Setup Menu - OHAUS Dafender 3000 i-DT33P Instruction Manual

Table of Contents

Advertisement

Defender 3000 Indicators
4.3

Setup Menu

Enter this menu S.E.t.U.P to set scale parameters. Default settings are in bold.
Menu
Sub-Menu
Reset
Capacity Unit
Range
>|1|< Capacity
>|1|< Graduation
Setup
>|2|< Capacity
S.E.t.U.P
>|2|< Graduation
Power On Zero
Power On Unit
Auto Tare
End
Note: / refers to not applicable in the table above.
reset
Reset [
]
Reset the Setup menu to factory defaults.
NO
= do not reset
yes
= reset
Capacity Unit [C.
Select the unit used for calibration.
kg
lb
rANGE
[
]
Range
Selects the number of weighing intervals.
This terminals can be configured to use single or dual interval. Each interval can be assigned its own
graduation. If dual interval is selected, the graduation will change when the weight reaches the second
interval.
When Single interval is selected, the additional parameters available are:
Capacity
Graduation
When Dual interval is selected, the terminal functions with two intervals, each with its own capacity and
graduation. In addition to the Interval 1 capacity and graduation parameters, the following two parameters
are available:
Capacity 2
Graduation 2
Cap
[
] /
Capacity
Set the capacity of the scale or the first capacity of the scale.
0.1-99999.9
GrAd
[
] /
Grad
Grad1
Set the scale readability or the first readability of the scale.
0.0001~100
Cap2
[
Capacity2
Set the second capacity of the scale.
Sub-Menu (in
segment)
reset
C.UNIt
rANGE
CAP
Grad
CAP
Grad
P.ZerO
P.UNIt
A.tare
End
UNIt
]
Cap1
[
]
Capacity1
GrAd1
[
]
]
Options
no, yes
kg, lb
Single, Dual
0.1-99999.9
0.0001~100
0.1-99999.9
0.0001~100
Off, On
Auto, g, kg, lb, oz, lb:oz, t
Off, On, Accept
/
EN-31
Options (in segment)
NO, yes
/
SINGLE, DUAL
/
/
/
/
Off, ON
AUtO
Off, ON, ACCept
/