Regular cleaning measures

For trouble-free operation, briefly operate the rinsing basin flush after each treatment and flush each suction hose with cold water to remove residues from the lines.
The following cleaning measures must be carried out regularly:
Measure
Cleaning and disinfection of the extraction
system
Vessel change
7.1.1.
Daily cleaning with GREEN&CLEAN M2
2 x daily (noon/evening) and after surgical interventions, disinfection must be carried out with the prescribed disinfectant and cleaning agent GREEN&CLEAN M2.
Ideally, GREEN&CLEAN M2 should be used prior to the treatment unit remaining at a standstill for extended periods (lunchtime, end of the day or holiday).
For information on use and safety instructions, see the GREEN&CLEAN M2 instructions for use.
7.2.

Maintenance and service

7.2.1.

Replacement of the collection vessel

1
Provide a new collection vessel and remove the disinfectant bag inside
2
Tilt the locking bracket on module 2 upwards, remove it and place it on a flat, non-slip surface.
3
Open the 4 yellow clip fasteners on module 2, hold the collection vessel and lift off the upper part.
18
Interval
2 x daily
If required
Assembly, modifications and repairs may only be carried out exclusively by authorised specialist personnel (see 3.2. Safety
instructions)! For further information and assistance in carrying out repairs, retrofitting, fault analyses, etc., the METASYS
technical customer service is also available!
Warning:
Switch off the main switch of the treatment unit!
Warning:
Avoid contact with the contents of the amalgam separator!
Warning:
Risk of contamination: To avoid infection, wear personal protective equipment (hand, eye, mouth-nose protection) and
disinfect and clean the device!
see 7.1.1. Daily cleaning with GREEN&CLEAN M2
See 7.3.1. Replacement of the collection vessel
and 7.3.2. Disposal of the collection vessel