14
Inspiration Healthcare
6. AlphaCore
AlphaCore
Controller Start-Up Cycle
5
After pressing the power button and the unit powers
up, the bootloader screen is displayed. At the final
stage of the bootloader process the HIGH, MEDIUM,
and LOW alarm audible signals should each be
initiated for a signal alarm burst. At the same time,
the visual alarm bar at the top of the unit should be
illuminated
RED
or
of the audible alert being sounded.

Bootloader Screen

In the first stage of the bootloader, the visual alarm
bar should be
GREEN
alarms sounding.

Bootloader HIGH PRIORITY ALARM Test

The visual alarm bar should be illuminated in RED.
The audible alarm should complete at least one full
sequence of the HIGH PRIORITY ALARM tone.
Inspiration Healthcare Limited
W inspiration-healthcare.com
Controller Start-Up
5
AMBER
according to the priority
and there should be no audible
© 2021 Inspiration Healthcare Limited

Bootloader MEDIUM PRIORITY ALARM Test

The visual alarm bar should be illuminated in AMBER.
The audible alarm should complete at least one full
sequence of the MEDIUM PRIORITY ALARM tone.

Bootloader LOW PRIORITY ALARM Test

The visual alarm bar should be illuminated in
AMBER.
The audible alarm should complete at least one full
sequence of the LOW PRIORITY ALARM tone.
700-307 Revision: 10