7. Connect the test unit to the PC with the USB cable.
NOTICE
The power plug must be freely accessible to ensure it
can be easily disconnected.
8. Connect the test unit to the power supply using the
power supply unit.
To disconnect the plug, pull the socket approx.
1 mm away from the test unit.
This releases the lock, allowing the plug to be
removed from the test unit.
5

Use

5.1 Preparations for use

5.1.1 Preparing the test unit
● Open the compressed air supply at the high-
pressure inlet before testing.
5.1.2 Start the test software on the PC
When a test is launched using the test software, the
test unit is switched on automatically. There is no on/off
switch on the test unit.
NOTICE
Do not connect any breathing safety product to the test
unit's low pressure measuring points before launching
the software.
Each time the software is started, a zero adjustment is
automatically performed before the first test.
CAUTION
Risk of crushing!
There is a risk of crushing to fingers when reaching into
the lung during operation.
► Only use a test unit with the test head attached.
1. Access the test software via the Start menu or the
desktop shortcut.
The startup screen appears. The program
components are loaded. If a system error is
detected, this is indicated by an error message.
A loading bar with status information provides
information on the progress of the startup process.
2. When the login screen appears, enter the user
name and password (see "Prior to first use",
page 17).
✓ After starting the program and successfully logging
on, the "Test due list" and "Test" tabs are displayed
in the function area.
Instructions for use
|
Dräger Quaestor 4000/6000/8000
5.1.3 Automatic preparation for use
Dräger recommends checking the software settings
before starting work with the test unit.
When a test is launched using the test software, the test
unit is switched on automatically for usage:
– When the software is started, all sensors are zeroed.
Except for the high-pressure inlet, the connections
must be depressurised at this point.
– The artificial lung is positioned.
When the test unit is connected to the PC and
supplied with power, the status LED lights up. The
different meanings of the colours of the status LED are
explained on the help pages (F1).

5.2 During use

CAUTION
Risk of injury!
Incorrectly attached components may be ejected when
under pressure.
► Always make sure that all connections have been
correctly attached.
The test procedure is described in the test software.

5.3 After use

5.3.1 Tasks to be carried out on completion
of testing
NOTICE
Do not completely close the high-pressure outlet (e.g.
with a screw plug). A pressure reducer may remain
connected.
1. Log off from the system.
2. Close the compressed air supply.
CAUTION
Risk of injury!
Injury may occur if the bleed screw is not opened with
caution.
► Make sure that the bleed screw is not directed
towards a person's hands or head.
3. Vent the test unit.
4. Disassemble the test setup.
1) Q6/8 only
|