3.2.11 Connection of sample outlet and sample inlet

The outlet funnel cover is attached to the outlet funnel.
1.
Loosen the cover of the outlet funnel from the fastening.
2.
Push the open drain tube from the last flow through armature through
the cover of the outlet funnel until it protrudes inside.
3.
Mount the cover of the outlet funnel onto the outlet funnel.
figure 3-5 Outlet funnel with mounted drain tube
4.
Direct the water from the outlet funnel to a drain for waste water.
5.
Connect the sample feed tube to the sample inlet.

3.2.12 Tidying up the cables

The drinking water panel has cable passages to stow cables behind the drinking
water panel (see section 1.2.4).
CAUTION
The outlet of the DW/P drinking water panel may only be
connected to an open channel flow that is virtually pres-
sureless.