IDA-6 – User Manual – Maintenance

Maintenance

Cleaning
Turn off the product and disconnect the power adapter
before cleaning with a cloth dampened with a mild
detergent solution.
Storage and transportation
Before storing or transporting the product, drain the
channels as described in "Drain a channel" on page
31. For packaging recommendations, see "Notices" on
page 3.
Draining is particularly important if the product is at risk
of being exposed to freezing temperatures.
Drain a channel
To remove liquid from a channel, use a syringe with air:
1. Make sure that the product is turned on.
2. Disconnect the drain tube from the outlet port.
3. Connect a syringe filled with air to the inlet port.
4. Gently push air through the channel. The amount
and rate should be approximately 20–30 mL of air
during 30 seconds.
Note
Figure 24: Draining a channel.