5.
Preparation for use
5.1.
Transport and storage
The device is shipped in two polystyrene trays fixed together. This original packaging must be used for any onward and return transport. The device must always
be transported and stored in an upright position. The device must be transported to the installation site in a completely packed state. After unpacking the device,
check for completeness and possible transport damage.
70 °C
Transport and storage temperature
0 °C
80 %
Transport and storage humidity limitation
0 %
5.2.
Installation requirements
35 °C
Operating temperature: 10°C to 35 °C
10 °C
70 %
Humidity limit max. 70 %
0 %
Max. 3,000 m above sea level
≤ 3000 m
>
The following space requirement must be provided for the installation: H x W x T = 350 mm x 215 mm x 120 mm
>
To avoid vibrations, the attachment must only be made through the three holes on module 1 to load-bearing parts of the treatment unit! (Do not remove
vibration inserts!)
>
The device must be able to swing freely and must not have any contact with other parts of the treatment unit (cables, hoses, covers, etc.).
Caution:
A contact protection plate must be fitted behind module 1 if the subsurface is electrically
conductive!
5.2.1.
Assembly variants
Installation in the rinsing basin
Direct integration of the device into the rinsing basin should be used as the preferred solution, as the tubing
routes to the device can be kept as short as possible.
In the rinsing basin, ensure vibration-free suspension of the device from supporting parts of the treatment
unit.
Preparation for use
|
EN
11