Electrical Connection

4.1

Cable Colors and Pin Assignment (14 wire connection cables)

The electrical connection of the pyrometer (supply voltage and measuring signal) will be done via the con-
nector on the rear panel. For this purpose, pre-assembled connection cables are available as accessories (see
9.3
Accessories). To prevent accidental short circuits, not used cable wires need to be placed, for example to
the supplied screw terminals.
Cable colors Process
Sensors Corporation
pyrometers
White
Brown
Green
Yellow
White-green
Brown-green
Grey
Pink
Blue
Black
Grey-pink
Violet
Red-blue
Red
Orange
1)
Reference potential 0 V, brown
|NOTICE|
We generally recommend the use of the current 14-wire connection cables. Precaution when
using 12-wire connection cables that were delivered before 2015:
Connection cable for the previous models Metis and Sirius cannot be connected without restrictions on a M3 /
H3 pyrometer!
Old connection cables are bridged in the cable connector to pins E-C and D-F. Connecting such a cable to a
M3 pyrometer, serial interface and analog output 2 and digital input / output 3 are connected and the galvanic
isolation between the outputs is bypassing. Thereby the pyrometer is not destroyed, but the functions are no
longer available. Using such a cable only analog output 1 is available as output.
Cable colors old cable
White
Brown
Green
Yellow
Grey
Pink
Orange
Red
Black
Violet
Grey / pink
Red / blue
10
METIS M308 / M309 / M313 / M316 / M318 / H309 / H316 / H318 (12-pin)
No.
Function
1
+ 24 V DC power supply (18–30 V DC)
2
0 V DC power supply (ground)
3
+ Analog output 1 (0 / 4–20 mA)
4
- Analog output 1 (0 / 4–20 mA)
14
+ Analog output 2 (0 / 4–20 mA)
15
- Analog output 2 (0 / 4–20 mA)
5
Digital input / output 1
6
Digital input / output 2
13
Digital input / output 3 / Analog in
9
RS232: RxD
2)
RS485: B (+)
11
10
RS232: TxD
RS485: A (-)
2)
12
8
DGND (ground for interface)
Shield (connect only for cable extension,
7
do not connect in the control cabinet)
2)
H3 models only RS485
Function
Power supply +24 V DC
Power supply 0 V DC (ground)
+ Analog output (0/4–20 mA)
- Analog output (0/4–20 mA)
Targeting light, external switch (bridge to K)
Maximum value storage, external clearing (bridge to K)
Screen
DGND (Ground for interface)
RxD (RS232) or B1 (RS485)
TxD (RS232) or A1 (RS485)
RxD (RS232) or B2 (RS485) (bridge to F)
TxD (RS232) or A2 (RS485) (bridge to C)
Pin
K
A
L
B
D
B
1)
H
1)
J
1)
E
F
C
G
E
G
F
M
D
H
C
J
L
A
B
K
Connector pins
of the 12 pin
pyrometer connector
(view from outside)
M and plug housing
Pins
K
A
L
E
G
F
B
M H
D
H
J
C
J
L
M
A
B
K
G
F
Connector pins
C
cable socket (solder
D
side)
E