I.V. pole.
BLANKET DISPENSER
The blanket dispenser is available for use on all warming units, except the Model 505 unit. This
dispenser attaches to the warming unit and can store several Bair Hugger blankets. See the package
instructions for details.
500/OR,E warming units have a port that accommodates an Augustine
Medical, Inc.
500/OR and
I.V.
pole option
I.V. POLE
Figure I? Warming unit with
blanket dispenser
Model
31.9"C
Fluid input temperature: 20°C. Increase fluid temperature by an average of 3°C by tucking the
outlet tubing in Bair Hugger blanket channels.
Figure N. Fluid warming set inside warming unit hose
Optional Features
This section discusses optional features for Bair Hugger@ warming units. For more details about these
features, please call Augustine Medical or your local distributor.
Figure 0. Warming unit with
ml/hr
33.2-C
3000
ml/hr
34.7"C
2500
ml/hr
2000
36.1"C
ml/hr
36.8"C
1500
ml/hr
34.9"C
1000
ml/hr
500
*p
.~
-p
Fluid Warming Specifications
The average fluid output temperatures during normal operation of the 241 fluid warming set with
the warming unit set on HIGH are as follows:
INFUSION RATE
OUTPUT TEMPERATURE