Instructions for use
INTRA EVA head L61 G - 1.008.1828
INTRA EVA head L61 R - 1.008.1829

Advertisement

loading
Need help?

Need help?

Do you have a question about the INTRA EVA head L61 G and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo INTRA EVA head L61 G

  • Page 1 Instructions for use INTRA EVA head L61 G - 1.008.1828 INTRA EVA head L61 R - 1.008.1829...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................7 2.4 Accessories and combination with other equipment ............
  • Page 4 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
  • Page 5 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 1 User instructions Serial number Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices...
  • Page 6 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 1 User instructions WARNING In cases which – if not prevented – can lead to death or severe injury. CAUTION In cases which – if not prevented – can lead to minor or moderate injury.
  • Page 7 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
  • Page 8 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 2 Safety | 2.4 Accessories and combination with other equipment 4 If the following defects occur, stop working and have the service personnel carry out repair work: ▪...
  • Page 9 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 2 Safety | 2.6 Service and repair ▪ Service technicians of KaVo authorised dealers after the appropriate product training Observe all the following items during servicing work: 4 Have the service and testing tasks carried out according to the Medical Device Operator Ordinance.
  • Page 10 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 3 Product description | 3.1 Intended use 3 Product description INTRA EVA head L61 G (Mat. no. 1.008.1828) INTRA EVA head L61 R (Mat. no. 1.008.1829) 3.1 Intended use...
  • Page 11 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 3 Product description | 3.2 Technical Specifications L61 G NOTE Read and take note of Instructions for use of the KaVo shanks in which the head can be inserted.
  • Page 12 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 3 Product description | 3.4 Transportation and storage conditions 3.4 Transportation and storage conditions CAUTION Startup after refrigerated storage. Malfunction. 4 Prior to startup, strongly refrigerated products must be allowed to warm up to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
  • Page 13 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 4 Startup and shut down | 4 Startup and shut down WARNING Hazard from contaminated products. Patients, users or third parties could be infected by contaminated medical devices.
  • Page 14 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 4 Startup and shut down | 4.1 Checking the amount of water 4.1 Checking the amount of water CAUTION Overheating of the tooth due to insufficient amount of cooling water.
  • Page 15 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING The patient swallows or aspirates loose parts or substances. Danger of suffocation.
  • Page 16 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 5 Operation | 5.3 Inserting the dental tool in the head 5.3 Inserting the dental tool in the head NOTE Files from the firm Intensiv can be used.
  • Page 17 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 5 Operation | 5.4 Removing the dental tool 5.4 Removing the dental tool WARNING Axially moved files. Cuts and infections. 4 Do not touch file when moving in axial direction.
  • Page 18 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 5 Operation | 5.5 Application of head ▪ The removal of proximal fillings surfaces and crown edges with the Prox- oshape® file. ▪ With a Rootshape file for smoothing roots and odontoplasty.
  • Page 19 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions CAUTION Overheating of the device.
  • Page 20 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Press the O-ring between your fingers to form a loop. 4 Push the O-ring to the front, and remove it.
  • Page 21 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 7.1 Preparations at the site of use...
  • Page 22 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual reprocessing 7.2 Manual reprocessing WARNING Incomplete disinfection. Patients, users or third parties could be infected by contaminated medical devices.
  • Page 23 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual reprocessing 7.2.2 Manual internal and external cleaning and internal...
  • Page 24 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.3 Automated reprocessing 7.3 Automated reprocessing WARNING Sharp dental tool in the medical device.
  • Page 25 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing 7.3.3 Automated drying The drying procedure is usually part of the cleaning programme of the washer disinfector.
  • Page 26 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.5 Packaging 7.4.2 Servicing with KaVo QUATTROcare PLUS NOTE Head can be serviced alone (with head adapter) or attached to a shank.
  • Page 27 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.6 Sterilisation 7.6 Sterilisation Sterilisation in a steam steriliser (autoclave) in...
  • Page 28 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat.No. INTRA Instrument stand 3.005.5204 Service coupling for heads (QUATTROcare) 0.411.7941...
  • Page 29 Instructions for use INTRA EVA head L61 G - 1.008.1828, INTRA EVA head L61 R - 1.008.1829 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions:...