Quick Card
ONA-800 Platform
E4100-series OTDR SMART TEST ASSISTANT
This quick card describes how to connect to a fiber under test, configure SMART TEST OTDR settings,
run tests, and analyze results with a VIAVI ONA-800 and the 4100-series OTDR module.
Equipment Requirements:
• ONA-800 equipped with the following:
o Fiber Optics Software Release V18.94 or
greater
o E4100 Series OTDR Module
• Fiber optic cleaning and inspection tools
• Launch Cable (minimum 20-meter Fiber optic
patch cable) with connectors matching the OTDR
port and Fiber Under Test
• Optical Coupler to connect Launch Cable to Fiber
Under Test
The following information is required to complete the
test:
• Type of Fiber (Multimode or Single Mode)
• Type of Connectors (SC UPC, SC APC, LC UPC, etc.)
• Distance unit (feet, meters, miles, kilometers)
Fiber Inspection Guidelines:
• Use the VIAVI P5000i or FiberChek Probe
microscope to inspect both sides of every
connection being used (OTDR Port, Launch Cable,
bulkhead connectors, patch cords, etc.)
• Focus the fiber on the screen. If dirty, clean the
connector.
• If it appears clean, run inspection test.
• If it fails, clean the fiber and re-run inspection
test. Repeat until it passes.
Connect Launch Cable to the OTDR port:
1. Inspect the OTDR port on top of the ONA-800.
2. Inspect the fiber end face of Launch Cable.
3. Connect Launch Cable to the OTDR port.
Figure 1: Equipment Requirements
Figure 2: Inspect Before You Connect (IBYC)
VIAVI Solutions

Advertisement

loading

Summary of Contents for Viavi E4100 Series

  • Page 1 E4100-series OTDR SMART TEST ASSISTANT This quick card describes how to connect to a fiber under test, configure SMART TEST OTDR settings, run tests, and analyze results with a VIAVI ONA-800 and the 4100-series OTDR module. Equipment Requirements: • ONA-800 equipped with the following: o Fiber Optics Software Release V18.94 or...
  • Page 2 Note: The Launch Cable should not be connected to the Fiber Under Test during measurement. 9. Alarms: Choose User for Viavi recommended Pass/Fail limits. Figure 6: Launch Cable Setup 10. Distance Unit: If you wish to use a different unit of measure for the test results, change the unit of measure to the desired value.
  • Page 3 VIAVI Solutions Connect to Fiber Under Test (FUT): The Launch Cable may be connected to the FUT via an optical patch panel (OPP) or an optical coupler: ONA-800 1. If the interface to the FUT is a patch cord, connect the patch cord to an optical coupler with the same connector type.
  • Page 4 Figure 12: SMART TEST Setup © 2020 VIAVI Solutions Inc. Contact Us +1 844 GO VIAVI To reach the VIAVI office nearest you, Product specifications and descriptions in this (+1 844 468 4284) visit viavisolutions.com/contacts. document are subject to change without notice.

This manual is also suitable for:

Ona-800