6.
Use
6.1.
Normal operation
Ensure that the device for level measurement is switched off at least 1 x per working day. The 24 V AC power supply must be
connected after the device's or practice's main switch.
6.2.
Internal display (MST 1 ECO only)
1
2
3
4
6.3.
Error messages
Assembly, modifications and repairs may only be carried out exclusively by authorised specialist personnel (see 3.2. Safety
instructions)! For further information and assistance in carrying out repairs, retrofitting, fault analyses, etc., the METASYS technical
customer service is also available!
Error message
Suction current is not released
Bubble formation in the vessel when pump (module
2) is in operation
Pumping capacity not sufficient
Pump (module 2) does not start
Signal 1
Stacking signal
Signal 2
Solenoid valve in module 2
Signal 3
Pump module 2
Signal 4
No function
Possible cause
Solenoid valve (module 2) defective
Sieve for pump or water brake missing or dismantled Remonter ou remplacer le tamis ou le frein à eau
Sieve for pump missing or dismantled
Pump defective, probes dirty or defective
Suction hose lifted off
(12-24 V AC or DC at connector SV7)
The solenoid valve in module 2 is actuated
("Emergency stop" probe not addressed)
The pump in module 2 is actuated
No function
Countermeasures
Nettoyer ou remplacer l'électrovanne (voir 7.2.6.
Remplacement de l'électrovanne)
Remonter ou remplacer le tamis
Remplacer entièrement le module 2 ou nettoyer les
sondes (voir 7.2.5. Remplacement du module 2)
Use
|
EN
17