Withings BP-801 - Blood Pressure Monitor Manual

Withings BP-801 - Blood Pressure Monitor Manual

Also See for BP-801:

Advertisement

Withings BP-801 - Blood Pressure Monitor Manual

INTRODUCTION

Requirements

In order to use your Wireless Blood Pressure Monitor, you need an iOS (7.0 or higher) or Android (4.0 and higher) device.

Package content

Withings Wireless Blood Pressure Monitor
Four AAA alkaline (LR3) cells (already inserted) USB adapter

Intended use

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches (22cm- 42cm).
This device is not intended to be a diagnostic device.
Contact your physician if hypertensive values are indicated.

AFTER USE

Cleaning

  • Do not use an alcoholic-base or solvent agent to clean the device
  • Clean the device with a soft and dry cloth
  • The dirt on the cuff can be cleaned by a moisten cloth and soap
  • Do not flush the device and cuff with much water
  • Do not dismantle the device or disconnect the cuff or try to repair by yourself. If any problem happens, refer to the distributor
  • Do not operate the device under severe environment of extremetemperature or humidity, or direct sunshine
  • Do not shake the unit violently
  • Do not submerge the device or any of the components in water
  • Do not let the device under strong shocks, such as dropping the unit on the floor

Storage

  • If you are not using the device for an extended period, remove the alkaline cells from the aluminium tube for storage.
  • Store the device and the components in a clean and safe location

Maintenance

If you can't fix the problems using the troubleshooting instructions, request service from your dealer. Manufacturer will make available on requested circuit diagrams, component part lists, descriptions, calibration instructions, or other information that will assist manufacturer's staff or authorized representative for repair.
It is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy.

ERROR MESSAGES

Error message Causes Countermeasures
Measurement could not be performed. Please try again. If the problem occurs again, please contact customer service. Hardware auto- ‐check error.
  1. 1Measurement could not be performed. Please try again. If the problem occurs again, please contact customer service.
Connection error. Quit the application and reconnect your blood pressure monitor. Communication error between the Blood Pressure Monitor and the collateral device.
  1. Push the blood pressure monitor button for 3 seconds to turn it off. Quit the application and push the blood pressure monitor button to turn it on.
  2. If the problem occurs again, please contact customer service.
Please wait until the cuff is totally deflated before starting a new measurement. Stay still during measurement.
  • Cuff is not deflated before measurement.
  • Noise interference.
  • Excessive user movements.
  1. Please wait until the cuff is totally deflated before starting a new measurement. Stay still during measurement.
  2. If the problem occurs again, please contact customer service.
Check that the Blood Pressure Monitor is correctly positioned on your arm and that measurement is performed in good conditions.
  • User movement, or not relaxed, or talking...
  • Cuff is not correctly fastened
  • Cuff broken
  • Pump or valve failure
  • Pressure overflow
  • Noise interference
  • User movement, or not relaxed, or talking...
  • User special characteristics
  • Battery level is too low
  1. Check that the Blood Pressure Monitor is correctly positioned on your arm and that measurement is performed in good conditions.
  2. Low battery level. Unplug the Blood Pressure Monitor. Replace the alkaline cells.
  3. If user has special characteristics, please contact your physician.
  4. If the problem occurs again, please contact customer service.
Low battery level. Unplug the Blood Pressure Monitor. Replace the alkaline cells. Battery level is too low.
  1. Low battery level. Unplug the Blood Pressure Monitor. Replace the alkaline cells.
  2. If the problem occurs again, please contact customer service.

TROUBLESHOOTING

Problem Remedy
Though the batteries are installed, there is nothing happening when connected to iOS or Android device
  1. Check that your iOS or Android device is unlocked
  2. Check and correct the AAA alkaline cells polarities
  3. Remove the AAA alkaline cells and wait for one minute. Then install the AAA alkaline cells back.
  4. Replace the AAA alkaline cells
The inflation action cannot be done or the air pressure cannot rise
  1. Check the cuff position and fasten the cuff correctly ad remeasurethe pressure again
  2. Check the blood pressure monitor connection to the iOS or Android device

The low battery icon is shown in the application

Replace the old alkaline cells by four new AAA alkaline cells.
The blood pressure cannot be taken and the application shows an error message or wrong result
  1. Re-fasten the cuff
  2. Relax yourself and sit down
  3. Keep the cuff and heart at the same level during the measurement period
  4. Keep silent and still during measurement
  5. If the patient has sever heart beat problem, then the blood pressure may not be read correctly
Under normal measuring circumstance, the reading at home is different from that of the clinics or each measurement has various reading
  1. The variation is due to the different environments
  2. The blood pressure is changing according to the physiological or psychological status of the human body
  3. Show your recorded values to your physician

SPECIFICATIONS

Technical data

Product description: Digital automatic Blood Pressure Monitor
Model: BP-801 (Withings)
Measurement method: Cuff oscillometric method
Cuff inflation: Automatic inflation with air pump at 15 mmHg/s
Pressure sensor: Gauge sensor
Measurement range(pressure): 0 to 285 mmHg
Measurement range (pulse): 40 to 180 beats/min
Accuracy (pressure): Within +- 3 mmHg or 2% of reading
Accuracy (pulse): Within +-5% of reading
Sensor: Semiconductor pressure sensor
Operating conditions: 10 to 40°C, 15 to 90% RH, atmospheric 86Kpa~106kpa, altitude: 2000m
Storage and transport conditions: -25 to 70°C,10 to 95%RH,atmospheric 86Kpa~106kpa, altitude: 2000m
Arm type: Circumference 22-42cm (9"-17"), Dimensions: 150(L) x 140(W) x 100(H)
Power source: AAA alkaline cells (x4)
Weight: Approx. 600g without cells
Accessories: AAA Alkaline cells x4, Micro-USB to USB adapter, Instruction manual
Product life: 5 years
Wireless transmission V4.0 dual mode (V2.1 + EDR)

Note: Specifications are subject to change without prior notice or any obligation on the parts of the manufacturer

CAUTION

Important safety information

Consult your doctor during pregnancy, or if you suffer from arrhythmia and arteriosclerosis. Please read this section carefully before using the Withings Wireless blood pressure monitor.

General safety and precautions

  • Do not forcibly bend the arm cuff
  • Do not inflate the arm cuff when it is not wrapped around your arm
  • Do not apply strong shocks and vibrations to the blood pressure monitor or drop it
  • Do not take measurement after bathing, drinking alcohol, smoking, exercising or eating.
  • Do not immerse the arm cuff in water

General usage

Always consult your doctor. Self-diagnosis of measurement results and self treatment are dangerous People with severe blood flow problems, or blood disorders, should consult a doctor before using the blood pressure monitor. Cuff inflation can cause internal bleeding. Operational factors such as common arrhythmias, ventricular premature beats, atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-eclampsia or renal disease can affect the performance of the automated sphygmomanometer and/or its blood pressure reading. This device is a precision measuring equipment liable to be understood by lay user but it still should be handled with care. A long exposure of the device to lint, dust or sunlight might reduce its life time or damage it. Damaged cuff or sensor might lead to incorrect measurements.

  • Do not leave the blood pressure monitor unattended with infants or persons who cannot express their consent.
  • Do not use the blood pressure monitor for any purpose other than measuring blood pressure.
  • Do not disassemble the blood pressure monitor.– Do not operate the blood pressure monitor in a moving vehicle (car, airplane).
  • Do not use a cellular phone near the deivce. It may result in an operational failure.
  • Improper continuous pressure of cuff or too frequent measurements may interfere blood flow and result harmful injury to the users.
  • Do not apply the cuff over the users' arm bearing a wound or medical treatment and so on, as this can cause further influcence on the therapy.

AAA alkaline cells usage

  • If AAA alkaline cells fluid should get on your skin or clothing, immediately rinse with plenty of clean water.
  • Use only four AAA alkaline cells with this blood pressure monitor.
    Do not use any other types of AAA alkaline cells.
  • Do not insert AAA alkaline cells with their polarities incorrectly aligned.
  • Replace old AAA alkaline cells with new ones immediately.
    Replace all four AAA alkaline cells at the same time.
  • Do not use new and used AAA alkaline cells together

Replacing the AAA alkaline cells

If the low battery symbol appears in the Withings application, replace all four AAA alkaline cells at the same time.

  1. Remove the AAA alkaline cells cover at the lower end of the aluminum tube.
  2. Install or replace four AAA alkaline cells so that the + (positive) and – (negative) polarities match the polarities indicated on the AAA alkaline cells compartment.
  3. Put the AAA alkaline cells cover back in place.

If the device will not be used for a long period of time, it is advised to remove the alkaline cells.

YA HORNG ELECTRONIC CO., LTD.
ATTEN ELECTRONIC (DONGGUAN) CO., LTD
188 industrial district, Ping Shan Administrative District,
Tang Shia Town, Dongguan, Guangdong, China

Australia representative:
PICA Australia Pty. Ltd.
11/37 Keilor Park Dr, Keilor Park, VIC
3042 Australia
Tel: +61 3 9336 7336

USA incorporated address:
Withings Inc.
16192 Coastal Highway Lewes DE 19
United States of America

USA business office:
101 Main St
14th Floor - Withings Inc
CMBRIDGE, MA02142
United States of America

Kahl Handelsvertretung
Isarstr. 33 40699 Ekrath, Germany
Tel: +49 21 0447 754

Documents / Resources

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Withings BP-801 - Blood Pressure Monitor Manual

Advertisement

Need help?

Need help?

Do you have a question about the BP-801 and is the answer not in the manual?

Questions and answers

Table of Contents