Omron K3HB Manual page 11

Digital indicators k3hb series
Hide thumbs Also See for K3HB:
Table of Contents

Advertisement

■ Input Ranges (Measurement Range and Accuracy)
Input type
Range
K3HB-VLC
A
Load Cell, mV
B
C
D
Note: 1. The accuracy is for an ambient temperature of 23±5 ° C. For all ranges,10% or less of max. input ±0.1% FS.
2. The letters "rdg" mean "reading."
a lc
Input type
Connected
E2
terminals
(mV)
199.99
200.000
150.000
100.000
50.000
0.00
0.00
− 50.00
− 100.00
− 150.00
− 200.00
The area shown in dark shading indicates the factory setting.
■ Load Cell Wiring Example
B
E
E2
E3
E4
B5
E5
B6
E6
Set value
Measurement range
a Ud
0.00 to 199.99 mV
b Ud
0.000 to 19.999 mV
c Ud
±100.00 mV
d Ud
±199.99 mV
b lc
E6
E3
E6
19.999
0.000
+10 V
+IN
−OUT
+OUT
Load cell
+
−IN
0 V
Input impedance
1 M Ω min.
c lc
d lc
E4
E6
E5
E6
199.99
100.00
− 100.00
− 199.99
■ Scaling Example Using Range A
Indicated on the K3HB-V as 0 to 49N in the load cell specifications
(rated load 49N, recommended applied voltage 10 V, rated output 2
mV/V) (See note.)
Display value
49 N
Note: 2 mV/V indicates a load cell output of 2 mV for 1 V applied volt-
age for the rated load (when using a load of 1 N). When the ap-
plied voltage is 10 V, the load cell output is 20 mV (2 mV × 10).
Weighing Indicator
Accuracy
±0.1%rdg ± 1 digit max.
±200 V
±0.1%rdg ± 5 digits max.
±0.1%rdg ± 3 digits max.
±0.1%rdg ± 1 digit max.
Scaling values:
inp.a1=000.00
dsp.a1=00000
inp.a2=020.00
dsp.a2=49000
dp decimal point position: 00.000
Input value
20 mV
K3HB-V
Allowable
instantaneous
overload (30 s)
11