INTRODUCTION The Console Drive controls the speed of MASTERFLEX Pump Heads to provide ow rates from 0.12 to ® ® 17 L/min, or 26 L/min with two pump heads. Mount up to 2 MASTERFLEX EASY-LOAD or Standard Pump Heads or 1 MASTERFLEX High-Performance ®...
SETUP AND DRIVE OPERATION 1. Mount pump head and load tubing. (See pump head manual.) 2. Turn pump on and select tubing size by pressing “TUBING SELECT.” NOTE: If CAL LED is lit, that tubing size has been previously field calibrated. If LED is not lit, the drive is operating with the built-in factory calibration.
DISPENSE/COPY A first press of the DISP key results in the last entered dispense volume being displayed. The VOLUME annunciator will illuminate and flash. The INC/DEC (L, M) keys are used to change the dispense volume, if desired. The START/STOP key then initiates delivery of the set volume. The amount remaining to be dispensed will be displayed during countdown.
Page 6
5. To adjust the voltage or current scaling for other than zero to full scale, Press and hold the MODE key while pressing the FLOW key. This display will show “LO” for 2 seconds and then a flow rate. Use the UP/DOWN (L, M) arrow keys to set the minimum control level.
TROUBLESHOOTING AND MAINTENANCE Fuse Replacement 1. Place the power switch in the off position. 2. Disconnect the AC power input line cord from the receptacle. 3. Remove and check the fuse and replace if defective. 4. Reconnect the AC power input line cord to the receptacle. A.
Page 8
Troubleshooting (cont.) SYMPTOM CAUSE REMEDY B. (cont.) Motor does not rotate. B2. MODE control not properly set. 1. Check that the MODE control is Display lights. set to INTERNAL for operation with front panel control or to mA or V for operation with remote control.
Cleaning Keep the drive enclosure clean with mild detergents. Do not immerse. Replacement Parts and Accessories Description Part Number Fuse—T6.3A 77500-24 Hand-held remote control 07592-83 Remote control cable, 25 ft (7.62 m) 77300-32 Caster Kit 77420-01 Replacement Seal 77420-03 SPECIFICATIONS Output: Speed: 1 to 650 r/min...
WARRANTY Use only MASTERFLEX precision tubing with MASTERFLEX pumps to ensure optimum performance. Use of other tubing may void applicable warranties. The Manufacturer warrants this product to be free from signi cant deviations from published speci cations. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part, as determined by the Manufacturer.