Page 1
All manuals and user guides at all-guides.com Operating Instructions Laboratory Glassware Washer G 7883 To prevent accidents and machine damage read these instructions before installation or use. M.-Nr. 05 893 860...
All manuals and user guides at all-guides.com IMPORTANT SAFETY INSTRUCTIONS Installation and Service This appliance is only intended for The machine should only be specialized applications. Only use installed, maintained and repaired the appliance for its intended by a Miele authorized service purpose.
Page 4
All manuals and user guides at all-guides.com IMPORTANT SAFETY INSTRUCTIONS Avoid inhaling powdered detergent. If swallowed, it can Personnel operating the machine cause burning in the mouth and throat should be trained regularly. or inhibit breathing. Children and untrained personnel must not be allowed access to the machine Be careful when sorting items with or its controls.
Page 5
All manuals and user guides at all-guides.com IMPORTANT SAFETY INSTRUCTIONS Pre-treatments with cleaning or The machine is designed only for disinfecting agents can create operation with water and additive suds, as can certain rinsing agents. For cleaning agents. Organic solvents must pre-treatment and / or cleaning, only not be used in the machine.
Page 6
All manuals and user guides at all-guides.com IMPORTANT SAFETY INSTRUCTIONS Accessories Only genuine Miele parts or accessories should be used with the Laboratory Glassware washer. The performance and safety of non-genuine parts or accessories cannot be guaranteed, and use of such parts or items may void the machine warranty.
All manuals and user guides at all-guides.com Description of the machine The G 7883 is the "industry standard" undercounter washer designed to handle the vast majority of glassware cleaning applications.
All manuals and user guides at all-guides.com Guide to the lab washer a "On/Off" button j "Add liquid detergent" indicator * b "Door" button k "Add rinse aid" indicator c Display l "Add neutralizing agent" indicator d "Display" indicator m "Recharge Softener" indicator e "Drying"...
Page 9
All manuals and user guides at all-guides.com Guide to the lab washer a Connection for optional DOS-Module d Powdered detergent dispenser b Filter combination e Rinse aid reservoir (with dosage selector) c Salt reservoir f Rinse aid level indicators (water softener)
All manuals and user guides at all-guides.com Opening and closing the door Electric door lock To cancel a program The machine is equipped with an If a program has to be canceled in an electric door lock. The door can only be emergency, e.g.
All manuals and user guides at all-guides.com Opening and closing the door To open the door with the emergency release The emergency release should only be used when the door cannot be opened normally, e.g. in the event of a power failure.
All manuals and user guides at all-guides.com Water softener Before using the machine for the first Your local water authority will advise time, the water softener should be you on the water hardness in your programmed. The water needs to be softened to avoid calcium deposit area.
Page 13
All manuals and user guides at all-guides.com Water softener Settings Settings gr/gal Settings gr/gal 1 - 30 1 - 30 CaCO 31 - 60 31 - 60 CaCO 19 *) 1010 1030 1040 1060 1080 *) factory setting...
All manuals and user guides at all-guides.com Water softener Setting the water softener The setting will now be stored in memory. The display will go out. ^ Turn off the machine with the "On/Off" button. For future servicing, it is useful to make a note of your water hardness level.
All manuals and user guides at all-guides.com Water softener Filling the salt reservoir Only special, coarse-grained reactivation salt* should be used in this machine. *Available from Miele’s Professional Customer Service Department. Please see back page for more information. Do not use other types of salt, e.g. table salt, agricultural or gritting salt.
All manuals and user guides at all-guides.com Water softener Salt indicator When the salt reservoir is empty the "Recharge Softener" indicator, b, illuminates to remind you to fill the reservoir. Reactivation takes place automatically during a program. The "Recharging" indicator, a, illuminates while this is occurring.
All manuals and user guides at all-guides.com Adding rinse aid Rinse aid is not used in laboratory glassware cleaning applications. If rinse aid is required, the rinse aid dosage must be activated with assistance from Miele Technical Service. Rinse aid (e.g. Mielclear) is filled in the storage container and dispensed automatically.
All manuals and user guides at all-guides.com Adding rinse aid Setting the rinse aid dosage The dosage adjuster (see arrow in the illustration) in the opening has settings adjustable from 1 to 6 (1 - 6 ml). It is preset to "3" (3 ml). ^ Use a flat headed screwdriver to turn the dial to the correct setting.
All manuals and user guides at all-guides.com Adding neutralizing agent Neutralizer (pH-level: acidic) neutralizes detergent residue on the surface of glassware and other items being cleaned. It is automatically dispensed in the rinse following the wash step of a program. The relevant program (see "Program Guide"...
Page 20
All manuals and user guides at all-guides.com Adding neutralizing agent Setting the neutralizer dosage The dosage is factory set to 0.10, you may or may not be required to adjust it. and priming the neutralizer dispensing system Set the dosage concentration First you need to access the according to the detergent programming mode:...
Page 21
All manuals and user guides at all-guides.com Adding neutralizing agent Priming the neutralizing system ^ Access the programming mode level 2 (see "Adding neutralizer - To access the programming mode"). ^ Turn the program selector to "Universal". "do3" appears in the display. ^ Press the "Display"...
All manuals and user guides at all-guides.com Adding detergent Maintenance of the dispensing Only use cleaning agents formulated system for this machine. Do not use To ensure trouble-free operation, the detergents formulated for following regular maintenance should domestic dishwashers. be carried out by a Miele approved service technician.
All manuals and user guides at all-guides.com Adding detergent Powdered detergent If you are using powdered detergent, it must be placed in the detergent container prior to every program, except for "Rinse" and "Drain". ^ Add detergent to the compartment. Dosage: approx.
All manuals and user guides at all-guides.com Loading The following instructions relate only to Note basic preparation and loading of ^ Load items so that water will cover all glassware. There are additional surfaces. This ensures that they will requirements for particular applications. be properly cleaned.
Page 25
All manuals and user guides at all-guides.com Loading Important: Make sure that the spring adapter for the water connection engages correctly when a basket or injector unit is inserted into the machine. It must be 4 - 5 mm higher than the water connection inlet in the machine.
All manuals and user guides at all-guides.com Loading Loading examples O 188 Upper basket O 187 Injector wagon For various inserts. (Upper basket) For direct injection of narrow necked glassware, 34 injectors,160 mm high. E 350 Injector wagon For narrow necked glassware, 15 injectors, 160 mm high;...
Page 27
All manuals and user guides at all-guides.com Loading U 874 Lower basket E 106 For various inserts. Stainless steel half insert with 28 spring hooks in 2 different heights for various glassware, e.g. narrow necked flasks, graduated cylinders, medicine bottles, etc.
All manuals and user guides at all-guides.com Loading Adjusting the upper basket The upper basket can be adjusted above and below the middle position by 2 cm. Example: Upper basket O 188/1 and lower basket U 874 (measurements do not include inserts) Upper Upper basket Lower basket...
All manuals and user guides at all-guides.com Program guide Program Program cycle Pre- Main Main Rinse Rinse Final Final Drying Wash wash 1 wash 2 rinse 1 rinse 2 (Optional 1) 2) 1) 2) function) Custom DOS 1 DOS 3 75/1’...
Page 30
All manuals and user guides at all-guides.com Program guide Program Program cycle Pre-Wash Thermal Rinse Rinse Final rinse Drying Disinfection (Optional 1) 2) and/or main function) wash varioTD DOS 1 DOS 3 93°C/10’ 55°C/3’ 1’ SPECIAL DOS 1 DOS 3 93°C-10’...
All manuals and user guides at all-guides.com Operation Once the program has started, all other Turning on programs are blocked. If the program ^ Close the door. selector is turned to another program, ^ Press the "On/Off" button. the values shown in the display will go out.
All manuals and user guides at all-guides.com Operation Drying At the end of a program "Drying" 0 may be selected as an ^ The "Complete" indicator in the additional feature once a program has program sequence display illuminates, and the "Start" 6 been selected (except in the "Rinse"...
All manuals and user guides at all-guides.com Operation Canceling a program ^ Turn the program selector to the "Stop" f position (the program will stop after approximately 2 seconds). ^ Open the door. Caution: Water and items in the machine may be hot. Danger of burning or scalding.
All manuals and user guides at all-guides.com Programming special functions Accessing the programming For future servicing, please level document any alterations to the Before performing any of the special standard factory settings. functions that follow, the programming mode of the labwasher controller needs General note to be accessed.
Page 35
All manuals and user guides at all-guides.com Programming special functions ^ Turn the program selector to 1. Setting the dosage "Inorganica". The display shows concentration and priming the either the dosage time in seconds dosage system (for powdered detergent) or the set dosage concentration in % (for liquid Before performing any of the detergent).
Page 36
All manuals and user guides at all-guides.com Programming special functions 1.1.2 Priming the detergent system b) If the system is primed only (e.g. because the dosage system was a) If the dosage concentration is set: completely empty): ^ Turn the program selector to ^ Select programming level "E02"...
Page 37
All manuals and user guides at all-guides.com Programming special functions 1.2.2 Priming the neutralizer system Before performing any of the following special functions, the a) If the dosage concentration is set: programming mode of the controller ^ Turn the program selector to needs to be accessed.
Page 38
All manuals and user guides at all-guides.com Programming special functions ^ Turn the program selector to "STOP" 2. Changes in wash temperature and / or temperature holding time at "Main wash" (except ^ Turn the machine off. SPECIAL 93°C-10’ and "Final ^ Press and hold "Display"...
Page 39
All manuals and user guides at all-guides.com Programming special functions "Main wash" temperature "Main wash" temperature holding time ^ Main wash 1: Press and hold the "Drying" 0 ^ Turn the program selector to "STOP" button until "E09" appears in the display.
Page 40
All manuals and user guides at all-guides.com Programming special functions "Final rinse" temperature "Final rinse" temperature holding time ^ Turn the program selector to "STOP" ^ Turn the program selector to "Stop" ^ Final rinse 1: Press and hold "Drying" 0 until "E22" ^ Final rinse 1: Press and hold "Drying"...
Page 41
All manuals and user guides at all-guides.com Programming special functions ^ Turn the program selector to 3. Setting the buzzer "Inorganica". A number between "20" A buzzer can be set to signal the and "27" appears in the display. following: "20"...
All manuals and user guides at all-guides.com Programming special functions 4. Restoring the factory If the hardness level setting in the settings water softener had been changed, ^ Turn the program selector to "STOP" this will reset to the factory setting of 340 ppm (19 gr/gal).
All manuals and user guides at all-guides.com Machine care Filter combination Cleaning the fine, flat and micro-fine filters The machine must not be used ^ Remove the coarse filter. without all the filters in place. The ^ Remove the fine filter (if fitted) from be- filters protect the circulation pump from damage caused by foreign tween the flat and the micro-fine filters.
All manuals and user guides at all-guides.com Machine care Cleaning the non-return valve The drain pump is located beneath the non-return valve (see arrow). and drain pump If water has not been pumped away at the end of a program, the drain pump or the non-return valve may be blocked.
All manuals and user guides at all-guides.com Machine care ^ Replace filters and the seal. Make Cleaning the filters in the water sure they are sitting correctly. inlet ^ Reconnect the hose to the water inlet, Filters are incorporated in the screw making sure that it goes back on connection of the water inlet hose to straight.
All manuals and user guides at all-guides.com Trouble shooting With the aid of the following guide, minor problems can be fixed without a service call. The Miele control unit includes a wide range of fault codes to indicate specific errors. These fault codes are indicated on the display by a flashing F. . . (Ranging from F1 up to F30.) If one of these codes appears, please contact the Miele Technical Service Department at 1-800-991-9380 for assistance.
Page 47
All manuals and user guides at all-guides.com Trouble shooting What if … Possible fault the wash cycle finishes – The drain hose is – Turn the program selector to "Stop" f (the fault code too early, the "Fill/Drain" kinked. goes out). indicator flashes and –...
Page 48
All manuals and user guides at all-guides.com Trouble shooting What if … Possible fault I need to open the door – see "Opening and during a power failure? closing the door." the fault was caused by – Turn the program selector to "Stop"...
All manuals and user guides at all-guides.com After Sales Service In the event of a fault which you cannot correct yourself please contact the Miele Technical Service Department 1-800-991-9380 techserv@mieleusa.com 1-800-565-6435 service@miele.ca Please quote the model of your machine. This information can be found on the machine’s front panel.
Page 50
All manuals and user guides at all-guides.com...
All manuals and user guides at all-guides.com Caring for the environment Disposal of the packing Disposal of the appliance material Appliances contain materials which can be recycled. Please contact your local The protective packing materials are authorities about recycling in your area. environmentally safe and can be recycled.
All manuals and user guides at all-guides.com Installation Positioning and securing the Please refer to the installation diagram supplied with the machine. machine To ensure stability, the machine must be Furniture and fittings installed aligned and screwed to the counter. near the machine must be of a commercial standard (able to ^ Open the door.
Page 54
All manuals and user guides at all-guides.com Installation Removing the lid (if necessary) Important for machines with steam condenser (depending on model) If necessary, remove the lid from the ^ To prevent countertop damage by machine to make installation possible. steam, the supplied self adhesive ^ Open the door.
All manuals and user guides at all-guides.com Plumbing ^ Water valves with Water inlet " male hose thread are to be provided on site. ^ The machine must be connected to They should be easily accessible so the water supply in accordance with the water supply can be turned off if local and national regulations.
Page 56
All manuals and user guides at all-guides.com Plumbing DI-Water connection The machine comes with a standard connection for a pressure-resistant system. The recommended water pressure (flow rate) is 25 - 60 psi however, water pressure of 10 - 147 psi is acceptable.
Page 57
All manuals and user guides at all-guides.com Plumbing Drainage ^ The drainage system is fitted with a non-return valve which prevents dirty water from flowing back into the machine via the drain hose. ^ The machine should preferably be connected to a separate drainage system onsite.
All manuals and user guides at all-guides.com Electrical connection It can be converted to use a single All electrical work must be phase power supply in accordance performed by a qualified person in with the conversion and wiring diagram. accordance with local and national The cord must be changed from a safety regulations.
All manuals and user guides at all-guides.com Electrical connection Grounding connection Removing the service panel ^ The ground lead must be connected Disconnect the machine from the to the screw connection point main electrical supply. (marked with the ground symbol 8) at the back of the machine.
All manuals and user guides at all-guides.com Technical data English Metric Height: " (32 ") 85 (82) cm Width: " 60 cm Depth: " 60 cm Voltage: 208 V Power cord approx. 5’ 11" 1.8 m Water pressure (flow rate): Recommended 25 - 60 psi 1.72 - 4.13 bar...
Page 61
All manuals and user guides at all-guides.com...
Page 62
All manuals and user guides at all-guides.com...
Page 63
All manuals and user guides at all-guides.com...
Page 64
All manuals and user guides at all-guides.com All rights reserved / 1104 M.-Nr. 05 893 860 / 01 M.-Nr. 05 893 860 / 01 This paper has been bleached without the use of chlorine.
Need help?
Do you have a question about the G 7883 and is the answer not in the manual?
Questions and answers