Special Product Features - CAIRE FreeStyle Comfort User Manual

Portable oxygen concentrator with autosat, ultrasense and autodose
Hide thumbs Also See for FreeStyle Comfort:
Table of Contents

Advertisement

FreeStyle Comfort
A brief audible alarm will sound to indicate the
device has powered on.
Note: When the FreeStyle Comfort is initially turned
on the default flow setting is 2. If the setting is
changed, the unit will power on to the previous flow
setting.
6. Position the nasal cannula on your face per the
manufacturer's instructions provided with the
cannula and breathe normally. You should feel a
puff of oxygen delivered through the nasal cannu-
la each time you inhale.
7. To change the flow setting, press the increase (+)
or decrease (-) flow buttons to the desired flow
rate. It is normal to hear a difference in sound as
you change the flow settings.
autoSAT Pulse Volume (mL)*Setting
Breath
1
2
Rate
15
10.5
21.0
20
10.5
21.0
25
10.5
21.0
30
10.5
21.0
35
10.5
21.0
40
10.5
21.0
*Tolerance is +/- 15%
14 - EN PN MN232-C4 L | User Manual
3
4
5
31.5
42.0
52.5
31.5
42.0
52.5
31.5
42.0
42.0
31.5
35.0
35.0
30.0
30.0
30.0
26.3
26.3
26.3
non-autoSAT Pulse Volume (mL)* Setting
Breath
1
2
Rate
15
14.0
28.0
20
10.5
21.0
25
8.4
16.8
30
7.0
14.0
35
6.0
12.0
40
5.3
10.5
8. Turn the FreeStyle Comfort off by pressing the
power button.
Note: The display screen will enter standby mode and
dim after 20 seconds. Press any button to wake the
screen out of standby mode.
9. For Advanced Features see Menu Navigation
Instructions

SPECIAL PRODUCT FEATURES

autoSAT
®
CAIRE's proprietary autoSAT feature allows the
FreeStyle Comfort to deliver a consistent volume
of oxygen (bolus) per breath, up to a maximum
breath rate. As the user's breath rate increases
during activities like walking, the autoSAT feature
adjusts the device operation to maintain a set bolus
volume.
UltraSense
®
This sensitive pulse dose trigger detects the
pressure change during your breath, and releases
the pulse of oxygen when you inhale.
autoDOSE
®
Should a "no breath alarm" occur, this feature will
pulse oxygen at 20 breaths per minute at the unit's
current flow setting until the alarm state is resolved.
3
4
5
42.0
56.0
70.0
31.5
42.0
52.5
25.2
33.6
42.0
21.0
28.0
35.0
18.0
24.0
30.0
15.8
21.0
26.3

Advertisement