Pdataout - Racal Instruments 1260-64 Manual

18ghz microwave switch module
Table of Contents

Advertisement

PDATAOUT

Module Specific Syntax 3-2
number. For Bank A and Bank B, this value must be between 00
and 15. For relays S1, S2, S3, and S4, this must be between 00
and 05. Note the leading 0 for relays 00 through 09 is required.
Refer to Figures 4-1,4-2, and Table 4-1 for banks, relay
numbers, and connector pins for the 1260-64 module.
If more than one connection is to be made or broken on the
1260-64 with contiguous relays, the following format is
supported:
<mod addr>.<bank no><relay no>-<bank no><relay no>
Multiple groups of relays can be specified on a single command
line by separating the path designators by commas. Command
lines terminate at the end of the line.
EXAMPLE:
OPEN 3.000,004-015,100-1 15,201,303
The PDATAOUT command causes the specified module to
transmit the CLOSED state of the relays in the 1260-64 module.
The syntax used is:
PDATAOUT <mod addr>[;<mod addr>][;<mod addr>]....
The response to the PDATAOUT command for the 1260-64 is as
follows:
<header>
<mod addr>. <bank no><group no>[,...]
<bank no><group no>[,...]
<mod addr>.END
where <header> is as follows:
1260-64A: <mod addr>. 1260-64A Quad 1x6 SWITCHING
MODULE
1260-64B: <mod addr>. 1260-64B Dual 1x6 SWITCHING
MODULE
1260-64C: <mod addr>. 1260-64C Single 1x6 SWITCHING
MODULE
Note the actual <header> sent is determined by the setting of the
ID Byte DIP switches on the module, and is independent of the
number of microwave relays installed.
User Manual 1260-64