L-4
Figure C.
2) Insert the hose of the warming unit in the inlet port on the blanket. Use a twisting motion to
ensure a snug fit (see Figure C).
3) Connect the warming unit to a properly grounded power source.
4) If so equipped, turn the Isolation Switch to the ON position (see the section titled Isolation Switch).
5) Press the System ON/OFF button to turn the warming unit ON and select the appropriate
temperature setting.
6) Place a cotton blanket over the Bair Hugger blanket for maximum effectiveness.
7) Monitor the patient's temperature at least every 10 to 20 minutes and adjust the temperature
setting of the warming unit as required.
SAFETY INSPECTION
Perform a safety inspection after making repairs to the warming unit, and before returning the
warming unit to service. A safety inspection must include a test of the operating temperatures
(described in the Service Manual), the Over Heat Alarm system, as well as a leakage current test.
PROPER USE AND MAINTENANCE
Augustine Medical, Inc. assumes no responsibility for the reliability, performance, or safety of the
equipment if:
l
Modifications or repairs are performed by unauthorized personnel.
l
The equipment is used in a manner other than that described in the Operation or Service Manuals.
l
The equipment is installed in an environment that does not meet the relevant grounding requirements.
Setup and Operation
The Bair Hugger' Total Temperature Management@ System is easy to set up and to use. Follow the
instructions provided with each blanket for specific information.
1)
Place the Bair Hugger blanket on the patient with the perforated side (the side with small holes)
against the patient's skin.

Advertisement