S7 / OPMI Sensera Safety Measures Safety Measures Key to symbols .................. 6 Hazard symbols ...................6 Information symbols..................6 Target group ..................7 Field of application ................7 Intended use ....................7 Normal use....................7 Reasonably foreseeable misuse..............8 Liability and warranty................8 Notes for the operator............... 9 Requirements for operation.............
Safety Measures S7 / OPMI Sensera Key to symbols We would like to inform you about safety aspects which must be observed when handling this device. This chapter contains a summary of the most im- portant information concerning matters relevant to instrument safety.
Normal use OPMI Sensera is a surgical microscope designed to provide the user with op- tical magnification and illumination of the surgical area during surgical proce- dures in ENT surgery.
Safety Measures S7 / OPMI Sensera Reasonably foreseeable misuse CAUTION Injury to the patient's eye! Never use xenon illumination for ophthalmic procedures! • Make sure that no xenon light enters the patient's eyes. CAUTION Not for diagnostic purposes! Do not use the recorded video sequences, video clips (cut sequences) and single images for diagnostic purposes.
S7 / OPMI Sensera Safety Measures Notes for the operator The correct use of the device is absolutely vital for safe operation. Therefore, please thoroughly familiarize yourself with the content of these Instruction for Use before starting up the device. Please also observe the Instruction for Use of additional instrument equipment.
Page 10
(see section "Care and Maintenance"). • Modifications and repairs of this device or any equipment operated to- gether with this device may only be performed by Carl Zeiss service staff or other persons authorized by Carl Zeiss.
Page 11
S7 / OPMI Sensera Safety Measures • Start with the lowest brightness setting and gradually increase brightness up to the necessary and still admissible level. • With increasing age of the light source, the actual illumination intensity delivered at the respective setting decreases (this is a normal property of the system).
S7 / OPMI Sensera Requirements for operation Prior to the very first use Carl Zeiss Service or a person authorized by Carl Zeiss will install the device. Please make sure that the following requirements continue to be met for fur- ther operation: ...
S7 / OPMI Sensera Safety Measures Before every use • When adding accessories and/or components please ensure the permissi- ble total weight of the device is not exceeded (see label "Maximum load" or section "Technical Data"). • Before every use, make certain to compensate for any added weight; this will enable the surgical microscope to maintain its balance in all positions of the working range.
Page 14
• Do not force electrical connections. If connection is not readily possible, check whether the plug fits the socket. If any of the connectors are dam- aged, have the Carl Zeiss service team or authorized specialist staff repair them. • Position the suspension system in such a way that you can slide or swing it away from the patient at any time.
S7 / OPMI Sensera Safety Measures • Never look directly into the light source, e.g. a microscope objective lens or a light guide, as this might cause phototoxic injury to the eye. • Never leave a device unattended with the light source still switched on.
Page 16
Safety Measures S7 / OPMI Sensera luminated tissue moist and make sure it is sufficiently rinsed. Carefully monitor the effect of the illumination on the tissue, particularly under the following cir- cumstances: – During longer interventions on skin and tissue using objective lenses with short focal depth (short working distance), –...
Page 17
The area of the incision should be constantly irrigated. Final remark Carl Zeiss recommends to reduce illumination of the surgical field in that pos- Version 11.0 G-30-1434-en...
Page 18
Safety Measures S7 / OPMI Sensera sible extent to ensure the patient's security and a good microscopic image. Please refer to the warnings and precautions in this case. Version 11.0 Page 18 G-30-1434-en...
S7 / OPMI Sensera Safety Measures Safety devices of the suspension systems Halogen illumination system 1 Manual switching to the backup lamp The lamp housing contains a backup lamp which is automatically swung into the illumination beam path when the first lamp fails. If this automatic function fails, you can switch on the backup lamp by pressing this button.
Page 20
Safety Measures S7 / OPMI Sensera Xenon illumination system CAUTION The xenon lamp has a limited service life of 500 h! If used beyond its maximum service life, the xenon lamp may explode. • Please replace the xenon lamp in due time.
Page 22
Safety Measures S7 / OPMI Sensera Ceiling mount with lifting column Lifting column Selector switch for setting the optimum viewing height of the surgical microscope or for downward movement into the working position. As long as you keep the selector switch in the turned position, the lifting column moves upward (2) or downward (1), depending on the switch posi- tion.
Label on the knob for tilting the microscope upwards. 4 "Balance setting of front-to-back tilt axis" label Label on the knob for tilting the microscope downwards. 5 "OPMI Sensera" device label The device label contains the following information: OPMI Sensera –...
Page 25
S7 / OPMI Sensera Safety Measures 7 "Friction adjustment of front-to-back tilt axis" label Label on the knob for setting the required friction of the front-to-back tilt axis. 8 "Friction adjustment of lateral tilt axis" label Label on the knob for setting the required friction of the lateral tilt axis.
Page 26
Safety Measures S7 / OPMI Sensera 9 "Filter selector" marking The illustration on the left shows the marks on the filter selector knob: 10 "Focusing knob" marking The illustration on the left shows the marks on the focusing knob: 11 Labeling of zoom and focusing switches Labeling of the zoom and focusing switches on the handgrips.
Safety Measures S7 / OPMI Sensera Labels on S7 floor stand 1 "Risk of crushing" label Fingers may be pinched between the carrier and suspension arms. Do not touch this area while the surgical microscope is being moved. 2 "Maximum load" warning label The maximum load on the suspension arm must not exceed 14 kg.
Page 30
Safety Measures S7 / OPMI Sensera 7 "Observe Instructions for Use" label Observe the Instructions for Use or accompanying documents. 8 Year of manufacture This label indicates the year of manufacture of the device. 9 Rating label The rating label provides the following information: –...
Page 32
Manufacturer (company name) – Your options to contact the device manufacturer, i.e., phone number, fax number and email address of the local contact of the national Carl Zeiss sales organization. XXXX XXXXXX – SIP No. A unique identification number assigned to your device.
Safety Measures S7 / OPMI Sensera Additional labels on S7 floor stand (instrument tray option) 1 "Maximum load on instrument tray" indicating label The maximum load of accessories on the instrument tray must not exceed 13 kg. Version 11.0 Page 34...
Safety Measures S7 / OPMI Sensera Labels on S7 ceiling mount 1 "Risk of crushing" label Fingers may be pinched between the carrier and suspension arms. Do not touch this area while the surgical microscope is being moved. 2 "Maximum load" warning label The maximum load on the suspension arm must not exceed 14 kg.
Page 38
Manufacturer (company name) – Your options to contact the device manufacturer, i.e., phone number, fax number and email address of the local contact of the national Carl Zeiss sales organization. XXXX XXXXXX – SIP No. A unique identification number assigned to your device.
Page 39
S7 / OPMI Sensera Safety Measures 12 Rating label (for S7 ceiling mount with lifting column) The rating label provides the following information: – Manufacturer's symbol – Manufacturer (company name) Carl Zeiss Meditec AG – Manufacturer's address Goeschwitzer Strasse 51-52 07745 Jena, Germany –...
Safety Measures S7 / OPMI Sensera Additional label on S7 ceiling mount with lifting column 1 "Lifting column operation" indicating sign Switch for moving the vertical position upward or downward. Version 11.0 Page 40 G-30-1434-en...
Safety Measures S7 / OPMI Sensera Labels on the light sources of the suspension systems Xenon light source 1 "Warning of a hazard area" indicating label The Superlux light source is equipped with an integrated xenon illumina- tion system. The xenon illumination must not be used for ophthalmic sur- gery.
Page 44
Safety Measures S7 / OPMI Sensera Halogen light source 1 "Filter selector for halogen light source" This label indicates the position of the filter wheel. 2 "Instructions for Use" label Observe the Instructions for Use or accompanying documents. Version 11.0...
Page 45
System Overview System Overview Sensera surgical microscope on S7 suspension systems ....48 Features ....................48 System components ..................50 OPMI Sensera surgical microscope ..........52 Design .......................52 Controls, displays, connections ..............54 Tiltable binocular tube and eyepieces ............60 S7 floor stand ..................64 Design .......................64...
Page 46
System Overview S7 / OPMI Sensera Control panel and menu overview ...........98 Structure of the menus ................100 Foot control panel (option) ............122 Features....................122 Design ....................122 Foot control panel with 14 functions ............124 Foot control panel with 8 functions ............125 Version 11.0...
Page 47
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 47...
NTSC video system. Suspension systems The S7 floor stand is a suspension system for Zeiss surgical microscopes. It comprises a suspension arm, a carrier arm, a stand column and a stand base. A maneuvering handle mounted on the stand column is used to move the stand and to attach the foot control panel.
Page 49
S7 / OPMI Sensera System Overview The suspension arm of all types of suspension system permits almost effortless positioning of the surgical microscope. The spring force of the suspension arm can be varied in a range from 6 to 14 kg, permitting reliable balancing of the microscope even with accessory equipment attached.
1 S7 ceiling mount with lifting column 2 S7 ceiling mount 3 S7 floor stand 4 OPMI Sensera surgical microscope with 180° tiltable tube and magnetic eyepieces The overall system comprises a suspension system and a microscope. All op- tions described in these Instructions for Use can be ordered separately.
Page 51
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 51...
• For this reason, hold the surgical microscope tightly at its handgrips be- fore loosening the friction adjustment screw (see page 66) on the sus- pension arm. 7 Motorized fine focusing (option), recommended for the use of OPMI Sensera with laser micromanipulators. Version 11.0 Page 52 G-30-1434-en...
Page 53
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 53...
System Overview S7 / OPMI Sensera Controls, displays, connections 1 Friction adjustment for vertical axis Use this knob to adjust the friction for the vertical axis as required. 2 Balance setting of the lateral tilt motion Use this knob to adjust the balance setting of the lateral tilt motion.
Page 55
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 55...
Page 56
System Overview S7 / OPMI Sensera 8 Dust cover 9 Handgrips (removable) for moving the surgical microscope and triggering the motorized OPMI functions (see points 14 to 17). 10 Locking screw for optimum positioning of the handgrips. 11 Zoom wheel (emergency operation) If the motorized functions fail, you can manually adjust the magnification after removing the plastic cover from the zoom wheel.
Page 57
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 57...
Page 58
System Overview S7 / OPMI Sensera 15 Release buttons, freely programmable In combination with an S7 suspension system, specific functions of the suspension system can be assigned to these buttons (e.g.: focus, zoom, brightness control, still and video camera release, etc.). Configuring these buttons is described on page 108 (Handgrip Settings menu).
Page 59
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 59...
System Overview S7 / OPMI Sensera Tiltable binocular tube and eyepieces 180° tiltable tube 1 PD adjustment knob The correct position has been set when the two eyepiece images merge into one. 2 180° tiltable tube 3 Eyepiece tube Version 11.0...
Page 61
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 61...
Page 62
System Overview S7 / OPMI Sensera Widefield eyepieces with magnetic coupling NOTE Eyepieces have a magnetic field! Please note that the usual rules for the handling of magnets must be ob- served for eyepieces removed from the tube: • Do not place the eyepiece near instruments which may be magnetizable.
Page 63
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 63...
System Overview S7 / OPMI Sensera S7 floor stand Design 1 Suspension arm 2 Lamp housing (either xenon or halogen illumination) 3 Carrier arm 4 Control panel 5 Connector panel 6 Handgrip 7 Cable supports (2x, on the right and left of the stand column)
Page 65
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 65...
System Overview S7 / OPMI Sensera Suspension arm 1 Lamp housing with halogen or xenon illumination The suspension system is equipped with an illumination system for light guides. Each lamp housing contains a backup lamp. When a halogen illu- mination system is used, the backup lamp will be automatically swung in if the first lamp fails.
Page 67
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 67...
System Overview S7 / OPMI Sensera Carrier arm, stand column, stand base 1 Control and display panel The control and display panel permits you to control all electrical functions of the S7 floor stand and the surgical microscope. 2 Friction Knob for adjusting the friction of the carrier arm's swivel movement.
Page 69
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 69...
Connector panel of S7 floor stand CAUTION Danger! Electrical voltage! • Only connect accessories and medical devices intended by Carl Zeiss for use with this system to power outlet socket (5). When connecting other devices, make sure that safety is guaranteed regarding admissible touch currents and ground leakage currents as per IEC 60601-1:2005.
Page 71
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 71...
Page 72
System Overview S7 / OPMI Sensera Connectors for USB option 1 USB socket for USB memory stick Note: Only use sticks conforming to the USB-2.0 specification. 2 DV output port (FireWire) Digital, compressed video signal for recording on a digital video recorder or PC.
Page 73
S7 / OPMI Sensera System Overview 1, 2 Version 11.0 G-30-1434-en Page 73...
The instrument tray (1) can carry a maximum of 13 kg. The tray has been de- signed, for example, for mounting MediLive Trio by Carl Zeiss. Further acces- sories can be installed on the instrument tray using the strap provided.
Page 75
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 75...
System Overview S7 / OPMI Sensera S7 ceiling mount Design 1 Ceiling flange 2 Stand column 3 Control panel 4 Connector panel 5 Lamp housing (either xenon or halogen illumination) 6 Suspension arm 7 Carrier arm Version 11.0 Page 76...
Page 77
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 77...
System Overview S7 / OPMI Sensera Carrier and suspension arms 1 Lamp housing with halogen or xenon illumination The suspension system is equipped with an illumination system for light guides. Each lamp housing contains a backup lamp. When a halogen illu- mination system is used, the backup lamp will be automatically swung in if the first lamp fails.
Page 79
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 79...
System Overview S7 / OPMI Sensera Connector panel of S7 ceiling mount 1 Opening for the system cable from an external video camera 2 Remote socket for controlling external devices (e.g. MEDIALINK 100) with a max. breaking capacity of 24 V/0.5 A.
Page 81
S7 / OPMI Sensera System Overview 5, 6 Version 11.0 G-30-1434-en Page 81...
System Overview S7 / OPMI Sensera Ceiling flange with lifting column 1 Lifting column 2 Selector switch for setting the optimum viewing height of the surgical microscope or for downward movement into the working position. As long as you keep the selector switch in the turned position, the lifting column moves downward (3) or upward (4), depending on the switch po- sition.
Page 85
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 85...
System Overview S7 / OPMI Sensera Carrier and suspension arms 1 Lamp housing with halogen or xenon illumination The suspension system is equipped with an illumination system for light guides. Each lamp housing contains a backup lamp. When a halogen illu- mination system is used, the backup lamp will be automatically swung in if the first lamp fails.
Page 87
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 87...
System Overview S7 / OPMI Sensera Connector panel of S7 ceiling mount with lifting column 1 Opening for the system cable from an external video camera 2 Remote socket for controlling external devices (e.g. MEDIALINK 100) with a max. breaking capacity of 24 V/0.5 A.
Page 89
S7 / OPMI Sensera System Overview 6, 7 Version 11.0 G-30-1434-en Page 89...
System Overview S7 / OPMI Sensera Illumination system Two different types of illumination are available for the suspension system. 1 Halogen illumination The halogen light source is equipped with an illumination system for fiber illumination. The lamp housing contains a backup lamp which is automat- ically swung into the illumination beam path when the first lamp fails.
Page 91
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 91...
System Overview S7 / OPMI Sensera Halogen illumination system The illumination system has been designed for fiber illumination. The lamp housing contains a backup lamp which automatically swings into the illumi- nation beam path when the first lamp fails. 1 Lamp module...
Page 93
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 93...
System Overview S7 / OPMI Sensera Xenon illumination system CAUTION Risk of injury caused by xenon light in the facial area Surgery using xenon light in the facial area may seriously harm the patient's eyes. • The patient's eyes must not be exposed to xenon light.
Page 95
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 95...
Page 96
System Overview S7 / OPMI Sensera When inserting a new lamp module, make sure that the knob (2) is set to “1". If the first lamp fails, switch to the second lamp in logical sequence. 3 Indicator: Backup lamp is in use When the segment in knob (2) lights up, the backup lamp is in use.
Page 97
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 97...
System Overview S7 / OPMI Sensera Control panel and menu overview The control panel consists of a graphic display (1) with control keys (3). All functions can be interactively controlled via menus. The display shows the se- lected functions and settings. Using control buttons (3) you can move through the menus and activate a selection field.
Page 99
S7 / OPMI Sensera System Overview Version ..300 mm Please wait ... USER 1 USER 2 NONE USER 3 NONE 50 % 50 % 300 mm 50 % 50 % 4,9 x 300 mm 20 % Close Standard Press Enter...
After you have switched on the system, the software is loaded. The growing size of the ZEISS logo indicates that this loading process is still in progress. The release number is displayed in the top left corner. The system then performs a self-test.
Page 101
S7 / OPMI Sensera System Overview Edit User menu Three different users or applications (e.g. larynx, nose, ear) and their profiles can be stored in the User menu. The Edit User menu allows you to enter your personal user name. When Edit icon (1) after the user name is activated, the Dr.
Page 102
System Overview S7 / OPMI Sensera Surgery menu The Surgery menu is constantly displayed during operation. The menu in- cludes the icons (from left to right) for the following menus: Zoom / Focus / 4.9x 300 mm Illumination. Bottom: User / Options / Video und Balance. The displays below...
Page 103
S7 / OPMI Sensera System Overview Zoom menu You can adjust the settings for the following functions in the Zoom menu: Speed: The zoom speed from minimum to maximum speed from 5% to 100% in steps of 5%. Start at: The zoom start value.
Page 104
System Overview S7 / OPMI Sensera Focus menu You can adjust the settings for the following functions in the Focus menu: Zoom Link: The Zoom Link function allows the focusing speed to be coupled to the cur- 50 % 300 mm rent magnification.
Page 105
S7 / OPMI Sensera System Overview Illumination menu You can adjust the settings for the following functions in the Illumination menu: Zoom Link Intensity: 50 % 50 % The illumination intensity from minimum to maximum brightness from 5 % to 100 % in steps of 5 %. This setting can be adjusted for the xenon il- lumination and the halogen illumination.
Page 106
System Overview S7 / OPMI Sensera Options menu Activate the Options icon and press the Enter button to access the Options menu. The Options menu comprises the following submenus (from the top): – Image Capture menu – Handgrip Settings menu –...
Page 107
S7 / OPMI Sensera System Overview Index option The Index option permits you to access the graphic index. Saved images are displayed as a graphic index with 12 distortion-free, demag- nified images (thumbnails) per screen page. For the Index functions, see page 188.
Page 108
System Overview S7 / OPMI Sensera Handgrip Settings menu This menu allows you to assign one of the listed functions to each of the four buttons (A, B, C, D) for every user or application: – Light on/off: switching the illumination on and off.
Page 109
S7 / OPMI Sensera System Overview Foot Control menu Your suspension system permits the connection of a foot control panel (op- tion). This menu allows you to assign one of the listed functions to each of the buttons C and D on the foot control panel for every user: –...
Page 110
System Overview S7 / OPMI Sensera Binocular Settings menu If you select Binocular Setting in the Options menu, the following three setting options are available: – Selection of the eyepieces used (10x or 12.5x) – Selection of the tube used (170 mm or 200 mm) –...
Page 111
S7 / OPMI Sensera System Overview SpeedFokus autofocus option The SpeedFokus autofocus option permits you to select and automatically focus on a feature in the ROI (region of interest, measuring field) in the image of the surgical field. The SpeedFokus autofocus function is based on the evaluation of the video image contrast: during focusing of the microscope's Varioskop optics, Speed- Fokus continuously determines the video image contrast in the selected ROI.
Page 112
System Overview S7 / OPMI Sensera When you start your system, the ROI is always positioned at the center of the field of view. The ROI is visualized as a square in the graphic display. Use the arrow buttons to shift the ROI.
Page 113
S7 / OPMI Sensera System Overview Video Settings menu (option) The Video Settings menu permits you to configure the integrated 1CCD video camera in the Config submenu, and to enter or edit text using the OSD func- tion. Activate the Video icon and press the enter button to access the Video Set-...
Page 114
System Overview S7 / OPMI Sensera Freeze/Live setting The freeze button permits you to freeze the current live video image and to Press Enter Auto Auto White Balance display it as a full-screen freeze frame on the video monitor. During this time,...
Page 115
S7 / OPMI Sensera System Overview OSD menu (On Screen Display) Press Enter Auto The OSD function allows you display freely selectable texts, a ruler and micro- White balance Auto Freeze/Live Live Ready Ready scope parameters such as magnification and working distance on the video...
Page 116
System Overview S7 / OPMI Sensera If "Ruler" selection field (4) has been activated, a ruler including a dimension is displayed in the bottom left corner of the screen. The ruler and the dimen- sion specified automatically change with the zoom and focus settings. The nu- merical value of the dimension displayed can be 2.5 / 5 / 10 / or 20 mm.
Page 117
S7 / OPMI Sensera System Overview Balance menu The Balance menu has been designed as a user prompt and consists of six im- ages explaining the balancing procedure. You can scroll through the Balance menu using the arrow buttons ( ).
Page 118
System Overview S7 / OPMI Sensera Laser menu The User Selection menu permits you to access the laser selection menu via the respective Edit button using the arrow button ( ). Here, the working dis- tance (focus) of the Varioskop in the surgical microscope is adjusted to the...
Page 119
S7 / OPMI Sensera System Overview When you select "TwinER", you access the "Focus TwinER" menu (option). The "Focus TwinER" menu only contains the following displays: – Internal Focus is fixed – external focus active The Varioskop automatically adjusts to the focal length of the TwinER laser and switches off.
Page 120
System Overview S7 / OPMI Sensera Malfunctions (Error message menu) This window appears on your display when a malfunction or incorrect opera- tion has occurred. At the same time, this is also indicated by three beeps. In addition, the type of malfunction present and a code number are displayed in field (1).
Page 121
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 121...
System Overview S7 / OPMI Sensera Foot control panel (option) Features The foot control panel permits you to operate various functions of a suspen- sion system or surgical microscope. The assignment of functions to the con- trols on the foot control panel is described on the next page. Only functions provided in the system configuration used (suspension system, surgical micro- scope) can be controlled.
Page 123
S7 / OPMI Sensera System Overview Version 11.0 G-30-1434-en Page 123...
System Overview S7 / OPMI Sensera Foot control panel with 14 functions The illustration shows the standard assignment of functions to the foot con- trol panel with 14 functions. The focus/zoom assignment can be changed by our service staff, on request.
S7 / OPMI Sensera System Overview Foot control panel with 8 functions The illustration shows the standard assignment of functions to the foot con- trol panel with 8 functions. The focus/zoom assignment can be changed by our service staff, on request.
Page 126
System Overview S7 / OPMI Sensera Version 11.0 Page 126 G-30-1434-en...
Page 127
S7 / OPMI Sensera Preparations for use Preparations for use Attaching the equipment...............130 Mounting the tube and eyepieces............130 Connections................... 132 Connecting the strain relief device ............132 Connecting the suspension system - S7 floor stand........134 Connecting the suspension system - S7 ceiling mount with rigid column ....................136...
Page 128
Preparations for use S7 / OPMI Sensera Applying drapes..................156 Surgical microscope with a laser micromanipulator.......157 Configuring the camera settings (config) ........159 Configuring Image Capture............164 Relocating the S7 floor stand ............171 Version 11.0 Page 128 G-30-1434-en...
Page 129
S7 / OPMI Sensera Preparations for use Version 11.0 G-30-1434-en Page 129...
Preparations for use S7 / OPMI Sensera Attaching the equipment Mounting the tube and eyepieces CAUTION Risk of injury caused by lowering of the surgical microscope or accessories falling down! • Never change modules and accessories during a surgical procedure or above the patient.
Page 131
S7 / OPMI Sensera Preparations for use Version 11.0 G-30-1434-en Page 131...
Preparations for use S7 / OPMI Sensera Connections Connecting the strain relief device You can secure the power plug and the multipoint connector of the switching component against inadvertent loosening by installing the two cables in strain relief device (1).
Page 133
S7 / OPMI Sensera Preparations for use 320 mm Version 11.0 G-30-1434-en Page 133...
Preparations for use S7 / OPMI Sensera Connecting the suspension system - S7 floor stand • Check the voltage indicated at (5). The suspension system is set at the factory to the rated voltage used in the country of destination. The rated voltage indicated at window (5) must correspond to the rated voltage available on the site of installation.
Page 135
S7 / OPMI Sensera Preparations for use Version 11.0 G-30-1434-en Page 135...
Preparations for use S7 / OPMI Sensera Connecting the suspension system - S7 ceiling mount with rigid column • Check the voltage indicated at (2). The suspension system is set at the factory to the rated voltage used in the country of destination.
Page 137
S7 / OPMI Sensera Preparations for use Version 11.0 G-30-1434-en Page 137...
Preparations for use S7 / OPMI Sensera Connecting the suspension system - S7 ceiling mount with lifting column • Check the voltage indicated at (2). The suspension system is set at the factory to the rated voltage used in the country of destination.
Page 139
S7 / OPMI Sensera Preparations for use Version 11.0 G-30-1434-en Page 139...
Preparations for use S7 / OPMI Sensera Connecting the video monitor and the MEDIALINK 100 (option) NOTE Use of video signal isolation When video signals are led out of the OR, e.g. for a conference, this may cause excessive leakage and contact currents.
Page 141
S7 / OPMI Sensera Preparations for use (Option) VIDEO IN VIDEO IN BNC(FBAS) MEDIALINK 100 BNC(FBAS) Remote Version 11.0 G-30-1434-en Page 141...
Preparations for use S7 / OPMI Sensera Adjusting the system - S7 floor stand Balancing the suspension arm CAUTION Risk of crushing - mind your fingers! Fingers may be crushed between the carrier and suspension arms. • Never touch the area between the suspension and carrier arms while you are performing the balancing procedure or moving the suspension arm.
Page 143
S7 / OPMI Sensera Preparations for use • Firmly retighten locking lever (2) for the upward/downward movement. • Lower the surgical microscope again to the lowest limit stop and check the safety distance. – The limit of downward travel is only effective in the lower part of the sus- pension arm's vertical movement range (starting from the arm's horizontal position).
Preparations for use S7 / OPMI Sensera Adjusting the system - S7 ceiling mount with rigid column Balancing the suspension arm CAUTION Risk of crushing - mind your fingers! Fingers may be crushed between the carrier and suspension arms. • Never touch the area between the suspension and carrier arms while you are performing the balancing procedure or moving the suspension arm.
Page 145
S7 / OPMI Sensera Preparations for use • Firmly retighten locking lever (2) for the upward/downward movement. • Lower the surgical microscope again to the lowest limit stop and check the safety distance. – The limit of downward travel is only effective in the lower part of the sus- pension arm's vertical movement range (starting from the arm's horizontal position).
Preparations for use S7 / OPMI Sensera Adjusting the system - S7 ceiling mount with lifting column Setting an ergonomic working height CAUTION Risk of injury caused by lifting column Activating the lifting column may cause injuries. • Do not activate the lifting column during surgery.
Page 147
S7 / OPMI Sensera Preparations for use Version 11.0 G-30-1434-en Page 147...
Preparations for use S7 / OPMI Sensera Balancing the suspension arm CAUTION Risk of crushing - mind your fingers! Fingers may be crushed between the carrier and suspension arms. • Never touch the area between the suspension and carrier arms while you are performing the balancing procedure or moving the suspension arm.
Page 149
S7 / OPMI Sensera Preparations for use – The limit of downward travel is only effective in the lower part of the sus- pension arm's vertical movement range (starting from the arm's horizontal position). – If you wish to use the suspension arm's full range of vertical movement, lower the suspension arm as far as it will go and then firmly retighten lock- ing lever (2).
Preparations for use S7 / OPMI Sensera Balancing the surgical microscope NOTE Injury to the patient! With an incorrectly balanced system, brake release may lead to uncontrolled movements of the suspension arm. • Use the surgical microscope in perfectly balanced condition only! •...
Page 151
S7 / OPMI Sensera Preparations for use Version 11.0 G-30-1434-en Page 151...
Page 152
Preparations for use S7 / OPMI Sensera Balancing the front-to-back tilt motion NOTE Injury to the patient! With an incorrectly balanced system, brake release may lead to uncontrolled movements of the suspension arm. • Use the surgical microscope in perfectly balanced condition only! •...
Page 153
S7 / OPMI Sensera Preparations for use bservation tube) it may become necessary to correct the balance setting. – If the microscope body tends to tilt forward (in your direction), press key (2). – If the microscope body tends to tilt backward, press key (1).
Preparations for use S7 / OPMI Sensera Adjusting the surgical microscope Notes on adjustment – It is recommended to provide the interpupillary distance as well as the re- fraction values for the different users to adjust the surgical microscope more quickly and to preset the values in the scope of the preparation.
Page 155
S7 / OPMI Sensera Preparations for use • Repeat the entire procedure for the second eyepiece. Now, the microscope is adjusted in such a way that the image is sharply defined over the complete magnification range without any renewed fo- cusing during the magnification setting.
Attaching asepsis caps For sterile use, the system can be equipped with resterilizable products. The asepsis sets available from Carl Zeiss contain caps and hand grips which can be sterilized in autoclaves. For detailed information on sterilization please see the enclosed instructions "Preparation of resterilizable products" for the re- spective asepsis sets.
S7 / OPMI Sensera Preparations for use Surgical microscope with a laser micromanipulator • Please observe the separate Instructions for Use for your micromanipula- tor. Before using the surgical microscope with a laser micromanipulator, you must adjust the focal plane of the microscope (Varioskop) to the focal plane of the micromanipulator.
Page 158
Preparations for use S7 / OPMI Sensera The therapy beam of the laser displays its maximum effect in the focal plane of the micromanipulator, where it burns a spot on the wooden spatula. The wooden spatula must be sharply focused in this position.
S7 / OPMI Sensera Preparations for use Configuring the camera settings (config) The configuration menu permits you to configure the camera settings of the 1 chip video camera. The user program is set by default on delivery. 4.9 x 300 mm 20 % •...
Page 160
Preparations for use S7 / OPMI Sensera Navigation in the Config mode using the control buttons Control buttons Press the control buttons to access the submenu required. To activate the selected submenu, press the enter button. Enter button The enter button permits you to jump on from all submenus (except Chroma).
Page 161
S7 / OPMI Sensera Preparations for use Lightmeter The use of the Lightmeter function will only be meaningful if Brightness has been set to the automatic mode (Auto). You can select between three metering patterns provided by the video camera in the Lightmeter selection field: Integral, small, or large.
Page 162
Preparations for use S7 / OPMI Sensera Chroma submenu The Chroma submenu permits you to adjust the overall color impression pro- vided by the camera. Chroma Use the control buttons for continuous adjustment of the color intensity in a range between 0% and 100%. The value currently set is displayed on the right of the Chroma submenu bar.
Page 163
S7 / OPMI Sensera Preparations for use Factory settings After the selection of this function, all settings of the selected user can be reset to the factory settings. If you select "Factory Settings" (using the Enter button), the display shows "processing"...
Preparations for use S7 / OPMI Sensera Configuring Image Capture Image Capture permits you to save, view and delete image data in the JPEG or TIFF format. The Config menu permits you to configure the Image Capture function. • Select the Options icon in the Surgery menu and press the Enter button.
Page 165
S7 / OPMI Sensera Preparations for use Date and Time submenu The current date and time are displayed in the yyyy-mo-dd_hh-mi-ss format in large numerals on the video monitor. (yyyy = year, mo = month, dd = day, hh = hours, mi = minutes, ss = seconds). The time is constantly updated.
Page 166
Preparations for use S7 / OPMI Sensera Remote Control Here, you can switch the acoustic signal on and off and select the operation mode of the remote switch on the Remote input. Hand grip buttons, buttons on a foot control panel or a separate foot switch can be used as remote control switch.
Page 167
S7 / OPMI Sensera Preparations for use Record Mode Here, you can select the format and the quality in which the video images will be stored. – JPEG / TIFF Files – Still / Moving Subject JPEG / TIFF Files Do you wish to store the video images in JPEG format only or in both JPEG and TIFF formats? Two modes are available.
Page 168
Preparations for use S7 / OPMI Sensera second field is computed by doubling the lines and then displayed on the video monitor. The same applies to the storage process. Delete mode Confirm Delete "On/Off" Would you like to have to expressly confirm the "Delete" command for de-...
Page 169
S7 / OPMI Sensera Preparations for use File Name To extend the file name, you can enter a freely selectable name with up to 14 characters after the date and time. The file name extension is separated from the remaining name (date/time) by an underscore. In the index mode (graphic index) the extension is shown below the file name (date/time).
Page 170
Preparations for use S7 / OPMI Sensera over again. On every system restart, the upper case mode is set automat- ically. The delete key of the virtual keypad always deletes the last character of the file name entered. The OK key brings you to the Configuration main menu. The file name you en- tered is buffered for the time being, and will be saved when you exit the con- figuration main menu, after an appropriate inquiry.
S7 / OPMI Sensera Preparations for use Relocating the S7 floor stand CAUTION Risk of crushing - mind your fingers! Fingers may be crushed in the areas marked with the "risk of crushing" label. • Do not touch these areas while the system is being moved or brought in its working / transport position.
Page 172
Transport over long distances! Over long distances (e.g. removal, return for repair, etc.) the device must al- ways be transported in the original packaging or in special return packaging. • For details, please contact your dealer or the Carl Zeiss Service. Version 11.0...
S7 / OPMI Sensera Operation Operation Checklist for S7 floor stand ............174 Checklist for S7 ceiling mount with rigid column ......178 Checklist for S7 ceiling mount with lifting column ......182 Procedure ..................186 Working with the SpeedFokus autofocus option........187 Display and deletion of images (USB option)..........188...
Operation S7 / OPMI Sensera Checklist for S7 floor stand CAUTION Risk of injury! Incorrect configuration settings may lead to unexpected behavior of the de- vice and may cause injury to the patient. • Always check the settings of the selected user profile before using the system.
Page 175
S7 / OPMI Sensera Operation – If operated for 8 hours/day on 5 days/week, the lamp needs to be ex- changed after 12 weeks maximum. Checking the illumination system Check that the lamp and the backup lamp are intact.
Page 176
Operation S7 / OPMI Sensera Checking the functions of the knobs • Check the proper functioning of focusing knob (3) and filter selector knob (4). Eyepieces / binocular tube Check that binocular tube (2) has been securely mounted. Check that securing screw (1) has been firmly tightened.
Page 177
S7 / OPMI Sensera Operation Limiting downward travel Check that the minimum working distance (height) from the surgical field has been adjusted. Stand base Check that at least three of the locking tabs have been pressed and pre- vent the stand from starting to move by itself.
Operation S7 / OPMI Sensera Checklist for S7 ceiling mount with rigid column CAUTION Risk of injury! Incorrect configuration settings may lead to unexpected behavior of the de- vice and may cause injury to the patient. • Always check the settings of the selected user profile before using the system.
Page 179
S7 / OPMI Sensera Operation – If operated for 8 hours/day on 5 days/week, the lamp needs to be ex- changed after 12 weeks maximum. Checking the illumination system Check that the lamp and the backup lamp are intact.
Page 180
Operation S7 / OPMI Sensera Checking the functions of the knobs • Check the proper functioning of focusing knob (3) and filter selector knob (4). Eyepieces / binocular tube Check that binocular tube (2) has been securely mounted. Check that securing screw (1) has been firmly tightened.
Page 181
S7 / OPMI Sensera Operation Limiting downward travel Check that the minimum working distance (height) from the surgical field has been adjusted. Foot control panel (option) The plug of the foot control panel has been connected. The power switch of suspension system has been switched on.
Operation S7 / OPMI Sensera Checklist for S7 ceiling mount with lifting column CAUTION Risk of injury! Incorrect software settings may lead to unexpected behavior of the device and may cause injuries to the patient. • Always check the selected user profile settings of the device prior to its use.
Page 183
S7 / OPMI Sensera Operation – If operated for 8 hours/day on 5 days/week, the lamp needs to be ex- changed after 12 weeks maximum. Checking the illumination system Check that the lamp and the backup lamp are intact.
Page 184
Operation S7 / OPMI Sensera Checking the functions of the knobs • Check the proper functioning of focusing knob (3) and filter selector knob (4). Eyepieces / binocular tube Check that binocular tube (2) has been securely mounted. Check that securing screw (1) has been firmly tightened.
Page 185
S7 / OPMI Sensera Operation Checking the balance setting – Check that the suspension arm has been properly balanced. When fully balanced, the surgical microscope should remain stationary in all positions within the work area and move neither upward nor downward.
Operation S7 / OPMI Sensera Procedure CAUTION Risk of injury caused by xenon light in the facial area Surgery using xenon light in the facial area may seriously harm the patient's eyes. • The patient's eyes must not be exposed to xenon light.
S7 / OPMI Sensera Operation Working with the SpeedFokus autofocus option The SpeedFokus autofocus option permits you to select an area of special in- terest to you in the surgical field. SpeedFokus focuses on this selected area at the press of a button. Focusing is triggered on the buttons of the handgrip or foot control panel as configured in the Options menu.
Operation S7 / OPMI Sensera Display and deletion of images (USB option) Index menu After selection of the Index menu, the display shown on the left appears. At the same time, the thumbnails are displayed on the monitor. Use the arrow keys to navigate in the thumbnails. When you press OK, the program goes to the next higher level.
Page 189
S7 / OPMI Sensera Operation Full Screen / Delete Image If you press the Enter button for more than 2 seconds, the display shown on the left appears: Delete image is deleted Next next image is displayed on the monitor...
S7 / OPMI Sensera What to do in the event of malfunctions What to do in the event of malfunctions Failure of main functions ...............192 Failure of the focusing function ...............192 Failure of the zoom function..............194 Failure of the halogen lamp ..............196 Lamp failure of the xenon illumination.............198...
What to do in the event of malfunctions S7 / OPMI Sensera Failure of main functions NOTE Risk of injury caused by failure of main functions! In case one of the main functions (focusing, zoom) is no longer available or does not function correctly, deactivate the electronic components and pro- ceed in manual mode.
Page 193
S7 / OPMI Sensera What to do in the event of malfunctions Version 11.0 G-30-1434-en Page 193...
What to do in the event of malfunctions S7 / OPMI Sensera Failure of the zoom function Motor running constantly • Bring the suspension arm into its horizontal position and tighten locking knob (4) firmly. • Loosen and remove knurled nut (1).
Page 195
S7 / OPMI Sensera What to do in the event of malfunctions Version 11.0 G-30-1434-en Page 195...
What to do in the event of malfunctions S7 / OPMI Sensera Failure of the halogen lamp NOTE Overheating of the lamp module! If ventilation grids are covered e.g. by drapes, this may lead to overheating of the lamp housing and deactivation of the lamp.
Page 197
S7 / OPMI Sensera What to do in the event of malfunctions Version 11.0 G-30-1434-en Page 197...
What to do in the event of malfunctions S7 / OPMI Sensera Lamp failure of the xenon illumination CAUTION Risk of injury caused by lamp rupture! Lamp rupture (audible as a loud bang) may lead to jamming of the lamp module and/or failure of the electronics modules.
Page 199
S7 / OPMI Sensera What to do in the event of malfunctions Version 11.0 G-30-1434-en Page 199...
Page 200
What to do in the event of malfunctions S7 / OPMI Sensera • Reset the service hour counter to “0". Use a pointed object and press it into the recess of reset button (1). • Switch the suspension system back on again at the power switch.
Failure of the SpeedFokus autofocus option with otherwise correct function of the OPMI Sensera / S7 system: You can continue using your OPMI Sensera / S7 system, but you can no longer benefit from the SpeedFokus autofocus option until it has been repaired by the service staff.
What to do in the event of malfunctions S7 / OPMI Sensera Causes of malfunctions and remedies For your safety • This system is a high-grade technological product. To ensure optimum performance and safe working order, we recommend having it checked by our service representative as part of regular scheduled maintenance.
Page 203
S7 / OPMI Sensera What to do in the event of malfunctions S7 suspension system Problem Possible cause Remedy No function at all. Line power failure. Contact in-house electrician. Suspension system not Switch on the system at its page 98 switched on.
Page 204
What to do in the event of malfunctions S7 / OPMI Sensera Problem Possible cause Remedy No video image, only color Camera head not connected. Connect the camera head. page 140 bars. Image with color tinge. No white balance Performing the white bal-...
Page 205
S7 / OPMI Sensera What to do in the event of malfunctions Problem Possible cause Remedy Brightness of CCU set too Increase brightness of CCU. low. Surgical field illumination too Increase brightness of illumi- page 105 dark. nation. Image too bright / Monitor not correctly set.
Page 206
What to do in the event of malfunctions S7 / OPMI Sensera Problem Possible cause Remedy Poor focusing result or none Focusing of the video camera • Focus the microscope im- at all. not matched to focusing of age, then the video im- the surgical microscope.
Page 207
S7 / OPMI Sensera Care and Maintenance Care and Maintenance Care of the device................208 Cleaning ....................208 Sterilization .....................209 Disinfection .....................210 System maintenance..............211 Maintenance intervals................211 Safety check ....................212 Replacing the halogen lamp ..............214 Changing the xenon lamp module............216 Spare parts ....................218 Disposal....................219...
• Remove dust from the optical surfaces using a squeeze blower or a clean, grease-free brush. For the regular cleaning of objective lenses and eyepieces of the surgical mi- croscope, we recommend the optics cleaning set available from Carl Zeiss. For the catalog number, please see the section "Device data". Prevention of fogging To keep the objective lenses from fogging, we recommend using an anti-fog- ging agent.
• Make sure that the operating staff is briefed about the sterile operation of the system. The asepsis sets available from Carl Zeiss contain rubber caps and hand grips which can be sterilized in autoclaves. For detailed information on sterilization please see the enclosed instructions "Preparation of resterilizable products"...
Care and Maintenance S7 / OPMI Sensera Disinfection NOTE Risk of damaging the paint on the device! • Use a disinfectant based on aldehyde or alcohol. The addition of quater- nary compounds is acceptable. To avoid damaging the surfaces, do not use any disinfecting components other than those specified below.
All maintenance work not detailed here requires system-specific expert know- how. It is frequently necessary to open the system for this work. Please contact your local Carl Zeiss service representative in due time for the performance of these maintenance activities.
Care and Maintenance S7 / OPMI Sensera Safety check CAUTION Risk of injury! Cables and plug connections may come loose and parts of the housing may become live. • Make sure that the regular technical safety checks required for this sys- tem in accordance with the applicable national regulations are performed on schedule and to the stipulated extent.
Page 213
S7 / OPMI Sensera Care and Maintenance Version 11.0 G-30-1434-en Page 213...
Care and Maintenance S7 / OPMI Sensera Replacing the halogen lamp NOTE Risk of burns caused by hot lamp! If you change the lamp shortly after it has failed, the lamp will still be very hot. • Allow the lamp to cool down before changing it.
Page 215
S7 / OPMI Sensera Care and Maintenance Version 11.0 G-30-1434-en Page 215...
Care and Maintenance S7 / OPMI Sensera Changing the xenon lamp module CAUTION Risk of injury caused by lamp rupture! Lamp rupture (audible as a loud bang) may lead to jamming of the lamp module and/or failure of the electronics modules.
Page 217
S7 / OPMI Sensera Care and Maintenance • Pack the old lamp module (5) in the transport packaging of the new lamp module. Fill in the enclosed return card and send the old lamp module to the nearest Zeiss service representative.
Care and Maintenance S7 / OPMI Sensera Spare parts Asepsis sets Description Cat. No. Asepsis caps, 27 mm (pack of 6) 305810-9008-000 Asepsis caps, 22 mm (pack of 6) 305810-9001-000 Asepsis caps for 180° tiltable tube (pack of 6) 305810-9003-000...
S7 / OPMI Sensera Care and Maintenance Disposal User information on the disposal of electrical and electronic devices This symbol means that the product must not be disposed of as normal do- mestic waste. The correct disposal of electrical or electronic devices helps to protect the en-...
Page 220
Care and Maintenance S7 / OPMI Sensera Version 11.0 Page 220 G-30-1434-en...
Page 221
Electromagnetic immunity for ME equipment and ME systems ....238 Electromagnetic immunity for non-life-supporting ME equipment and ME systems..................240 Recommended safety distances between portable and mobile RF communication equipment and the S7 / OPMI Sensera .......242 System combinations..............243 Version 11.0 G-30-1434-en...
System Data S7 / OPMI Sensera Technical data Surgical microscope Optics Component Features Magnification Motorized zoom system, zoom ratio 1:6, magnification factor γ = 0.4x - 2.4x. Total magnification 2.4x - 14.6x at a working distance of 200 mm (10x eyepiece) 1.5x - 8.7x at a working distance of 415 mm...
Page 223
S7 / OPMI Sensera System Data Optionally available Motorized fine focus Focusing range ±10 mm (Recommended when laser micromanipulators are used.) Version 11.0 G-30-1434-en Page 223...
System Data S7 / OPMI Sensera Suspension systems Electrical data (of all S7 suspension systems) Component Features Power connection Only connect the floor stand to wall outlets which are provided with a properly connected protective ground conductor. Rated voltage 115 V∼ (100...125 V~ ±10%) 230 V~ (220...240 V~ ±10%)
Page 225
S7 / OPMI Sensera System Data S7 floor stand Mechanics Component Features Carrier arm Length: 450 mm Swivel angle: 360° Suspension arm Length: 865 mm Swivel angle: 296° Vertical lift: -320 mm / +400 mm Coupling Swivel angle: 350° Stand height...
Page 226
System Data S7 / OPMI Sensera S7 ceiling mount Mechanics Component Features Carrier arm Length: 450 mm Swivel angle: 296° Suspension arm Length: 865 mm Swivel angle: 296° Vertical lift: -320 mm / +400 mm Coupling Swivel angle: 350° Admissible max. load...
Page 227
S7 / OPMI Sensera System Data S7 ceiling mount with lifting column Mechanics Component Features Carrier arm Length: 450 mm Swivel angle: 296° Suspension arm Length: 865 mm Swivel angle: 296° Vertical lift: -320 mm / +400 mm Coupling Swivel angle: 350°...
Page 228
System Data S7 / OPMI Sensera Technical data of lifting column Rated voltage 24 VDC Current consumption Vertical lift, total 350 mm Lifting force 2000 N Lifting speed 5.5 mm/sec Interval (operation / pause) 1 min / 9 min Version 11.0...
S7 / OPMI Sensera System Data Integrated 1 CCD PAL video camera (option) Component Features Video focal length 48mm with 1/4 in. CCD image sensor Optical separation 80% observer / 20% camera Image sensor 1/4 in. interline transfer CCD image sensor...
S7 / OPMI Sensera Ordering data Only operate the system with the accessories included in the delivery package and approved by Carl Zeiss. You will find the contact responsible for orders in your country on this website: www.meditec.zeiss.com The product configurations, accessories and spare parts that are applicable for authorized Carl Zeiss partners are stipulated by contract.
Page 231
S7 / OPMI Sensera System Data Eyepieces Description Cat. No. 10x push-in widefield eyepiece 305542-0000-000 12.5x push-in widefield eyepiece, asph. 305543-9901-000 The eyepieces are also available with a focusing aid (1 eyepiece with "format" reticle). Suspension systems Description Cat. No.
Page 232
System Data S7 / OPMI Sensera Further accessories Description Cat. No. Instrument tray for S7 floor stand 000000-1352-922 Wall socket for S7 ceiling mount 000000-1351-565 Adapter with double iris diaphragm (for increasing 303354-0000-000 the depth of field) Dovetail mount / adapter plate for ext. compo- 303360-9903-000 nents (e.g.
Page 233
S7 / OPMI Sensera System Data Asepsis Description Cat. No. 22 mm asepsis caps, pack of 6 305810-9001-000 27 mm asepsis caps, pack of 6 305810-9008-000 Asepsis caps PD 180° tiltable tube, pack of 6 305810-9003-000 Zeiss sterile drapes, type 70,...
System Data S7 / OPMI Sensera Regulatory information S7 / OPMI Sensera Electrical safety The medical device complies with the requirements according to – IEC 60601-1:2005, – IEC 601-1:1988+A1:1991+A2:1995, – CAN/CSA-C22.2 No. 601.1-M90. Classification according to level of protection against electrical shock: Protection class 1 Protection rating (acc.
S7 / OPMI Sensera System Data Ambient conditions For operation Feature Permissible range Temperature + 10 °C ... + 40 °C Rel. Humidity 30 % ... 75 % Air pressure 700 hPa ... 1060 hPa For transportation and storage Feature...
When using the device, the EMC precautions specified below must be ob- served. • Only use spare parts approved by Carl Zeiss for this device. • Do not use any portable or mobile RF communication equipment in the vicinity of the device as this may impair the device's function.
The S7 / OPMI Sensera is intended for operation in an electromagnetic envi- ronment as specified below. The customer or the user of S7 / OPMI Sensera is responsible for ensuring that the device is operated in such an environment.
S7 / OPMI Sensera Electromagnetic immunity for ME equipment and ME systems The S7 / OPMI Sensera is intended for operation in an electromagnetic envi- ronment as specified below. The customer or the user of the S7 / OPMI Sensera is responsible for ensuring that the device is operated in such an en- vironment.
Page 239
S7 / OPMI Sensera System Data Immunity IEC 60601 - Compliance level Electromagnetic environment - tests test level guidelines NOTE: U is the AC supply voltage prior to application of the test level. Version 11.0 G-30-1434-en Page 239...
Electromagnetic immunity for non-life-supporting ME equipment and ME systems The S7 / OPMI Sensera is intended for operation in an electromagnetic envi- ronment as specified below. The customer or the user of the S7 / OPMI Sensera is responsible for ensuring that the device is operated in such an en- vironment.
Page 241
S7 / OPMI Sensera System Data Immunity IEC 60601 - Compliance Electromagnetic environment - tests test level level guidelines Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is influ- enced by absorption and reflection by structures, objects and persons.
Recommended safety distances between portable and mobile RF communication equipment and the S7 / OPMI Sensera The S7 / OPMI Sensera is intended for use in an electromagnetic environment in which RF disturbances are controlled. The customer or the user of S7 / OPMI...
16.2, the power outlet socket of this system is a multiple connector that is intended for an ME system. Connecting electrical devices not approved by Carl Zeiss can lead to a reduced level of safety on the ME system. When configuring your ME system (use of the power outlet socket), make sure...
Page 244
System Data S7 / OPMI Sensera Version 11.0 Page 244 G-30-1434-en...
Page 245
S7 / OPMI Sensera Index Index Numerics 1. Freeze / 2. Store .................166 12-digit user name, entering ..............101 180° tiltable tube ..................60 Acoustic signals ..................120 Address of manufacturer ................2 Adjusting the eyepieces ................154 Adjusting the interpupillar distance ............154 Adjusting the surgical microscope ............154 Adjusting the system ..............142, 144, 146...
Page 246
Index S7 / OPMI Sensera Ceiling anchor ...................12 Ceiling flange ..................76, 82 Ceiling mount with lifting column .............22 Checklist for S7 ceiling mount with lifting column ........182 Checklist for S7 ceiling mount with rigid column ........178 Checklist for S7 floor stand ..............174 Chroma submenu ................162, 165...
Page 247
S7 / OPMI Sensera Index Error message menu ................120 Eyecups ....................62 Eyepieces, technical data ..............222, 223 Factory settings ..................163 Failure of main functions ................192 Failure of the video autofocus option .............201 Failure of the zoom function ..............194 Field of application ..................7 File Name ....................169...
Page 248
Index S7 / OPMI Sensera Illuminated-field diameter ...............222 Illumination intensity ................105 Illumination menu ...................105 Illumination system ...................90 Image Capture, configuration ..............164 Image Orientation ...................162 Image sensor ..................229 Images per burning process ..............168 Index function (USB option) ..............188 Index menu ....................188 Index menu (USB option) ................107...
Page 249
S7 / OPMI Sensera Index Magnification ..................222 Magnification factor ...............222, 223 Main beam path ..................112 Malfunctions ..................120 Manual activation of the backup lamp ..........92, 94 Manufacturer .....................2 Manufacturer’s address ................2 Measuring field (ROI) ................111 MEDIALINK 100, connection of ..............140 Menus, relationship ..................98 Mobile phones ..................236...
Page 250
Index S7 / OPMI Sensera Record Mode ..................167 Regulatory information ................234 Relocating the system ................171 Remote Control ..................166 Remote socket ................70, 80, 88 Remote Sound ..................166 Remote switch on Remote input .............166 Replacing the halogen lamp ..............214 Risk of burn injuries caused by high illumination intensity ......15 ROI (region of interest, measuring field) ..........111...
Page 251
S7 / OPMI Sensera Index Surgery menu ..................102 Surgical microscope with a laser micromanipulator .........157 Suspension arm ............64, 66, 76, 78, 82, 86 Suspension arm, limit of downward travel ......142, 144, 148 Suspension systems ................224 Switching to the backup lamp ............19, 20, 198 Symbols and labels on the device .............24...
Page 252
Index S7 / OPMI Sensera Weight ....................222 White balance ..................113 Widefield eyepieces with magnetic coupling ..........62 Xenon illumination system ..............20, 94 Xenon lamp module, changing ...............216 Y/C video port ................72, 80, 88 Zero position ..................24, 54 Zoom function, failure ................194 Zoom Link ..................104, 105...
Page 253
S7 / OPMI Sensera (Blank page, for your notes . . .) Version 11.0 G-30-1434-en Page 253...