Fault/message
The device cannot be
switched on.
Volume too low/
sample invalid.
Device outside tem-
perature range
(over range)
Volume too small
Volume too great
Sample interruption
Fault number
Dräger Alcotest 6000
Cause
Battery drained.
Test subject is blow-
ing intermittently or
not hard enough.
Device is too hot or
too cold.
Measuring range
exceeded. Alcohol
residues that are still
in the mouth may
cause incorrect mea-
suring values.
Breath sample vol-
ume too small.
Measuring range
exceeded.
Irregular exhalation,
abrupt end, or inhala-
tion
Device fault
Troubleshooting
Remedy
Replace battery.
Blow into the mouth-
piece harder and con-
tinuously.
Allow the device to
cool down or warm
up.
A wait time of at least
15 minutes after alco-
hol was last orally
consumed must have
elapsed (see "Prereq-
uisites for use",
page 8). To repeat,
press the
button.
Repeat breath sam-
ple.
Repeat breath sam-
ple.
Repeat breath sam-
ple.
Switch the device off
and on again. If the
fault remains, contact
DrägerService.
|