Display”.It is convenient to view the process and running status.
Page 29
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Chunit:set measurement unit of current channel. Chsymbol:set bit number of current channel, can not exceed eight characters. Cont: set regulation mode value meaning symbol remarks Bit regulation or Current AT_NOT...
Page 30
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL This parameter also can judge the self-tuning. If its value is modified automatically after self-tuning is done, that indicates self-tuning fails. You need to restart self-tuning. Input : Table7.1 value...
Page 31
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL B) To ascertain least value when linear current regulation (transmitter) mode. If ”output type” is set at 1, unit would be 0. 1mA, viz value × 0. 1=minimum practical current.
Page 32
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Besides the mode 0 mentioned above, mode 4 has additional function that can run from current measurement; the measurement usually is not the same with the given value when program running or is modified. Take the temperature control as...
Page 33
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL HdAL(Positive deviation alarm):To set the positive deviation (PD). E.g: we need a alarm signal when measuring value(MV) 10℃ higher than setting value(SV), and we set PD at 10. So if SV is 500, we get a signal when MV≥510℃.
Page 34
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 4.3 Permis(MUNU→Permis) level:there are two levels provided in“level”: “Engner”and“Operat”. “Engner”engineer can set and modify all items; “Operat”operator can set the given value, first time self-tuning, manual/automatic conversion; Code:input a new password and press“ ”to quit and make it become effective.
Page 35
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Facnumb:Instrument number. The user can not change; Version:software version; Challnum:total number of channels. The user can not change; Probatin:system reset 4.5 Record(MENU→Record) Recintvl: Time format is Hours: Minutes: Seconds; Recmode:select the recording mode.
Page 37
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL A is to control two event status: A=0,No effect ( only jump) A=1,trigger event 1; A=2,trigger event 2; A=3,trigger event 1 and 2; A=5,event 1 trigger is released; A=4,stop(shut all output);...
Page 38
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Trigger event Release event 1 Natural cooling thermostatic thermostatic heating thermostatic cooling Natural cooling heating 4.8.5 Import files by U disk BT1009 color programmable regulator allow users to import and export files.
Page 39
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL Insert SD card or U disk will pop out the dialog. There are 3 choices: Craftinp:import files on U disk to meter,including process 1~process 10(selected in the dialog box); Craftout:export files to U disk or SD card;...
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 7.5 Instructions for data management software Copy record data to U disk Insert the supplied SD card or U disk to the socket in the lower right corner of the meter.
7. Do not open the furnace when it is being heated above 300° C to prevent its insulation materials from cracking. 8. MTI furnaces are only suitable for using with air or inert gas environments. Other flammable or toxic gases, such as H...
Seven Zones 1200° C Max. 5" Split Tube Furnace OTF-1200X-VII-5UL 9. Trouble Shooting 1. Furnace is unable to start up If the furnace cannot be started up when you press the power button: ① Make certain that the furnace is plugged into an AC outlet with adequate power.