MULTI SYSTEM TYP 1 ECO
MULTI SYSTEM TYP 1 ECO Light
Instructions for use
EN | ZK-58.205/01| 2022-10
eIFU:
www.metasys.com/downloads

Advertisement

loading
Need help?

Need help?

Do you have a question about the MULTI SYSTEM TYP 1 ECO and is the answer not in the manual?

Questions and answers

Summary of Contents for Metasys MULTI SYSTEM TYP 1 ECO

  • Page 1 MULTI SYSTEM TYP 1 ECO MULTI SYSTEM TYP 1 ECO Light Instructions for use EN | ZK-58.205/01| 2022-10 eIFU: www.metasys.com/downloads...
  • Page 2: Table Of Contents

    5.4. Electronics ............................................15 5.4.1. Electrical connections ......................................15 5.4.1.1. MULTI SYSTEM TYP 1 ECO .................................... 15 5.4.1.2. MULTI SYSTEM TYP 1 ECO Light ..................................16 Use ..............................17 6.1. Normal operation ..........................................17 6.2. Internal display (MST 1 ECO only) ....................................17 6.3.
  • Page 3 ZK-58.205/01 If you would like to request a printed copy of the instructions for use, please contact us at ifu@metasys.com or use the order form at www.metasys.com/downloads A hardcopy of the instructions for use will be made available to you free of charge and within seven calendar days of receipt of the request.
  • Page 4: Notes

    Notes 1.1. General instructions METASYS can only guarantee the safety, reliability and performance of the dental device if the following instructions are adhered to: > The product is only to be used in accordance with the instructions for use. >...
  • Page 5: Copyright Notice

    Name and address of the Fault on end device Use eye protection authorized representative‘s registered office in Switzerland 1.3. Copyright notice All names and contents are protected by copyright. Distribution, duplication or alternative use of this document is only permitted with the written consent of METASYS Medizintechnik.
  • Page 6: Intended Use

    Intended use Intended use The MULTI SYSTEM TYP 1 ECO / ECO Light is used to separate water and air in the suction flow of dental treatment units. Furthermore, the integrated place selection valve enables the interruption of the suction current. 2.1. Indication Not applicable. 2.2. Contraindication Not applicable. 2.3. Target group The device shall be used by trained and qualified personnel exclusively.
  • Page 7: Safety-Related Information

    Safety-related information Safety-related information 3.1. General safety-related information All serious incidents related to the device must be reported to the manufacturer and the competent authority of the Member State where the user and/or the patient is resident. 3.2. Safety instructions Assembly, modifications or repairs shall only be carried out by authorised qualified personnel who guarantee compliance with the EN 60601-1 standard (international standard on Medical Electrical Equipment and Systems, in particular part 1: General requirements for basic safety).
  • Page 8: Product Description

    Product description Product description 4.1. Product description MULTI SYSTEM TYP 1 ECO / ECO Light is a single-place, air/water separator with integrated heavy parts collector and place selection valve. For simplified readability of the texts, the abbreviations MST 1 ECO and MST 1 ECO Light are used below. 4.2. Technical data / performance data Height [H] Width [W]...
  • Page 9: Layout

    > Station selection valve > Diagnostics board (only MST 1 ECO) MULTI SYSTEM TYP 1 ECO MULTI SYSTEM TYP 1 ECO Light Cover: Cover (optional) 4.5. Functional description Lifting a suction hose opens the place selection valve. The suction current is fed into module 2, where the air/water separation takes place by means of the cyclone principle.
  • Page 10 Product description Station selection valve Module 2 (suction current input) Air/water separation Air outlet (to suction motor) Collection vessel Conductivity probe Pump motor Waste water output...
  • Page 11: Preparation For Use

    Preparation for use Preparation for use 5.1. Transport and storage The device is shipped in two polystyrene trays fixed together. This original packaging must be used for any onward and return transport. The device must always be transported and stored in an upright position. The device must be transported to the installation site in a completely packed state. After unpacking the device, check for completeness and possible transport damage. 70 °C Transport and storage temperature 0 °C...
  • Page 12: Pipe And Hose Connections

    Preparation for use Installation in the back-of-the-head position If there is no rinsing basin at the treatment site and the suction is mounted in the behind-the-head position, it is a good idea to install the device in a setting. In this case, the connection opening for the rinsing basin drain must be closed with a blind plug. Set-up with cover If it is not possible to install the device directly in the treatment unit, the set-up can also be carried out in the optionally available cover.
  • Page 13: Pre-Filter

    Meet spatial requirements see 5.2. Installation requirements Installing the pre-filter (if not present in the hose rack) MULTI SYSTEM TYP 1 ECO / ECO Light Built-in version Fitting a contact protection plate behind module 1 (if the subsurface is electrically conductive) Connecting the rinsing basin drain hose Connecting the suction line from the hose tray (suction hose)
  • Page 14 Preparation for use Connecting the drain hose Connecting the suction line to the suction machine (vacuum hose) MULTI SYSTEM TYP 1 ECO / ECO Light with cover Example picture Placement of the MST 1 ECO / ECO Light with cover: Rinsing basin drain above I (red), amalgam separator drain below O (yellow), min.
  • Page 15: Installation And Assembly Of Optional Accessories, Retrofit Parts And Spare Parts

    For further information and assistance in carrying out repairs, retrofitting, fault analyses, etc., the METASYS technical customer service is also available! 5.3.2. Connections with other devices Connecting the METASYS device to other devices or systems can be dangerous. Therefore, it must be ensured that no hazards arise for the user or the patient and that the environment is not affected. The specifications of the manufacturer of the device or system to be connected must be observed. 5.4. Electronics The mains connection may only be carried out by a qualified electrician. The electrical installation must be carried out in accordance with the applicable local regulations.
  • Page 16: 5.4.1.2. Multi System Typ 1 Eco Light

    Preparation for use 5.4.1.2. MULTI SYSTEM TYP 1 ECO Light X1: Power supply 24 V AC X2: Connector for stacking signal X1: Solenoid valve X2: Pump SV1: Connection for probes Stacking signal 24 V DC Main fuse T3, 15 A/UN 250 V/ ICN 35 A...
  • Page 17: Use

    No function No function 6.3. Error messages Assembly, modifications and repairs may only be carried out exclusively by authorised specialist personnel (see 3.2. Safety instructions)! For further information and assistance in carrying out repairs, retrofitting, fault analyses, etc., the METASYS technical customer service is also available! Error message Possible cause Countermeasures Nettoyer ou remplacer l‘électrovanne (voir 7.2.6.
  • Page 18: Care And Maintenance

    Ideally, GREEN&CLEAN M2 should be used prior to the treatment unit remaining at a standstill for extended periods (lunchtime, end of the day or holiday). For information on use and safety instructions, see the GREEN&CLEAN M2 instructions for use. 7.2. Maintenance and service Assembly, modifications and repairs may only be carried out exclusively by authorised specialist personnel (see 3.2. Safety instructions)! For further information and assistance in carrying out repairs, retrofitting, fault analyses, etc., the METASYS technical customer service is also available! Warning: Switch off the main switch of the treatment unit! Warning:...
  • Page 19: Disposal Of The Collection Vessel

    Disposal of the collection vessel The full collection vessel can be submitted to the company‘s own disposal organisation, METASYS logistics & collection GmbH. When the complete device is disassembled at the end of its service life, it can be returned to the manufacturer for correct disposal (see 8.2. Recycling and disposal).
  • Page 20: Normal Operation Test (Only For Mst 1 Eco)

    Empfänger / Consignee For other countries please visit: www.metasys.com/collection_centers Visit www.metasys.com/collection_centers for more information on disposal in your country! 7.2.3. Normal operation test (only for MST 1 ECO) The normal operation test tests the operating functions of the device. This test should be carried out after installation as well as after every repair! Switch on the main switch.
  • Page 21: Mst 1 Eco / Eco Light Service Kit

    Care and maintenance 7.2.4. MST 1 ECO / ECO Light service kit Check the correct installation and connection of the amalgam separator automatic separator in see 5.2. Installation requirements accordance with the installation requirements see 7.1.1. Daily cleaning with GREEN&CLEAN Carry out disinfection and cleaning of the suction hoses and mouth rinsing basins Carry out a normal operation test (only on MST 1 ECO) see 7.2.3.
  • Page 22: Decommissioning

    Risk of contamination: To avoid infection, wear personal protective equipment (hand, eye, nose and mouth protection) and disinfect and clean the device! If it is necessary to return the device to the depot or to METASYS, the original METASYS packaging must be used. Before packing the METASYS device to be transported, clean and disinfect it. Possible openings where residual fluids could escape must be closed.
  • Page 23: Annex

    MST 1 ECO Light - Tekmil Automatic separator with Tekmil accessories and instructions for use 9.1.1. Accessories, service kits, collection containers and spare parts MST 1 ECO / ECO Light = Abbreviation for MULTI SYSTEM TYP 1 ECO / ECO Light Accessories Order number Designation...
  • Page 24 Tube nozzle, 15 mm, 90°, 5 pcs. 40010100 Adapter, 15/18 mm, O-ring, 2 pcs. 40010012 Angle-piece water outlet, 15 mm 40010102 Adapter, for METASYS, female 40010015 Double nozzle, 22-25 mm 40010103 Lock clip, adapter 40010017 Extension air duct, 72.5 mm...
  • Page 25 Adapter inlet frame, 15-16 mm, 5 pcs. 40010130 Adapter, 15/10 mm, o-ring 40010089 Adapter inlet frame, 16-17 mm, 5 pcs. 40010131 T-adapter, mini-nozzles MULTI SYSTEM TYP 1 ECO - see General spare parts 50110003 MULTI SYSTEM TYP 1 ECO Light 50020205 50020045 50020102 50120003...
  • Page 26 Annex Cover 50020042 50020050 40500001 50020020 Hose nozzles and adapters META Connect ø 21 ø 26 ø 26 ø 26 ø 26 ø 26 ø 26 ø 21 ø 26 40010114 ø 21 ø 21 ø 21 ø 21 40010115 40010116 40010117 40010118...
  • Page 27 Annex Hose nozzles and adapters META Connect ø 21 ø 21 ø 26 ø 21 ø 26 ø 21 ø 26 ø 26 ø 31 ø 24 ø 25 ø 32 ø 16 ø 26 ø 26 ø 26 ø 26 ø...
  • Page 28: Warranty Conditions

    In this case the warranty period begins with commissioning. In the event of installation without returning the installation proof to METASYS, any warranty claim is forfeited. Installation and return of installation proof must be completed within 24 months from the date of sale from METASYS.
  • Page 29: Change History

    The customer sends the device or component in question to METASYS at his own expense. METASYS checks whether there is a warranty claim. METASYS will repair the device or component if it is cost-effective. The customer is charged for the costs incurred for the repair, but not the replacement parts covered by the guarantee.
  • Page 30 METASYS Medizintechnik GmbH Florianistraße 3 | 6063 Rum bei Innsbruck | Austria Subject to printing and setting errors! T +43 512 205420 | info@metasys.com | metasys.com...