Setting The Gravity Factor [374|703|284|954] - Seca 7031321007 Service Manual

Table of Contents

Advertisement

Maintenance
seca
2.4

Setting the Gravity Factor [374|703|284|954]:

.
The gravity factor can only be set on non-approved models [374|703|284]
Before the gravity factor can be set, the scale must first be switched to
adjustment mode (see section 0 or 2.1.4).
While the calibration count is displayed, a key must be kept pressed (for more
than 1.5 seconds) to switch the scale to adjustment mode.
As soon as the text "CAL" is displayed, press a key briefly (less than 1.5
seconds). The display shows "GAL".
Then press a key (for more than 1.5 seconds) to switch the scale to the mode
for gravity factor setting. The display shows "dec" and then the gravity factor
currently set. The gravity factor is shown in mm/s².
The scale is in decrementing state.
If the gravity factor displayed matches the target value, no further setting is
required and the menu item can be closed.
If the gravity factor shown on the display needs to be adjusted, it must be
altered by pressing a key until it matches the gravity factor required. Estimate
whether the required gravity factor is easier to reach by decrementing or by
incrementing. To toggle between decrementing mode and incrementing mode,
press a key (for more than 1.5 seconds.) (Display shows dec or inc.)
Press the key for less than 1.5 seconds to increase or decrease the gravity
factor displayed by 5 mm/s². Repeat the procedure until the gravity factor
displayed and the target gravity factor match.
Gravity factor setting can now be completed by pressing a key until the display
shows "Sto". The scale stores the new gravity factor and switches off
automatically.
05.10.12/RLEU/ARI/MRE
16/38
30-34-00-812b