Coagulation Analyzer:
Yumizen G400 / Yumizen G400 DDi
Technical Manual
Ref: 1300047286
Int. Ref. Doc.: RAA068AEN

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Yumizen G400 and is the answer not in the manual?

Questions and answers

Summary of Contents for horiba Yumizen G400

  • Page 1 Coagulation Analyzer: Yumizen G400 / Yumizen G400 DDi Technical Manual Ref: 1300047286 Int. Ref. Doc.: RAA068AEN...
  • Page 3: Table Of Contents

    Temperature measurement..........11 ................ 11 1. Problem 1 ................ 11 2. Problem 2 Incubator heating...............12 ......12 1. The incubator block is not getting warm ..... 12 2. Only one side of the incubator block is being heated Yumizen G400...
  • Page 4 ......18 1. The instrument does not detect the pendrive ......18 2. The PC does not recognise the instrument Measurement position signal........... 19 ....19 1. Measurement position signalling does not work Serial protocol for Yumizen G400........20 Yumizen G400...
  • Page 5: Service Menu

    After entering it press the “OK” button. If you have entered an incorrect service code the instrument will allow you to re-enter the code. 6. After successfully entering the service code the instrument will show the service menu. Yumizen G400 - Service menu...
  • Page 6: Parts And Functions Of The Service Menu

    Place the vessel into the reagent incubation position ■ Switch on continuous mixing using the mixer button ■ Check that the mixing magnet starts to rotate and whether mixing stops when the button is pressed again. Yumizen G400 - Service menu...
  • Page 7 Password: You can change the default password of the instrument. DDimer: After inserting a pendrive containing the appropriate authorisation file using this button the authorisation of DDIMER measurement can be switched on and off. Yumizen G400 - Service menu...
  • Page 8: Display

    Measure the voltage on display connector (J8). The measured value should be above 9V. If the voltage is correct use a reference display and if this operates properly the problem is caused by the display assembly. Yumizen G400 - Display...
  • Page 9: Touch Screen

    Pin 15 (Chip Select) must be on ground potential and the voltage must be fallen to ground potential when the screen is touched. You need to check whether there is digital communication at position IC 12 when the screen is touched. Yumizen G400 - Touch screen...
  • Page 10: Printing

    There should be 9 V at pin 1 and 4 of the U4 IC which controls the burning. You should be able to measure 9 volts at pin 5, 6, 7 and 8 of the U4 IC Yumizen G400 - Printing...
  • Page 11 If the measured voltage is different then check if the control signal appears at the base of the T1 transistor. If the printer complies with all these measurements but it still does not print, check the instrument with a reference printer. Yumizen G400 - Printing...
  • Page 12: Date/Time

    32.7 kHz at pin 7 of the U3 IC. Set the current time, save it and switch the instrument off and on again. Wait for 1-2 minutes and check that the time displayed by the instrument is the same as the real time. Yumizen G400 - Date/Time...
  • Page 13: Measurement Result, Calibrations

    It is worthwhile doing a software update on the instrument, and then checking if the instrument is working properly before replacing the EEPROM-s. If the EEPROM-s are replaced you will have to press the “Setup defaults” button in the SERVICE menu. Yumizen G400 - Measurement result, calibrations...
  • Page 14: Mixer

    If there are no square wave signals check that the control is arriving at the U19 input. If there is control at pin 2, 6, 13 and 9, and there are no breaks in the coils then the U19 controller is defective. Yumizen G400 - Mixer...
  • Page 15: Temperature Measurement

    . You can measure the output voltage at pin 1 of the U20 circuit. If the measured value is not appropriate on the basis of these parameters then there may be a software fault. Yumizen G400 - Temperature measurement...
  • Page 16: Incubator Heating

    T8 and T9 check the power supply of the transistors as well. Only one side of the incubator block is being heated Perform the measurements described in the The incubator block is not getting warm chapter on the colder side of the incubator block. Yumizen G400 - Incubator heating...
  • Page 17: Controller, Program

    It is not possible to update the software Check that the name of the file copied onto the pendrive is coag.ahx. How to update the software: Yumizen G400 - Controller, Program...
  • Page 18 Place the pendrive containing the new software into the instrument ■ Switch on the instrument and allow the update ■ After the update has been completed the instrument will restart automatically ■ After this remove the pendrive Yumizen G400 - Controller, Program...
  • Page 19: Measuring Head, Measurement Results

    On pin 1 of the connector 3.3. V you should be able to measure while pin 5 is at ground potential. Measure pins 1 and 7 Yumizen G400 - Measuring head, measurement results...
  • Page 20 Finally check the change of voltage on pin 1 of the U304, U305, U306, U307 ICs, and measure the power supply on the ICs, on pin 8 you should measure 3.3 V, and on pin 4 you should measure ground potential. Yumizen G400 - Measuring head, measurement results...
  • Page 21: Serial Connection (Rs-232)

    The receiver should support zero modem connection. The serial cable used for communication must be in faultless condition. Check the settings needed for communication. The settings of Yumizen G400 and the receiver must be identical. The Yumizen G400 RS-232 settings are as below: Serial line basic values...
  • Page 22: Usb Connection

    ■ Restart the PC operation system, then switch on the Yumizen G400 and connect it to the PC. If the PC still does not recognise the Yumizen G400, check the PC operation system, whether it supports USB connection. If there is still no connection, check the Yumizen G400 as described in the The instrument does not detect the pendrive chapter.
  • Page 23: Measurement Position Signal

    PCB. You must measure voltage on pins 15, 1, 2, 3, 4, 5, 6 and 7 of the U17 IC, if no signal is measured on these pins, then the U17 IC has probably failed. Yumizen G400 - Measurement position signal...
  • Page 24: Serial Protocol For Yumizen G400

    ’ --- ’ if not defined Separator ’;’ (ASCII=59) Result field: Ratio pppp,pp or ’ --- ’ if not defined Separator ’;’ (ASCII=59) Result field: INR pppp,pp or ’ --- ’ if not defined Separator ’;’ (ASCII=59) Yumizen G400 - Serial protocol for Yumizen G400...
  • Page 25 00 0000 1000: Curve error (suspicious data during measure phase) 00 0001 0000: Out of range error( time is out of normal range) 00 0010 0000: Incubation error (the time was too long) Yumizen G400 - Serial protocol for Yumizen G400...
  • Page 26 00 0100 0000: Expired lot ( e.g.: „Expired lot” and the „Curve error” in same time: 01001000 hex: 48h, the error code is „072”) Yumizen G400 - Serial protocol for Yumizen G400...