Baker Hughes Druck DPI705E User Manual

Digital pressure indicator
Hide thumbs Also See for Druck DPI705E:

Advertisement

Quick Links

DPI705E
DPI705E-IS
Digital Pressure Indicator
User Manual
Druck.com

Advertisement

Table of Contents
loading

Summary of Contents for Baker Hughes Druck DPI705E

  • Page 1 DPI705E DPI705E-IS Digital Pressure Indicator User Manual Druck.com...
  • Page 3 PM700E (Absolute, Gauge) DPI705E Series Digital Pressure Indicator The Druck DPI705E pressure indicator uses a silicon transducer to produce a pressure reading in various pressure measurement units. These user instructions include the operations for all DPI705E Pressure Indicators, safety instructions and the requirements for intrinsically safe instruments.
  • Page 4 Please see datasheet for full details. Part Code Description IO-ADAPT-G1/4 G1/4 female adaptor IO-ADAPT-1/4NPT 1/4 NPT adaptor IO-ADAPT-1/8NPT 1/8 NPT adaptor IO-ADAPT-QF Quick-fit hose adaptor Copyright 2019 Baker Hughes Company. ii | DPI705E / DPI705E-IS User Manual–English...
  • Page 5 The instrument requires 4 x AA/LR6 batteries, fitted as shown below. Use a small Posidrive screwdriver to loosen the screws in the battery compartment cover and fit as per the battery orientation shown on the back of the instrument as illustrated below: Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | iii...
  • Page 6 The DPI705E M12 connector is to be connected to an external RTD probe as shown below. The RTD-INTERFACE is optionally supplied with a field-rewirable M12 connector for customers to connect their own wire-ended RTD’s into. This is accessory part number IO-RTD-M12CON. Copyright 2019 Baker Hughes Company. iv | DPI705E / DPI705E-IS User Manual–English...
  • Page 7 PIN, display lock behavior and factory reset. Also (Press Simultaneously) check sensor full-scale values, system date/time, software version and other General settings. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | v...
  • Page 8 12. Custom Units Indicator 13. Pressure Units Indicator 14. Tare Indicator 15. Sensor Type Absolute/Gauge Indicator 16. Minimum Symbol Indicator 17. Negative Value Indicator 18. Maximum Symbol Indicator Copyright 2019 Baker Hughes Company. vi | DPI705E / DPI705E-IS User Manual–English...
  • Page 9: Table Of Contents

    Auto shutdown 3.2.5 Battery Type 3.2.6 Date 3.2.7 Time 3.2.8 Custom Unit 3.2.9 Software Version Accessories Error Codes & Diagnostics Error Codes Diagnostics Approved Service Centers Return Goods/Material Procedure Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | vii...
  • Page 10 Copyright 2019 Baker Hughes Company. viii | DPI705E / DPI705E-IS User Manual–English...
  • Page 11: Basic Functionality

    The battery level is displayed. The positive full-scale of the sensor is briefly displayed. The number of days remaining to the next calibration due date are then briefly displayed. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 1...
  • Page 12: Pressure Measurement

    The DPI705E can be used to measure temperature using the RTD-INTERFACE. To measure temperature: Connect the RTD-INTERFACE via the port on the right of the instrument, and connect an RTD-PROBE or PT100 to the RTD-INTERFACE. Copyright 2019 Baker Hughes Company. 2 | DPI705E / DPI705E-IS User Manual–English...
  • Page 13: Zero

    RTD-INTERFACE. To perform a zero: Open all pressure ports to atmospheric pressure. Press the keys together. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 3...
  • Page 14: Tare

    Press the key again to turn the filter functionality off. The filter symbol will disappear. 1.7 UNITS To change the pressure units of the display: Copyright 2019 Baker Hughes Company. 4 | DPI705E / DPI705E-IS User Manual–English...
  • Page 15: Backlight

    2.1 Maximum Reading The maximum reading is the highest read by the instrument since power up. The value will be reset each time the device is turned off. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 5...
  • Page 16: Minimum Reading

    To view / reset the minimum pressure: Ensure the display is on the sensor reading screen as shown in the example below. Press the key. Copyright 2019 Baker Hughes Company. 6 | DPI705E / DPI705E-IS User Manual–English...
  • Page 17: Leak Test

    A clock symbol appears in the top left corner of the display. The leak test time appears with default at 60 seconds. To change the timer value, press the key, the display will cycle 60, 120, 300 seconds. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 7...
  • Page 18: Alarm

    By default, the alarm sounds for 60 seconds but the display flashes until the alarm condition is removed. To turn the alarm feature on or off: Ensure the display is on the sensor reading screen. Press the keys simultaneously. Copyright 2019 Baker Hughes Company. 8 | DPI705E / DPI705E-IS User Manual–English...
  • Page 19: Set High Alarm

    Ensure the display is on the sensor reading screen. Press the keys simultaneously. The current state of the alarm will be displayed as shown below. Ensure the alarm is set to Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 9...
  • Page 20: Set Low Alarm

    2.4.2 Set Low Alarm The low alarm allows an alarm to be set for when pressure reading goes below the alarm low threshold To set a low alarm: Copyright 2019 Baker Hughes Company. 10 | DPI705E / DPI705E-IS User Manual–English...
  • Page 21 Setting an invalid threshold will result in an error message. For example, Error.03 which means the value is out of bounds. If this happens press the key and enter the value again. Press the key to go to the main screen. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 11...
  • Page 22: Settings

    There are two categories of Settings within DPI705E, General Settings and Advanced Settings. To access these settings: Ensure the display is on the sensor reading screen. Press the keys simultaneously. Copyright 2019 Baker Hughes Company. 12 | DPI705E / DPI705E-IS User Manual–English...
  • Page 23: Advanced Settings

    Note: All sensors allow a two point calibration only. It is recommended to set calibration point “ ” (Calibration Point 1) near zero and “ ” (Calibration Point 2) near the full-scale value of the sensor. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 13...
  • Page 24 When the fourth digit has been set press the key to submit the PIN value. If the PIN is correct “ ” (Adjust) is displayed on the screen as shown below. Copyright 2019 Baker Hughes Company. 14 | DPI705E / DPI705E-IS User Manual–English...
  • Page 25: Calibrate Sensor

    Once the clock symbol stops rotating and disappears, press the key. Enter the pressure/resistance value applied. Use the key to increment the current digit value, and the key to move to the next digit. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 15...
  • Page 26 DPI705E. 16. Press the key again to return to the “ ” (Advanced) screen. 17. Press the key again to return to the sensor reading screen. Copyright 2019 Baker Hughes Company. 16 | DPI705E / DPI705E-IS User Manual–English...
  • Page 27: Set Calibration Date

    Setting an incorrect calibration interval may cause the instrument to be out of calibration leading to incorrect pressure or temperature measurements leading to possible compliance, environmental or health and safety issues. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 17...
  • Page 28: Change User Pin

    12. Press the key again to return to the sensor reading screen. 3.1.5 Change User PIN This feature allows the default user PIN to be changed. To change user PIN: Copyright 2019 Baker Hughes Company. 18 | DPI705E / DPI705E-IS User Manual–English...
  • Page 29 12. If the PIN is accepted, the user will be redirected back to the “ ” (PIN) screen as shown above. Note: If the PIN values don’t match, the error screen shown below will be displayed. Press key to re-enter the PIN. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 19...
  • Page 30: Units Lock

    Note: The screen shows the current measurement unit and indicates its state with text “ ” (On) - enabled and “ ” (Off) - disabled. To disable the unit, press the key. 10. Press the key to go to the Next Unit. Copyright 2019 Baker Hughes Company. 20 | DPI705E / DPI705E-IS User Manual–English...
  • Page 31: Lock Display

    ” (Off) which means the lock is disabled. 10. To enable the lock, press the key. The display will then show “ ” (On) which means lock is enabled. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 21...
  • Page 32: Factory Reset

    If the PIN is correct, the display will show “ ” (Adjust). Press the key 6 times to navigate to the “ ” (Reset) screen as shown below. Copyright 2019 Baker Hughes Company. 22 | DPI705E / DPI705E-IS User Manual–English...
  • Page 33: General Settings

    Turn audio on or off Auto Shutdown Enable or disable auto-shutdown Battery Type Allows setting the battery type to get maximum usage and accurate capacity readings Date Allows setting of system date Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 23...
  • Page 34: Full-Scale Value

    3.2.2 Calibration Due Date The calibration due date is the date when the sensor is next due for calibration. To see the days until the calibration due date: Copyright 2019 Baker Hughes Company. 24 | DPI705E / DPI705E-IS User Manual–English...
  • Page 35: Audio

    Ensure the display is on the sensor reading screen. Press the keys simultaneously. The display will show “ ” (Advanced) which indicates that the unit is now in the Settings option. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 25...
  • Page 36: Auto Shutdown

    Ensure the display is on the sensor reading screen. Press the keys simultaneously. ” (Advanced) which indicates that the unit is now in the Settings The display will show “ option. Copyright 2019 Baker Hughes Company. 26 | DPI705E / DPI705E-IS User Manual–English...
  • Page 37: Battery Type

    Note: The Intrinsically Safe DPI705E-IS can only be used with the first two battery technologies listed (Al and LiFeS) - refer to safety manual. To change the battery type: Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 27...
  • Page 38: Date

    10. Press the key to return to the sensor reading screen. 3.2.6 Date This setting changes the system date. To change the date: Copyright 2019 Baker Hughes Company. 28 | DPI705E / DPI705E-IS User Manual–English...
  • Page 39 To change the year value, press the key to increment the current (flashing) digit, and key to move to the next digit. Note: Only the last 2 digits of the year can be changed. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 29...
  • Page 40: Time

    Ensure the display is on the sensor reading screen. Press the keys simultaneously. The display will show “ ” (Advanced) which indicates that the unit is now in the Settings option. Copyright 2019 Baker Hughes Company. 30 | DPI705E / DPI705E-IS User Manual–English...
  • Page 41: Custom Unit

    The scaling factor will only be applied when the custom unit is selected during pressure measurement. To set a custom scaling factor: Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 31...
  • Page 42 • The unit starts flashing which means it is the current unit. • The default scaling factor for whichever unit the device is currently set to is 1.0000 Copyright 2019 Baker Hughes Company. 32 | DPI705E / DPI705E-IS User Manual–English...
  • Page 43: Software Version

    3.2.9 Software Version This setting shows the application software version on the instrument. To see the application software version: Ensure the display is on the sensor reading screen. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 33...
  • Page 44 The current software version will be displayed as shown below. Press the key to go back to the previous screen. Press the key again to return to the sensor reading screen. Copyright 2019 Baker Hughes Company. 34 | DPI705E / DPI705E-IS User Manual–English...
  • Page 45: Accessories

    2 adaptors AMC Quick Connect to M14F & M20F IO620-COMP 2-way comparator adaptor AMC Quick connect male to 2 X AMC quick connect Female – for comparing 2 gauges simultaneously Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 35...
  • Page 46: Error Codes & Diagnostics

    (OK) key or the (Back) key to return to measurement screen. Err.80 Critical Software Fault Power off and back on. If problem persists return unit to service center. Copyright 2019 Baker Hughes Company. 36 | DPI705E / DPI705E-IS User Manual–English...
  • Page 47: Diagnostics

    (and therefore cannot display (Date) at startup or Cal key to ignore. Message will repeat Cal Due days) Due days indicates as “---” every power-up if ignored Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 37...
  • Page 48: Approved Service Centers

    Provide the following information for a RGA or RMA: • Product (e.g. DPI705E) • Serial number. • Details of defect/work to be undertaken. • Calibration traceability requirements. • Operating conditions. • Operating conditions. Copyright 2019 Baker Hughes Company. 38 | DPI705E / DPI705E-IS User Manual–English...
  • Page 50 Copyright 2019 Baker Hughes Company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third- party product and company names are trademarks of their respective holders. bakerhughes.com...