Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741
7 Processing steps in accordance with ISO 17664 | 7.3 Disinfection
7.2.4 Automated internal cleaning
KaVo recommends washer disinfectors in compliance with ISO 15883-1, which are
operated using alkaline cleaning agents.
▶ For program settings as well as cleansers and disinfectants to be used, please re-
fer to the Instructions for Use of the washer disinfector.
▶ In order to prevent negative effects on the KaVo medical device, make sure that
the interior and the exterior of the medical device are dry after completion of the
cycle, and then lubricate it immediately with care products from the KaVo care
system.
7.3 Disinfection
Incomplete disinfection.
Infection hazard.
▶ Principally, KaVo recommends carrying out an final disinfection of the unpack-
aged item to guarantee the complete disinfection.
Using the disinfectant bath or chlorine-containing disinfectants.
Functional damage and property damage.
▶ Do not disinfect the device in the disinfection bath or with chlorine-containing dis-
infectants.
7.3.1 Manual external disinfection
KaVo recommends the following products based on material compatibility. The micro-
biological efficacy must be ensured by the disinfectant manufacturer and proven by
an expert opinion.
▪ CaviCide made by Metrex
Consumables required:
▪ Cloths for wiping the medical device.
▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow
the disinfectant to act in accordance with the instructions of the
disinfectant manufacturer.
Note
Comply with the instructions for use of the disinfectant.
7.3.2 Manual internal disinfection
Not applicable.
This product is suitable for automated disinfection only.
7.3.3 Automated external and internal disinfection
▶ For program settings as well as cleansers and disinfectants to be used, please re-
fer to the Instructions for Use of the washer disinfector.
WARNING
NOTICE
18 / 26