2
Testing Performance
Test External Trigger Sensitivity
This test applies to all models.
Verify the external trigger sensitivity at these settings:
1 Connect the equipment (see
38
100 MHz, 200 mV
pp
200 MHz, 350 mV
pp
a Use the N cable to connect the signal generator to the power splitter input.
b Connect one output of the power splitter to the Aux Trig input through a 50Ω
feedthrough termination.
c Connect the power sensor to the other output of the power splitter.
Figure
7).
M9241/42/43A PXIe Oscilloscopes Service Guide