Editing an Application Setup Interface
Edit a soil digestion method in Soil category, heat from room temperature to 100℃ and hold in 100 ℃ until
switch off the unit.
A. In home page, touch Setup to enter method set interface.
B. Enter Method Edit interface, touch Soil and Method1
C. Step 01: Touch the Temp. Text box, input 100 via the digital key. Then touch Time text box to input
-2 which means heat in max. heating speed (Normally, the average heating speed is around
5℃/min).
Step 02: Touch the Temp. text box and input 100 via the digital key. Then touch Time text box to
input 0 which means hold the temperature at previous step setpoint.
www.labtechsrl.com
DigiBlock – User Manual
Page 20 of 25
Need help?
Do you have a question about the DIGIBLOCK ED-S Series and is the answer not in the manual?
Questions and answers