Linearity Calibration [ Lin ]; Geo Code Adjustment [ Geo ] - OHAUS Dafender 3000 i-DT33P Instruction Manual

Table of Contents

Advertisement

Defender 3000 Indicators
c)
Repeat the process until all the digits are correct.
d)
Press the Yes button to accept calibration point. It is flashing on the display.
6.
Place a calibration mass of the specified weight on the pan and press the Yes button.
7.
The display shows
8.
Then the display shows
Note:
 If calibration is failed,
 If after waiting for 40 seconds the calibration is still not successful,
and the previous calibration data is restored.
4.2.4
Linearity Calibration
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the pan.
The mid calibration point is established with a weight equal to half of the full calibration weight on the pan. The
zero calibration point is established with no weight on the pan. The full calibration and mid calibration points
can be altered by users during the calibration procedure.
Calibration procedures:
1.
Long press the Menu button until you see
2.
Release the button and wait for the display to show
3.
Short press the No button several times to navigate until you see
0
4.
kg and the calibration unit are flashing on the display. With no weight on the pan, press the Yes button to
establish the zero point. The display shows
5.
Then the first calibration point and the unit are flashing on the display based on the scale capacity and unit
you set in the setup menu. (For example,
skip to step 7.
6.
To change the calibration point:
a)
Short press the No button several times until the desired number appears.
Note: press the Back button can decrease the digit.
b)
Short press the Yes button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
d)
Press the Yes button to accept calibration point. It is flashing on the display.
7.
Place a calibration mass of the specified weight on the pan and press the Yes button. The display shows
--C--
.
8.
The second calibration point and the unit are flashing on the display based on the capacity and unit you
set in the setup menu. (For example,
Note:
If after waiting for 40s the calibration is still not successful,
previous calibration data is restored.
9.
Repeat step 6 and 7.
10. The display shows
11. After that the display shows
4.2.5
GEO Code Adjustment
Geographical Adjustment Factor (GEO) code is used to adjust the calibration based on the current location.
Settings from 0 to 31 are available with 12 being the default.
Please refer to the Table of Geo Code Values section in the Technical Data chapter to determine the GEO
factor that corresponds to the indicator's location.
To set the GEO factor:
1.
Long press the Menu button until you see
2.
Release the button and wait for the display to show
3.
Short press the No button several times to navigate until you see
4.
The Geo point is flashing on the display (For example,
--C--
-DONE-
, and then
Lin
. Press the Exit button to exit.
CAL E
is displayed for 3 seconds and the previous calibration data is restored.
LIN
[
]
015.000
030.000
--C--
-DONE-
, and then
GEO
. Press the Exit button to exit.
[GEO]
when the calibration is finished.
mM.E.N.U
.
C.A.L
. Press the Yes button.
--C--
.
kg). If you do not need to change the calibration point,
kg)
CAL E
when the Linearity calibration is finished.
mM.E.N.U
.
C.A.L
. Press the Yes button.
12
).
CAL E
is displayed for 3 seconds
LIN.
Press the Yes button.
is displayed for 3 seconds and the
GEO.
Press the Yes button.
EN-29