Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................7 2.4 Accessories and combination with other equipment ............
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
Page 5
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 1 User instructions Serial number Manufacturer UDI symbol Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices...
Page 6
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 1 User instructions WARNING In cases which – if not prevented – can lead to death or severe injury. CAUTION In cases which – if not prevented – can lead to minor or moderate injury.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 2 Safety | 2.4 Accessories and combination with other equipment 4 If the following defects occur, stop working and have the service personnel carry out repair work: ▪...
Page 9
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 2 Safety | 2.6 Service and repair ▪ Service technicians of KaVo authorised dealers after the appropriate product training Observe all the following items during servicing work: 4 Have the service and testing tasks carried out according to the Medical Device Operator Ordinance.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 3 Product description | 3 Product description INTRA Endo head L3 (Mat. no. 1.008.1822) INTRA Endo head L3 Y (Mat. no. 1.008.1823) 10 / 28...
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 3 Product description | 3.1 Intended use 3.1 Intended use Indications for use: The medical device is: ▪ Intended for dental treatment only. All other types of use of or modifications to the product are not permitted and can be hazardous.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 3 Product description | 3.2 Technical Specifications L3 3.2 Technical Specifications L3 Drive speed max. 3,000 rpm Identification yellow Tool movement is converted into an oscillating motion of approx.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 4 Startup and shut down 4 Startup and shut down WARNING Hazard from contaminated products. Patients, users or third parties could be infected by contaminated medical devices.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating dental tool. Cuts, infection and burn injury. 4 Never push the press-button while the dental tool is rotating.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 5 Operation | 5.3 Inserting the dental bur 5.3 Inserting the dental bur NOTE Only use root canal instruments with handles that comply with DIN EN ISO 3630-1: - Handle diameter: 3.5 to 4.0 mm...
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 5 Operation | 5.4 Removing the dental bur CAUTION Dental tool shaft slips inside the chucking system due to excessive speed of the dental tool or abrupt engagement of the dental tool.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions CAUTION Overheating of the device.
Page 18
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Press the O-ring between your fingers to form a loop. 4 Push the O-ring to the front, and remove it.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 7.1 Preparations at the site of use...
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual reprocessing 7.2 Manual reprocessing WARNING Incomplete disinfection. Patients, users or third parties could be infected by contaminated medical devices.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.3 Automated reprocessing ▪ Brush, e.g. medium-hard toothbrush 4 Brush off under flowing tap water at least 10 seconds.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing CAUTION Never reprocess this medical device in an ultrasonic cleaner.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing CAUTION Improper service and care.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.5 Packaging 4 Close the front flap and press the chuck service button for at least three seconds until the spray canister control LED flashes three times consecut- ively.
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.7 Storage – at least 10 minutes at 134 C -1 C / +4 C (273 F -1.6...
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat.No. INTRA Instrument stand 3.005.5204 Service coupling for heads (QUATTROcare) 0.411.7941...
Instructions for use INTRA Endo head L3 – 1.008.1822, INTRA Endo head L3 Y – 1.008.1823 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...