Warming Fluids Using The 241" Fluid Warming Set - Bair Hugger 505 Operator's Manual

Warming units
Hide thumbs Also See for 505:
Table of Contents

Advertisement

fit.
5b) To perform fluid warming alone: Attach the hose cap to end of hose. Direct the hose end
away from patient.
6) Make patient connection. Do not entrap air.
7) Press the System ON/OFF (ON/STANDBY) button to turn the Series 500 Bair Hugger@ unit
ON, and select the appropriate temperature setting. Tuck the outlet and extension tubing between
the blanket channels to insulate.
8) Begin infusion. Use the temperature indicator to monitor the fluid temperature. The green square
indicates fluid temperature.
1 fluid warming ser. The 24 1 fluid warming set is latex-free.
1) Connect the 241 fluid warming set to the fluid source using the end of tubing without the
temperature indicator. Thoroughly prime all tubing.
2) Push the tab to open the notch on the hose end.
3) Slide the coil completely into the hose until the lever on the 241 fluid warming set reaches the
base of the notch.
4) Rotate the lever (1) of the fluid warmer towards the hose collar until it is parallel with the hose
collar. Push the tab (2) to close the notch on the hose end.
5a) To perform fluid warming with skin surface warming: Insert the end of the hose in the opening
for the hose on a Bair Hugger blanket. Use a twisting motion to ensure a snug
ml/hr.
Before initiating fluid warming therapy, read the package insert for the
24
W-ing Set
Series 500 warming units are equipped with a special hose for 241 fluid
warming. This allows the Bair Hugger' Total Temperature Management@
system to warm blood and intravenous fluids delivered through a
Bair Hugger 241 fluid warming set at infusion rates up to 3,000
241@ Fluid
Warming Fluids Using the

Advertisement

Table of Contents
loading

This manual is also suitable for:

500/0r

Table of Contents