4 Sensor specifications
3
The RBRduo
C.T | uv instruments use conductivity and temperature sensors.
3
The RBRconcerto
C.T.D | uv instruments use conductivity, temperature, and pressure sensors.
.
4.1 Conductivity (C)
RBR CT | uv and CTD | uv instruments use integrated inductive conductivity sensors which measure the ability of
seawater to conduct electric current.
Parameter
Range
Initial accuracy
Resolution
Typical stability*
Max depth rating
* For a clean cell with no biofouling.
The CT cell is rugged and durable. It is not affected by surface contaminants or freezing conditions, and can be frozen
into ice.
While 80% of its volumetric measurements happens inside the cell, they also extend up to 15cm away and thus may be
affected by conductive and non-conductive objects within this distance. RBR calibrates conductivity sensors to account
for static objects, such as cages, guards, and other sensors.
To maintain optimal accuracy, deploy your instrument at least 15cm away from other objects.
Conductivity measurements are temperature compensated.
RBR#0015699revB
Value
0 to 85mS/cm
±0.003mS/cm
<0.0001mS/cm
±0.010mS/cm/year
200m
CTD | uv cell, transparent view
9