4.Product operation is carried out according to the instruction manual. Disclaimer Biobase shall not be liable for any equipment failure or damage, or for any direct or indirect damage that may occur during the use of the equipment. 1.Malfunction or damage due to violation of the instructions, precautions, and intended use of this manual.
1.Introduction 1.1 Attention This operation manual contains necessary and sufficient information to operate the sterilizer safely, like optimal usage, safe and reliable operation, regular and correct servicing requirements. Read and understand all instructions in this manual before attempting to use the product. ...
2.Safety 2.1 Explanation Symbol Device Icon Description “ATTENTION” – Intended to alert the user to refer to the operation manual or other instructions when complete information cannot be provided on the label. “ATTENTION” – Pay attention to the high temperature in the chamber, and the sterilizer exterior when exhausting system is running.
Manual Prompt Note Indicates that concerning information which easier or helpful in the operation. Indicates that a potential hazard may exist which through inappropriate conditions or actions will or can cause: Caution • Minor injury; • Property damage; • Damage to machine. Indicates that a specific hazard may exist which through inappropriate conditions or actions will or can cause: Warning...
Scald risk Every time open the door after the sterilizer finishes the cycle, please keep an appropriate distance, because the chamber still has rudimental steam with high temperature, avoid to be scalded. Every time open the door after the sterilizer finishes the cycle, please do not touch the main door and chamber, because of high temperature, avoid to be scalded.
3.Receiving and Installation 3.1 Check the Package Please check package carefully when you receive the equipment. Front Flank Item 18 L 24 L Packing Size 660x525x495 mm 770x530x495mm N./G. Weight 25/28 KGs 30/33 KGs 3.2 Unpack the Accessories Open the door and take out all the accessories, item listed below: Part Name Tray rack 1 piece...
Page 9
The sterilizer should be set in a place with good ventilation; The temperature environment: 5-40℃; The humidity of environment: ≤85%; Atmosphere pressure: 860Hpa ~ 1060Hpa; An earth connection is essential.
4.Description and Specification 4.1 Overview NO. Item Description Door Handle Open or close the door Function Panel Show sterilization information for control and operation Leak pressure automatically when over working Safety Valve pressure Output heat and wasted water (Attention! High Discharge Port temperature!) Power Cable...
5.Panel and Function 5.1 Function Panel Display Description 1.Press. Real time in-chamber pressure 2.Temp. Real time in-chamber temperature and error code 3.Start lamp Flash when the program starts 4.Sterilize lamp Flash in the process of sterilization 5.End lamp Lighted up when the program ends 6.121℃...
Page 12
3、 Open the door and fill 1000ml distilled water into the chamber until the water immerse the heating stick. 4、Put the instruments which need to be sterilized in the chamber, ①Turn on the power switch; ②Turn the handle clockwise until it is tightened. ②...
① ③ ② 6、Light of “END” will be lighted up when the program ends, when it release automatically to the safe pressure below 0.1 bar, ①Manually press the release button to open, or rotate the handle to the door-opened position, the program will automatically open the exhaust valve;...
5.4 Sterilization Program Cycle time Default drying time Program Sterilizing time (without drying) (if request) 121℃ 20min approx.35min +20min 134℃ 5min approx.25min +20min When the program is running and the display window shows(134℃): :means alternate display Preheating 1st pressurize 1st Exhaust 2nd pressurize 2nd Exhaust Pr.0...
6.Error Code List The sterilizer will show Error information when malfunctions happen. Example Error Code: E01 Error Code List Error Description Code Er01 Sensor of the chamber defective Er02 In-chamber sensor defective Er03 Sterilization fail (high temperature) Er04 Sterilization fail (low temperature) Er05 Chamber over temperature Er06...
Appendix Piping Diagram Code Name Code Name Safety Valve Drain valve (Normal-Close) Pressure Exhausting Port Pressure sensor Air Release Valve Temperature Sensor(PT1000) (Normal-Close) OWV One-way Valve Air Pump Heating Tube Air Filter...
Page 18
1. Preparation before Operation: Add pure water into chamber in accordance with the requirements, and connect the power supply for this equipment. Connect the control-purpose power supply, turn the power switch of sterilizer to “|” side, and make proper preparation for operation of the program. ...