DigiBlock – User Manual
Installation
Site Requirements
Ambient Temperature Range: 10°C to 35°C (50°F to 94°F)
Relative Humidity Range : 10% to 80% (non-condensing)
Never place the unit in a location where excessive heat, moisture, or corrosive materials are present.
Positioning the instrument
Never place the unit in a location where excessive heat, moisture or corrosive materials are present.
Please keep the unit away from installations which release strong electro-magnetic field or high-frequency waves.
The unit should not be placed in a place with excessive corrosive gases such as chlorine, hydrochloric acid,
hydrogen sulphide, sulphurous acid gas, etc.
The installation platform must be horizontal and stable without vibration, capable of supporting the total weight of
the main unit and accessories. A distance at least 15 cm must be kept between the left side of the instrument and
the wall for undisturbed ventilation. Enough space around the unit fan should be left for smooth ventilation.
Do not place the unit in a place exposed to direct sunlight.
Do not place the unit in a dusty environment.
NOTE: improper working environment may significantly affect the measurement performance and
reduce the lifespan of the unit.
www.labtechsrl.com
Page 12 of 25
Need help?
Do you have a question about the DIGIBLOCK ED-S Series and is the answer not in the manual?
Questions and answers