Weekdays/working days
Fig. 20
Leica EG1150 H – Paraffin Embedding Station
The automatic switch-on feature is linked with the individual weekdays. It is
therefore necessary to define the days for which the automatic switch-on
feature shall operate.
Only on those weekdays that have been defined as working days,
the instrument will be at the required temperature and ready to
operate.
Setting the current weekday:
• Press the DAY button as many times as required until the green "TODAY"
LED lights up. The green LED of the current weekday is illuminated.
• If necessary, set the current day with the
ated LED lights up.
Defining the working days:
• Press the DAY button as many times as required until the green LED
"WORKING DAY" lights up. The LEDs of the weekdays defined as working
days light up, while the LED of the current day flashes.
• Use the
/
buttons to select the day to be defined or canceled as
the working day.
The LED of the selected day (FRI) flashes.
• Press the ON/OFF button.
If that day had not been a working day (LED off), it will now be added to
the working day list (LED on).
Otherwise (see
Fig.
day will no longer be defined as a working day (LED off).
If the programmed values are to be effective on the following days
(the instrument is operational at the programmed time and switches
off automatically when the end time is reached), the standby mode
must be active --> press ON/OFF.
20, FRI was defined as a working day, LED on), the