A proving cable is available for testing the meter using the same IrDA (infrared) interface used for communication with the meter. The prover cable allows for rapidly testing Compensated and Non-Compensated accuracy on the Dresser Transfer Prover and the Honeywell American SNAP Proving Systems.
3. Once back on the Configuration screen (Figure 1.3 Example), select Upload to unit. Figure 1.3 Example - Configuration screen in Dresser MeterWare 4. The chosen prove mode can then be selected on the LCD screen display of the meter.
5. When you exit the test setup screen, you should see Meter Size: the meter test file in the “OTHER” manufacturer cat- egory. After you verify the meter test is added, install the meter on the test bench. (Refer to Section 1.4.3) 6 | Dresser Measurement...
5. Install the meter on the test bench. (Refer to 3. Once the cable is connected to the meter the other Section 1.4.3) end is connected to the optical outlet port (J4) on the prover as shown in Figure 1.12 Example. This is a 6.
“ACCEPT” button and the meter test will be saved. Figure 1.13 Example - Light on IrDA cable indicates connec- tion is established 4. Once the meter tests are complete you can press the “ACCEPT” button and the meter test will be saved. 8 | Dresser Measurement...
1.5 Proving with Dresser Model 5 Transfer Prover 1.5.2. Model 5 Prover Software Configuration 1.5.1 Establish IrDA Cable Connection The Model 5 Prover software must be set up as circled on the left side of the screen shot in Figure 1.18 Example.
Note: When entering values, always move to the next box by either hitting “Enter” or using the cursor. Using “Tab” will cause errors in the test configuration. Note: Contact factory to request pre-configured test files if preferred. 10 | Dresser Measurement...
Page 11
Series D Prover Testing Guide NGS.MI.0041c | 11...