4. Press the"Start Key" , the instrument starts the shaking operation.
5. Press the"Stop Key" , the instrument stops running indicating that the unit is in adjustment
mode.
6. A variety of tray accessories can be replaced, and the tray can be replaced by locking nuts.
Troubleshooting and Error Code Information
Error
code
No
E01
response
(LED OFF)
The shaking board
E02
operation failure
Inaccurate
E03
speed
Shaking
E04
jammed
Problem
Power switch is OFF
operating
found
Switch position is OFF
Internal
disconnected
Target speed is not set
Driver board damage
Motor damage
Driver board damage
shaking
Photoelectric
damage
Driver board damage
speed
is
Motor damage
SCILOGEX.COM
Reason
Check and connect the power
supply before restart. Check the
adapter.
Turn the instrument on, and check
set speed displayed on the LED.
cables
Open the instrument and check all
connections before reconnecting.
Set the target speed, and ensure
that indicator lamp is ON.
Replace the driver board.
Replace the motor.
Replace the driver board.
switch
Replace the photoelectric switch.
Replace the driver board
Replace the motor.
Solution