DIAGRAM 0.3 ELECTRIC MOTORS
CHARACTERISTICS, CONNECTIONS, PROTECTIONS
Thr
'lowing electrical information
is intended as a guide : local
eld
_al
codes
; : NEMA
specifications
and
procedures
must
be
adhéred to for all electrical devices.
Use of an electrical protective device on the pump motor protects :
- The motor : in the event of overvoltage or pump seizure, the resul-
ting excess current can destroy the motor windings and possibly the
starter.
- The pump : in the event of a locked pump rotor due to contamina-
ted oil, presence of particles, or non- "lubrication, serious damage
may occur.
Type of protection
NEVER
USE
FUSES
TO
PROTECT
A
THREE-PHASE
MOTOR
UNLESS ALSO FITTED WITH A DIFFERENTIAL SYSTEM : Differential
thermal
circuit interupters
or motor
starter must
be used
; the
mechanism includes a quick-opening switch controlled by a bimetal-
lic strip, which is activated by excessive current draw.
Cholce of size
Two operating modes must be considered : starting and continuous
duty.
Starting :
During starting, the torque furnished by the motor to start the pump
is considerable :
- In a single-phase motor : use of the auxiliary phase and a starting
current which can reach § to 7 times the rated current :
- In a three-phase motor : excess
current 2 to 2.5 times the rated
current.
Starting In a Cold Environment (less than 18°C/65°F)
In certain cases, when the pump has to startin a cold environment, or
with slightly contaminated or viscious oil, the current after starting
cap
~mnain high until the pump oil heats up (10 to 20 minutes) :
- 4
»+phase motor : twice the rated
current.
- thn ve-phase motor : 1.5 times the rated current.
These values are given as examples, and can vary according to the
type of pump and oil used.
Continuous Duty
This is the normal pump operating mode after about 1 hour of opera-
tion. The current is then slightly less than the rated current of the
motor.
As a consequence, dual integrated protection should be used : one
for continuous duty (which will also protect the pump for starts under
normal conditions — oil clean, temperature above 18°C, etc.), and
one for start conditions.
When
starting : the time the pump
takes to get up to full speed
should not exceed 15 seconds.
Pumps on 60 Hz (European pumps)
MECHANICAL PUMPS
MOTOR
{| POWER
|VOLTAGE |
SIZE
2002A - 20028
Single phase| 0,170KW |
220V
2A
Woe ac aA
Single phasel 0.375KW|
220V |
4 A
MOIZAC btacm! = 2o12ay
(Single phase| o.a75KW|
220v |
6 A
eae 20044
Three phase| 0,375 kW
oy
TOA
MORAG -20124C--2012CPI"|Tvee phase| aTSKN | SOV | 9 A
Single phase} 0.55 kW]
220V
6A
pee
~ ROZDAC - 2020 CP1
Three phase| 0,55 kW
ey
} '
[
A
atl
Three phase{
1,1
kW
ay
é ;
ea CPT
63H
Three phase| 2,2
kW
av
iF
;
Pumps for 60 Hz (US Pumps)
a
MOTOR
power | VOLTAGE |
SIZE
2002A
.
120V
4A
0028 °
Single phase 60 Hz
200W
220
QA
1004A - 1004AC | Single phase 50Hz]
0,375 kW
100 V
10A
1012A
Single phase 60Hz |
0,550 kW
100V
10A
2008 A - 2008AC | Single phase 60Hz | 0,245 kW
115V
BA
2012 AH
2004A
Three phase 50Hz | 0,375 kW
200V
4A
2012A
Three phase GOHz }
0,245 kW
200V
4A
2012 AC
Three phase 60Hz | 0,375 kW
220V
4A
2012CP1
Three phase 60Hz | 0,550 kW
450V
4A
Single phase 50Hz | 0,550 kW
100V
10A
Single phase 60Hz |
0,660 kW
100 V
10A
wat
Single phase 60Hz | 0,660 KW
115V
BA
9020 CP1
Three phase 50Hz |
0,550 kW
200V
4A
Three phase 60Hz | 0,660 kW
220V
4A
Three phase 60Hz | 0,660 kW
460V
2A
2033
Three phase 60Hz |
1,1
kW
200V
6A
2033C
Three phase GOHz |
1,1
kW
220V
6A
2033 CP1 - 2033H | Three phase 6OHz {
1,1
kW
460V
4A
2063
Three phase SOHz |
2,2
kW
200V
10A
2063 C
Three phase 60Hz |
2,2
kW
200V
10A
2063 CP1
Three phase 60Hz }
2,2
kW
220V
10A
2063 H
Three phase 60Hz |
2,2
kW
460V
GA
Electrical Connections for Three-Phase Motors
The pumps can be equipped with different brands of motors, the
numbered connections are for your information. If there is any doubt,
only the specification
label in the electrical
box will supply th:
correct information.
1) CEI Terminal Box (Europe) :
LOW-VOLTAGE
CONNECTIONS
200 or 220 V - 50 or 60 Hz
Delta connection
HIGH-VOLTAGE
CONNECTIONS
380 V - 50 Hz or 460 V - 60 Hz
Star connection
x
Y
Vx
z
x
Y
Vv
o—e—-0
.
wl \we
x] z
u
vow
vue
ve
we
w7
Nu
|
| |
Connection
examples.
2) NEMA Terminal Box (USA) (9 wires) :
LOW-VOLTAGE
HIGH-VOLTAGE
CONNECTIONS
CONNECTIONS
200 or 220 V - 60 Hz
460
V - 60 Hz
Parallel connection
Serie connection
oti
3) NEMA Terminal Box (USA) (12 wires) :
LOW-VOLTAGE
HIGH-VOLTAGE
CONNECTIONS
CONNECTIONS
220 or 240 V - 50 or 60 Hz
380 or 460 V - 50 or 60 Hz
Parallel connection
Serie connection
fd
t
't
V7
6
4
|
4
90
8@
7
90
8
i
ee
Tay, TAN