SAFETY PRECAUTIONS ········································································ 2 PRECAUTIONS ON CE MARKING ······························································· 5 NOTATIONS USED IN THE INSTRUCTION MANUAL ·············································· 6 IMPORTANT PRECAUTIONS WHEN USING WGA-910A AND OPTIONAL UNITS ······················ 6 1. BCD OUTPUT ················································································ 7 1-1. OUTLINE OF PRODUCT ··································································· 7 1-2. CONTROLS AND INDICATORS (REAR PANEL) ·············································· 7 1-3.
STANDARD ACCESSORIES The following accessories are packed with the WGA-910A-1, WGA-910A-2, WGA-910A-3, WGA-910A-12. When unpacking, check the contents to ensure that all accessories are enclosed. Instruction Manual 1 (CD-R) *1 Product Warranty Software License Agreement *1. Including a PC software and a setting file for wave comparison mode.
PRIOR TO USE For safe use of the product, do not forget to read the “Safety Precautions” prior to use. Kyowa Electronic Instruments Co., Ltd. assumes no liability for any damage resulting from the user's failure to comply with the safety precautions.
Page 7
●Confirmation of Protective Ground To prevent an electric shock hazard, be sure to check safe protective ground function before operating the WGA-910A. If any abnormal matters are found in the protective ground function, fuse holder, etc., do not operate the WGA-910A.
Page 8
After the power ON, always preheat the product for approximately 30 minutes. ● Cleaning When the product gets dirty, clean the product with a dry soft cloth. When dust exists inside the WGA-910A, clean it by using an air blow gun. Do not touch electronic parts.
PRECAUTIONS ON CE MARKING NOTE When satisfying the following conditions, we evaluated the compatibility against the requirements of the EU Directives and confirmed that the product satisfied the requirements. ● To connect products, be sure to use the shielded D/A output cable, shielded BCD output cable and shielded RS-485 cable.
NOTE Essential precautions required when handling the product. MEMO Reference items required when handling the product. IMPORTANT PRECAUTIONS WHEN USING WGA-910A AND OPTIONAL UNITS NOTE The output signal (voltage/current) when the transducer cable is broken. When the transducer cable is broken, the product may output the following output signal (voltage/current).
1. BCD OUTPUT 1-1. OUTLINE OF PRODUCT The WGA-910A-1 is an instrumentation amplifier with the BCD option. These products are able to retrieve the indicated values as the BCD data. The product is able to control the PLC (sequencer), and be able to print data by using printers.
(Note 2): Detecting Block Output (Pin No. A13) When in detecting block: Open collector output is ON (Low). NOTE The model and manufacture of connectors in BCD option is different from that of BCD connectors in BCD and D/A option (Model: WGA-910A-12).
FUJITSU COMPONENT LIMITED NOTE The model and manufacture of connectors in BCD option is different from that of BCD connectors in BCD and D/A option (Model: WGA-910A-12). 1-6. SETTING BCD OUTPUT (1) Transmission Speed Sets the transmission speed of the BCD data output.
Page 14
2) Press the target key and Register key to register the target value. EO C O ut put Logi c Ret ur n M easur e Posi t i ve Logi c Negat i v Logi c For EO C out put l ogi c: negat i ve Negat i ve Logi c ( open-col l ect or out put : O N) For EO C out put l ogi c: posi t i ve...
(5) Data Form Sets to change BCD or binary for the type of outputted data. <Setting Range> BCD / Binary BCD O ut put Set t i ng Ret ur n M easur e <How to operate> The t r ansm i ssi on speed of BCD Tr ansm i ssi on Speed dat a i s set .
Page 16
(6) BCD Classification Outputs the target data to the BCD output. <Setting Range> Hold Value 1, Hold Value 2 or Measured value (See next page for details) <How to operate> 1) Press the BCD Classification key on the "BCD Output Setting BCD O ut put Set t i ng Ret ur n M easur e...
Page 17
< Setting Range of BCD Classification > When you change the operation mode, the items (screen) that can be set change. Items are as follows. Mode name Notation of setting Hold Value 1 Hold Value 2 Measured Value screen Normal Normal ―...
1-7. BCD SIGNAL TIMING The operation timing is as follows. Delay time (Delay time between input and BCD output) = (Delay time between input and D/A output) + (BCD conversion time) Delay time between input and D/A output : Max. 3 msec (Analog Filter: none, Moving Average: none) BCD conversion time: Max.
Page 19
The BCD data output COM and binary data output COM are connected to the control input COM of the WGA-910A. When the indicating value is in the detecting block or holding block, as the WGA-910A outputs the BCD data or binary data, the “Detecting Block Output” or “Holding Block Output” becomes ON.
(D/A) output to work with the measured value. In addition, the output circuit is insulated from the WGA-910A main circuit and the resolution is 14-bit or equivalent to -10 to +10 V and 4 to 20 mA.
2-3. D/A OUTPUT SPECIFICATIONS Output Voltage ....±10 V (load resistance 2 kohm or more), arbitrary scaling is available. Output Current ....4 to 20 mA (load resistance 500 ohm or less) 4 to 20 mA output is fixed when the voltage 0 to 10 V is applied. Insulation Voltage 250 VAC for 1 minute (output voltage and output current are non-insulated) Conversion Speed .....
2-6. D/A OUTPUT SET 2-6-1. D/A Display Sets the measured value (zero display) of 0 V or 4 mA output and the measured value (full display) of 10 V or 20 mA output. <Setting Range> Zero Dsp. (Zero Display): -99999 to 99999 O pt i on Set t i ng Ret ur n M easur e...
2-6-2. D/A Classification Outputs the target data to the D/A output. <Setting Range> Hold Value 1, Hold Value 2 or Measured value (See next page for details) <How to operate> 1) Press the D/A Classification key on the "D/A Output Setting" D/A O ut put Set t i ng Ret ur n M easur e...
Page 24
< Setting Range of D/A Classification > When you change the operation mode, the items (screen) that can be set change. Items are as follows. Mode name Notation of setting Hold Value 1 Hold Value 2 Measured Value screen Normal Normal ―...
The WGA-910A-3 is instrumentation amplifier with the RS-485 option. These products are able to set various functions and read measured values by using external devices. These products are able to connect multiple WGA-910A-3 units on the RS-485 network and able to control them. NOTE The RS-232C of the standard equipment is invalid and the RS-485 is valid when equipping the optional RS-485.
+ (Noninverting A) TD/RD+ - (Noninverting B) TD/RD- RS-232C - (Noninverting B) TERMINATOR N.C. Short N.C. TERMINATOR NOTE Turn OFF the both WGA-910A and external device before connecting a cable. Or, the WGA-910A may not function correctly or cause damage...
Page 27
Set external devices by conforming to signal specifications of the WGA-910A. RS-485 transceiver of the WGA-910A is provided with fail safe function that is designed to have the CPU output (RO) becomes 1 (Hi) when signal is received and when between [+ (noninverting A)] and [- (noninverting B)] is short-circuited or in float state.
Initial value of the baud rate: The recommended command interval of the measured value output command [GDT1] is 50 ms or more in 9600 bps. Command transmitted Command responded by from the PC etc. the WGA-910A Transmit 2 to 20 ms Response command 5 ms...
3-7. CONTROL COMMAND The following is an explanation of the formats and kinds of control commands which are output to the WGA-910A from external equipment such as PC. Sets the control command that is output from the external device to WGA-910A.
3-8 OPERATION AND FUNCTION OF THE COM. SETTING (COMMUNICATION SETTING) (1) Communication Speed Sets the communication speed (baud rate) of the RS-485. <Setting Range> 2400, 4800, 9600, 19200 (bps) <How to operate> 1) Press the Com. Setting key on the “Option Setting” window. O pt i on Set t i ng Ret ur n M easur e...
Push t he t ar get key and ' Regi st er ' key. Regi st er (3) Device ID Sets the ID of the WGA-910A. The Device ID key is displayed in gray when the optional RS-485 is not equipped. <Setting Range>...
4. BCD and D/A OUTPUT SETTING 4-1. OUTLINE OF PRODUCT The measured value can be acquired as BCD or binary coded data, and the measured value output converted to the analog signal can be taken out simultaneously by the optional BCD and D/A output. 4-2.
Compatible connectors of the cable side are as follows when connecting the external device. Compatible connector Connector Model: 10136-3000PE Manufacture: Sumitomo 3M Limited Cover Model: 10336-52F0-008 Manufacture: Sumitomo 3M Limited NOTE The model and manufacture of connectors in BCD and D/A option is different from that of connectors in BCD option (Model: WGA-910A-1).
The setting contents of D/A output of the optional BCD and D/A output is the same as that of the optional D/A output. (Refer to “2-6. D/A Output Set.”) In addition, the output circuit is insulated from the WGA-910A main circuit and the resolution is 14 bits equivalent. (Output range: -10 to +10 V, 4 to 20 mA) Conversion speed is 2000 times/sec.
If the product is damaged due to any causes other than as written in the Instruction Manual or if the user has disassembled or remodeled the product, Kyowa Electronic Instruments Co., Ltd. may not be able to offer repair service.
Maintenance service should be performed at the intervals specified below. (1) Calibration of apparatus An annual maintenance inspection is recommended to ensure the operation and measuring accuracy of the product. Must be returned to KYOWA for the scheduled maintenance inspection. (2) Replacement of limited life parts Item Content...