Calibration — Type 1A5
24. Check High Frequency Response
a. Equipment setup is shown in Fig. 6-27.
b. Connect the following equipment to the A INPUT con
nector in the sequence given.
50 Ω GR-to-BNC Termination
5 ns GR Cable
5:1 50 Ω GR Type Attenuator
Type 191 Constant Amplitude Signal Generator
c. Set the oscilloscope Time/cm control to .5 ms.
d. CHECK — High frequency response for each of the set
ups given in Table 6-9, using the following procedure:
(1) Set the VOLTS/CM control to indicated setting.
(2) Set the Constant Amplitude Signal Generator fre
quency to 50 kHz and adjust its amplitude to provide 4 cm
display amplitude.
(3) Set the frequency of the Constant Amplitude Signal
Generator to the value opposite the applicable VOLTS/CM
setting in Table 6-9.
6-30
(4) CHEHK — 2.8 cm or more dis pl ay am p litude. Repeat
step 20 if results are unsatisfactory.
TABLE 6-9
Type 1A5
VOLTS/CM
1 mV
2 mV
5 mV
10 mV
20 mV
e. Disconnect the equipment. This conclud e s t he calibra -
tion procedure.
NOTE
See the applicable active probe manual for instruc
tions regarding the PROBE STEP ATTEN BAL adjust
ment procedure.
Constant Amplitude
Signal Generator
Frequency (MHz)
40
45
50
50
50