18. Alarms, User Alerts and User Actions
Alarm On-Screen
Alarm
Message(s)
Description
Full screen warning.
Internal device
temperature
measures >85°C
Visual :
AMBER
(LOW)
Audible : LOW
Full screen warning.
Internal fan failure. When this alarm state
AMBER
Visual :
(LOW)
Audible : LOW
Full screen warning.
Battery Critical.
Visual :
AMBER
(LOW)
Audible : LOW
Inspiration Healthcare Limited
W inspiration-healthcare.com
Patient Warming System Instructions for Use
Actions to Clear/
Silence
When this alarm state
is initiated, the unit will
automatically shut-down
in 1 minute.
Operator can press the
power button to turn
the unit OFF before the
unit automatically shuts
down.
Operator must allow the
device to cool and ensure
air vents are not blocked
and then restart the
device.
is initiated, the unit will
automatically shut-down
in 1 minute.
Operator can press the
power button to turn
the unit OFF before
the unit automatically
shuts down.
Operator may attempt to
power cycle the device.
When this alarm state
is initiated, the unit will
automatically shut-down
in 1 minute.
Operator can press the
power button to turn
the unit OFF before
the unit automatically
shuts down.
Operator must connect
external power and then
restart the device.
© 2021 Inspiration Healthcare Limited
continued
Cause(s)
Actions
Failure of internal
Cut all power to
components.
warming system.
Failure to adequately
Alert sounds
cool (e.g. blocked
for 60 seconds
vents).
followed by system
shutdown.
Contact technical
support for repair.
Failure of internal fan.
Cut all power to
warming system.
Failure of internal
Alert sounds
components.
for 60 seconds
followed by system
shutdown.
Contact technical
support for repair.
Battery is critically
Cut all power to
depleted and no
warming system.
external power supply
Alert sounds
is connected.
for 60 seconds
Failure of battery
followed by system
status monitoring
shutdown.
circuit.
Contact technical
Loss of supply mains
support for repair.
when battery is not
yet appropriately
charged.
37
700-307 Revision: 10