If the problem is AC power-line transients (often from large electrical motors near the Interface 1010):
Try plugging your Interface 1010 into a different AC power branch circuit.
Plug your Reference into a power-line surge suppressor. Inexpensive surge suppressors are now
generally available because of their use with computer equipment.
Contact Gamry Instruments, Inc. if these measures do not solve the problem.

CE Compliance

The European Community has instituted standards limiting radio-frequency interference emitted by electronic
devices, setting limits for susceptibility of apparatus to RF energy and transient events, and mandating safety
requirements. Gamry Instruments, Inc. has designed and tested the Interface 1010 to comply with these
standards.
The relevant CE regulations include EN 61010 and EN 61326.

RoHS Compliance

The Interface 1010 has been built using lead-free components and lead-free solder. It is in compliance with the
European RoHS initiative.