18. Alarms, User Alerts and User Actions
Alarm On-Screen
Alarm
Message(s)
Description
No alarm message,
Temperature
however affected
Sensor
accessory should
measured
be highlighted in
>1°C under
BLUE
to indicate low
set point.
temperature status.
Visual :
GREEN
Audible : NONE
Low battery icon
Battery Low
below should be
flashing on screen.
Visual :
GREEN
Audible : LOW
External power and
Loss/
battery charging
disconnection
icons disappear.
of external
power.
Visual :
GREEN
Audible : Single 2
tone pulse
Inspiration Healthcare Limited
W inspiration-healthcare.com
Actions to Clear/
Silence
User must acknowledge
alert to clear audible
indicators. Audible alert
will be paused for 15
minutes.
Accessory will continue
to be highlighted
BLUE
to
indicate low temperature
status.
Visual and audible alert
indicators remain active
until device has been
connected to supply mains
or superseded by ALARM-
3-BATTERY-CRITICAL.
After connection to supply
mains this alert will
automatically be cleared.
The audible alert for this
condition may not be
paused.
No user action required
Single two tone audible
alert sounded to indicate
disconnection of mains.
© 2021 Inspiration Healthcare Limited
Patient Warming System Instructions for Use
continued
Cause(s)
Ambient conditions may
limit maximum achievable
temperature.
Failure of AlphaCore
5
Controller to supply
appropriate power.
Failure of mattress.
Failure of temperature
measurement circuit.
Failure of mattress thermistor.
Failure of warming system.
Failure of system to maintain
temperature.
Failure of temperature
measurement circuit.
Battery is depleted (normal
operation).
Failure of battery status
monitoring circuit.
Loss of supply mains
when battery is not yet
appropriately charged.
Mains power supply is
disconnected (either
deliberately or accidentally).
N.B. The device must be
fitted with a suitably charged
internal battery. With no
battery fitted and no external
power source the unit will
simply shut down.
41
Actions
System
continues to
operate all
connected
accessories.
System
continues to
operate all
connected
accessories
until battery
charge level no
longer supports
warming
operation.
User may
reattach power
supply or
continue to
operate unit
on battery
power subject
to maximum
accessory
load limits.