Mounting The Model 505 Warming Unit - Bair Hugger 505 Operator's Manual

Warming units
Hide thumbs Also See for 505:
Table of Contents

Advertisement

J).
Kpok tipping, catheter site
trauma, and patient injury.
USING A BED RAIL
The Model 505 warming unit can also hang on the edge of a bed. The safety strap is designed to
loop around the bed rail, keeping the Model 505 unit safely suspended even if the unit is
inadvertently dislodged from the bed rail (see figure
I.
wbeeIbase. Failure to do so may result in
(356cm) radius
14"
V
pole with a
minimum
I2 cm) on an I.
than 44" (
I
the unit no higher
that
provides stability. We recommend clamping
V
pole at a height
To prevent tipping, clamp the Model 505 warming unit to an I.
Figure I. Model 505 warming unit
attached to an I.V. pole
Figure J. Model 505 warming unit
attached to a bed rail
Mounting the Model 505 Warming Unit
USING AN I.V. POLE
The Model 505 warming unit clamps easily to an I.V. pole (see figure I). Simply turn the handle
clockwise to tighten the clamp onto an I.V. pole, counterclockwise to release.
WARNING: