Summary of Contents for Baker Hughes Druck DPI705E
Page 1
DPI705E DPI705E-IS Digital Pressure Indicator User Manual Druck.com...
Page 3
PM700E (Absolute, Gauge) DPI705E Series Digital Pressure Indicator The Druck DPI705E pressure indicator uses a silicon transducer to produce a pressure reading in various pressure measurement units. These user instructions include the operations for all DPI705E Pressure Indicators, safety instructions and the requirements for intrinsically safe instruments.
Page 4
Please see datasheet for full details. Part Code Description IO-ADAPT-G1/4 G1/4 female adaptor IO-ADAPT-1/4NPT 1/4 NPT adaptor IO-ADAPT-1/8NPT 1/8 NPT adaptor IO-ADAPT-QF Quick-fit hose adaptor Copyright 2019 Baker Hughes Company. ii | DPI705E / DPI705E-IS User Manual–English...
Page 5
The instrument requires 4 x AA/LR6 batteries, fitted as shown below. Use a small Posidrive screwdriver to loosen the screws in the battery compartment cover and fit as per the battery orientation shown on the back of the instrument as illustrated below: Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | iii...
Page 6
The DPI705E M12 connector is to be connected to an external RTD probe as shown below. The RTD-INTERFACE is optionally supplied with a field-rewirable M12 connector for customers to connect their own wire-ended RTD’s into. This is accessory part number IO-RTD-M12CON. Copyright 2019 Baker Hughes Company. iv | DPI705E / DPI705E-IS User Manual–English...
Page 7
PIN, display lock behavior and factory reset. Also (Press Simultaneously) check sensor full-scale values, system date/time, software version and other General settings. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | v...
Page 8
12. Custom Units Indicator 13. Pressure Units Indicator 14. Tare Indicator 15. Sensor Type Absolute/Gauge Indicator 16. Minimum Symbol Indicator 17. Negative Value Indicator 18. Maximum Symbol Indicator Copyright 2019 Baker Hughes Company. vi | DPI705E / DPI705E-IS User Manual–English...
The battery level is displayed. The positive full-scale of the sensor is briefly displayed. The number of days remaining to the next calibration due date are then briefly displayed. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 1...
The DPI705E can be used to measure temperature using the RTD-INTERFACE. To measure temperature: Connect the RTD-INTERFACE via the port on the right of the instrument, and connect an RTD-PROBE or PT100 to the RTD-INTERFACE. Copyright 2019 Baker Hughes Company. 2 | DPI705E / DPI705E-IS User Manual–English...
RTD-INTERFACE. To perform a zero: Open all pressure ports to atmospheric pressure. Press the keys together. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 3...
Press the key again to turn the filter functionality off. The filter symbol will disappear. 1.7 UNITS To change the pressure units of the display: Copyright 2019 Baker Hughes Company. 4 | DPI705E / DPI705E-IS User Manual–English...
2.1 Maximum Reading The maximum reading is the highest read by the instrument since power up. The value will be reset each time the device is turned off. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 5...
To view / reset the minimum pressure: Ensure the display is on the sensor reading screen as shown in the example below. Press the key. Copyright 2019 Baker Hughes Company. 6 | DPI705E / DPI705E-IS User Manual–English...
A clock symbol appears in the top left corner of the display. The leak test time appears with default at 60 seconds. To change the timer value, press the key, the display will cycle 60, 120, 300 seconds. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 7...
By default, the alarm sounds for 60 seconds but the display flashes until the alarm condition is removed. To turn the alarm feature on or off: Ensure the display is on the sensor reading screen. Press the keys simultaneously. Copyright 2019 Baker Hughes Company. 8 | DPI705E / DPI705E-IS User Manual–English...
Ensure the display is on the sensor reading screen. Press the keys simultaneously. The current state of the alarm will be displayed as shown below. Ensure the alarm is set to Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 9...
2.4.2 Set Low Alarm The low alarm allows an alarm to be set for when pressure reading goes below the alarm low threshold To set a low alarm: Copyright 2019 Baker Hughes Company. 10 | DPI705E / DPI705E-IS User Manual–English...
Page 21
Setting an invalid threshold will result in an error message. For example, Error.03 which means the value is out of bounds. If this happens press the key and enter the value again. Press the key to go to the main screen. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 11...
There are two categories of Settings within DPI705E, General Settings and Advanced Settings. To access these settings: Ensure the display is on the sensor reading screen. Press the keys simultaneously. Copyright 2019 Baker Hughes Company. 12 | DPI705E / DPI705E-IS User Manual–English...
Note: All sensors allow a two point calibration only. It is recommended to set calibration point “ ” (Calibration Point 1) near zero and “ ” (Calibration Point 2) near the full-scale value of the sensor. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 13...
Page 24
When the fourth digit has been set press the key to submit the PIN value. If the PIN is correct “ ” (Adjust) is displayed on the screen as shown below. Copyright 2019 Baker Hughes Company. 14 | DPI705E / DPI705E-IS User Manual–English...
Once the clock symbol stops rotating and disappears, press the key. Enter the pressure/resistance value applied. Use the key to increment the current digit value, and the key to move to the next digit. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 15...
Page 26
DPI705E. 16. Press the key again to return to the “ ” (Advanced) screen. 17. Press the key again to return to the sensor reading screen. Copyright 2019 Baker Hughes Company. 16 | DPI705E / DPI705E-IS User Manual–English...
Setting an incorrect calibration interval may cause the instrument to be out of calibration leading to incorrect pressure or temperature measurements leading to possible compliance, environmental or health and safety issues. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 17...
12. Press the key again to return to the sensor reading screen. 3.1.5 Change User PIN This feature allows the default user PIN to be changed. To change user PIN: Copyright 2019 Baker Hughes Company. 18 | DPI705E / DPI705E-IS User Manual–English...
Page 29
12. If the PIN is accepted, the user will be redirected back to the “ ” (PIN) screen as shown above. Note: If the PIN values don’t match, the error screen shown below will be displayed. Press key to re-enter the PIN. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 19...
Note: The screen shows the current measurement unit and indicates its state with text “ ” (On) - enabled and “ ” (Off) - disabled. To disable the unit, press the key. 10. Press the key to go to the Next Unit. Copyright 2019 Baker Hughes Company. 20 | DPI705E / DPI705E-IS User Manual–English...
” (Off) which means the lock is disabled. 10. To enable the lock, press the key. The display will then show “ ” (On) which means lock is enabled. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 21...
If the PIN is correct, the display will show “ ” (Adjust). Press the key 6 times to navigate to the “ ” (Reset) screen as shown below. Copyright 2019 Baker Hughes Company. 22 | DPI705E / DPI705E-IS User Manual–English...
Turn audio on or off Auto Shutdown Enable or disable auto-shutdown Battery Type Allows setting the battery type to get maximum usage and accurate capacity readings Date Allows setting of system date Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 23...
3.2.2 Calibration Due Date The calibration due date is the date when the sensor is next due for calibration. To see the days until the calibration due date: Copyright 2019 Baker Hughes Company. 24 | DPI705E / DPI705E-IS User Manual–English...
Ensure the display is on the sensor reading screen. Press the keys simultaneously. The display will show “ ” (Advanced) which indicates that the unit is now in the Settings option. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 25...
Ensure the display is on the sensor reading screen. Press the keys simultaneously. ” (Advanced) which indicates that the unit is now in the Settings The display will show “ option. Copyright 2019 Baker Hughes Company. 26 | DPI705E / DPI705E-IS User Manual–English...
Note: The Intrinsically Safe DPI705E-IS can only be used with the first two battery technologies listed (Al and LiFeS) - refer to safety manual. To change the battery type: Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 27...
10. Press the key to return to the sensor reading screen. 3.2.6 Date This setting changes the system date. To change the date: Copyright 2019 Baker Hughes Company. 28 | DPI705E / DPI705E-IS User Manual–English...
Page 39
To change the year value, press the key to increment the current (flashing) digit, and key to move to the next digit. Note: Only the last 2 digits of the year can be changed. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 29...
Ensure the display is on the sensor reading screen. Press the keys simultaneously. The display will show “ ” (Advanced) which indicates that the unit is now in the Settings option. Copyright 2019 Baker Hughes Company. 30 | DPI705E / DPI705E-IS User Manual–English...
The scaling factor will only be applied when the custom unit is selected during pressure measurement. To set a custom scaling factor: Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 31...
Page 42
• The unit starts flashing which means it is the current unit. • The default scaling factor for whichever unit the device is currently set to is 1.0000 Copyright 2019 Baker Hughes Company. 32 | DPI705E / DPI705E-IS User Manual–English...
3.2.9 Software Version This setting shows the application software version on the instrument. To see the application software version: Ensure the display is on the sensor reading screen. Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 33...
Page 44
The current software version will be displayed as shown below. Press the key to go back to the previous screen. Press the key again to return to the sensor reading screen. Copyright 2019 Baker Hughes Company. 34 | DPI705E / DPI705E-IS User Manual–English...
(OK) key or the (Back) key to return to measurement screen. Err.80 Critical Software Fault Power off and back on. If problem persists return unit to service center. Copyright 2019 Baker Hughes Company. 36 | DPI705E / DPI705E-IS User Manual–English...
(and therefore cannot display (Date) at startup or Cal key to ignore. Message will repeat Cal Due days) Due days indicates as “---” every power-up if ignored Copyright 2019 Baker Hughes Company. English–DPI705E / DPI705E-IS User Manual | 37...
Provide the following information for a RGA or RMA: • Product (e.g. DPI705E) • Serial number. • Details of defect/work to be undertaken. • Calibration traceability requirements. • Operating conditions. • Operating conditions. Copyright 2019 Baker Hughes Company. 38 | DPI705E / DPI705E-IS User Manual–English...