● If the device is used for maritime application, a distance of 5 m to
the compass must be complied with.
4.4

Preparations for use

1. Ensure that the rechargeable battery is charged or that the
batteries (for battery-operated devices) have been inserted.
2. Turn on the device by pressing the
 A self-test is carried out.
When the device is ready for measuring: Wait for the display to
indicate that the device is ready to measure (see the fold-out
page, Figure 2: A).
4.4.1 Before sampling
● The subject must be breathing calmly and steadily.
● The subject should be able to provide the minimum volume of
breath (>1.2 L) required for sampling.
4.5

Changing the measuring mode

By default, the device is set to measuring using the mouthpiece. The
measuring mode must be changed before measuring using the funnel.
The active measuring mode is indicated by a corresponding symbol
(device with mouthpiece or funnel) on the display.
To change the measuring mode:
1. Press the
2. Choose Funnel.
3. Press
to confirm.
✓ The Funnel measuring mode is now set.
The measuring mode remains set until it is changed in the menu or
the device is restarted. Restarting returns the device to the
mouthpiece measuring mode.
®