Model DS1200-AMT-Ex
9
Specifica
9.1
Model DS1200 Specifica
•
Temperature opera
•
Power supply = 20 – 300 VDC, or 22 to 253 VAC 50 – 60 Hz
•
3.2W internal power consump
•
3.5W maximum power consump
•
Isola
on voltage test / opera
•
Signal to noise ra
•
Display Response
•
Wire size, pin 45 & 46 (max) = 1 x 1.5mm2 stranded wire
•
Wire size, others (max) = 1 x 2.5mm2 stranded wire
•
Rela
ve Humidity = < 95% RH
•
Dimensions (HxWxD) = 48 x 96 x 120mm
•
Cut out dimensions = 44.5 x 91.5mm
•
Ingress Protec
•
Weight = 230g
•
Input
o
From Model AMT-Ex
o
Error detec
•
Display
o
Display Readout = AMT range
o
Scrolled error display
•
Current Output
o
Programmable signal ranges = 0...20 or 4...20mADC (Factory set)
o
Load (max) 20mA / 800
o
Load stability ≤ 0.01 % of span / 100
•
Relay
o
Relay func
o
Hysterisis 1% of range
o
Max voltage = 250 VRMS
o
Max Current = 2A /AC
o
Max AC power = 500VA
o
Max current at 24 VDC = 1A
9.2
Model AMT-Ex Specifica
Output Signal
Opera
ng Voltage
Maximum Series Resistance
Sensing element
AutoCal
Factory calibra
Accuracy
2148 - Iss 6 - 10/23
ons
ons:
ng range = -20°C to +60°C
o = Min 60dB (0-100kHz)
me (0-90%, 100 to 10%) = 0.4s.
on = (Front panel) IP65
on = Cable loop break
on
= setpoint
on
on
on
on
on = 2.3KVAC / 250 VAC
/ 16VDC
:
4 to 20mA Linear
:
18V - 28V DC. Reverse polarity
protected.
:
= {40 x (Supply Voltage – 7)}
:
Ultra High Capacitance - Aluminium
Oxide Type
:
Field calibra
:
Supplied with Cer
traceable to NPL / NIST
:
± 2°C dewpoint (NPL / NIST traceable
for range -90°C to 20°C)
on / Span check facility.
ficate of Calibra
on
19
Need help?
Do you have a question about the DS1200 and is the answer not in the manual?
Questions and answers