Updating Old Printers; Problems With Updating - Seca 7031321007 Service Manual

Table of Contents

Advertisement

Software
5.7

Updating Old Printers

The printers of the old generation cannot be updated. This means they cannot
be used in an updated wireless group. Re-programming of the printer at the
seca headquarters in Hamburg is necessary.
The first printout shows whether the printer
already belongs to the new generation. The
new version consists of the download
kernel, i. e. the program core "08-06-15-
306a (rev. 10)" and the actual printer
software "08-06-15-307b (rev.33)". Updating
the software via SD card is possible if the
kernel number 306a – or a more recent
number (i.e. 306b, 306c, etc.) – is given on
the printout.
5.8

Problems with Updating

If the wireless connection is unstable, the PC program "seca wireless updater"
repeatedly attempts to send the radio module software. If this is unsuccessful
you can try to restart updating.
Remedies if Updating Fails
- Hold the wireless USB adapter closer to the radio module
- Avoid metal surfaces in the vicinity
- Do not stay in the immediate line of sight between the USB wireless adapter
and the radio module
- Wait until interferences from electrical units (e.g. drills) or heavy wireless
traffic (WLAN, bluetooth) have stopped
NEVER
repeat set-up of the wireless group if an updating attempt has failed.
Set-up is not possible in updating mode – it would no longer be possible to
identify the corresponding device.
In rare cases, updating will fail entirely. The relevant radio module must then be
reaced. The radio module can be returned to the seca headquarters for repair.
05.10.12/RLEU/ARI/MRE
33/38
seca
30-34-00-812b

Advertisement

Table of Contents
loading