Device Connectors - Eizo CuratOR LL550W Instructions For Use Manual

8mp 55" lcd monitor
Table of Contents

Advertisement

NOTICE
Cable installation
Observe the following instructions:
• Only shielded cables are to be used for all signal connections.
• The connecting cables must not be kinked.
• The minimum bending radius of a connecting cable generally equals five times the ca-
ble diameter.
• Do not route signal cables and power cables next to one another. Otherwise, line power
subject to heavy interference could result in reversible pixel errors.
• The device must not share a line power supply with motors or valves (interference!).
• Externally connected cables can represent a trip hazard. Make sure that all incoming
cables are safely routed.
• If the device offers strain relief mechanisms for the cables, use them to prevent unin-
tended loosening of connected cables.

5.2 Device connectors

CAUTION
Opening the connection panel cover
• Only service may open the connection panel cover.
• The screw torque may not exceed 0.75 Nm +/- 0.05 Nm.
• Patients must not be present when the cover is open.
NOTICE
Isolating the device from mains supply
Means of isolating the device from mains supply is by disconnecting the AC power line.
• Make sure that the AC power line is easily accessible.
Power connector
(not shown)
Fig.: Lower connection panel
CuratOR LL550W
Instructions for Use, 02/2024
USB A
5.2 Device connectors
DVI